成人1型糖尿病患者治疗6个月时三种先进混合闭环系统的患者满意度:一项随访研究。

IF 1.8 4区 医学 Q4 ENDOCRINOLOGY & METABOLISM Endocrinologia Diabetes Y Nutricion Pub Date : 2023-10-01 DOI:10.1016/j.endien.2023.06.004
Victor Navas Moreno, Fernando Sebastian-Valles, Miguel Sampedro-Nuñez, Marcos Lahera Vargas, Monica Marazuela, José Alfonso Arranz Martin
{"title":"成人1型糖尿病患者治疗6个月时三种先进混合闭环系统的患者满意度:一项随访研究。","authors":"Victor Navas Moreno,&nbsp;Fernando Sebastian-Valles,&nbsp;Miguel Sampedro-Nuñez,&nbsp;Marcos Lahera Vargas,&nbsp;Monica Marazuela,&nbsp;José Alfonso Arranz Martin","doi":"10.1016/j.endien.2023.06.004","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p><span>Advanced hybrid closed-loop (AHCL) systems have demonstrated improved glycemic control in individuals with </span>Type 1 Diabetes Mellitus<span>. The aim of this study is to compare patient satisfaction among three available AHCL systems (Medtronic Minimed780 G, Roche Diabeloop DBLG1, and Tandem t:slim X2 Control IQ) after six months of treatment and to determine if it is related to glycemic control.</span></p></div><div><h3>Methods</h3><p>The data of 75 individuals were analyzed, including 15 using the DBLG1 system, 9 using Control IQ, and 51 using MM780 G. Patient satisfaction was assessed using the Diabetes Treatment Satisfaction Questionnaire for Diabetes Mellitus (DTSQc), a validated instrument.</p></div><div><h3>Results</h3><p><span>All systems demonstrated treatment satisfaction. The DBLG-1 system scored 14 (−15–21) points, while Control IQ scored 21 (9–24) and M780 G scored 19 (11–24) (p = 0.004). The multivariate analysis revealed that the DBLG-1 system is associated with a lower DTSQc score (OR 0.19, p = 0.019) independent of glycemic control, sex, age, duration of diabetes, duration as an </span>insulin pump user, and daily insulin dose.</p></div><div><h3>Conclusion</h3><p>AHCL systems are satisfactory treatments for users, with potential variations observed between each system regardless of the achieved glycemic control.</p></div>","PeriodicalId":48650,"journal":{"name":"Endocrinologia Diabetes Y Nutricion","volume":"70 8","pages":"Pages 548-555"},"PeriodicalIF":1.8000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Patient satisfaction in three advanced hybrid closed-loop systems at 6 months of treatment in adults with type 1 diabetes mellitus: a follow-up study\",\"authors\":\"Victor Navas Moreno,&nbsp;Fernando Sebastian-Valles,&nbsp;Miguel Sampedro-Nuñez,&nbsp;Marcos Lahera Vargas,&nbsp;Monica Marazuela,&nbsp;José Alfonso Arranz Martin\",\"doi\":\"10.1016/j.endien.2023.06.004\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p><span>Advanced hybrid closed-loop (AHCL) systems have demonstrated improved glycemic control in individuals with </span>Type 1 Diabetes Mellitus<span>. The aim of this study is to compare patient satisfaction among three available AHCL systems (Medtronic Minimed780 G, Roche Diabeloop DBLG1, and Tandem t:slim X2 Control IQ) after six months of treatment and to determine if it is related to glycemic control.</span></p></div><div><h3>Methods</h3><p>The data of 75 individuals were analyzed, including 15 using the DBLG1 system, 9 using Control IQ, and 51 using MM780 G. Patient satisfaction was assessed using the Diabetes Treatment Satisfaction Questionnaire for Diabetes Mellitus (DTSQc), a validated instrument.</p></div><div><h3>Results</h3><p><span>All systems demonstrated treatment satisfaction. The DBLG-1 system scored 14 (−15–21) points, while Control IQ scored 21 (9–24) and M780 G scored 19 (11–24) (p = 0.004). The multivariate analysis revealed that the DBLG-1 system is associated with a lower DTSQc score (OR 0.19, p = 0.019) independent of glycemic control, sex, age, duration of diabetes, duration as an </span>insulin pump user, and daily insulin dose.</p></div><div><h3>Conclusion</h3><p>AHCL systems are satisfactory treatments for users, with potential variations observed between each system regardless of the achieved glycemic control.</p></div>\",\"PeriodicalId\":48650,\"journal\":{\"name\":\"Endocrinologia Diabetes Y Nutricion\",\"volume\":\"70 8\",\"pages\":\"Pages 548-555\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2023-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Endocrinologia Diabetes Y Nutricion\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2530018023001348\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENDOCRINOLOGY & METABOLISM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Endocrinologia Diabetes Y Nutricion","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2530018023001348","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

摘要

引言:先进的混合闭环(AHCL)系统已证明改善了1型糖尿病患者的血糖控制。本研究的目的是比较三种可用AHCL系统(美敦力Minimed780G、罗氏Diabeloop DBLG1和Tandem t:slim X2 Control IQ)在治疗六个月后的患者满意度,并确定其是否与血糖控制有关。方法:对75名个体的数据进行分析,其中15名使用DBLG1系统,9名使用对照IQ,51名使用MM780G。使用经验证的糖尿病治疗满意度问卷(DTSQc)评估患者满意度。结果:所有系统均显示出治疗满意度。DBLG-1系统得分为14分(-15-21分),对照IQ得分为21分(9-24分),M780 G得分为19分(11-24分)(p=0.004分)。多变量分析显示,DBLG-1与较低的DTSQc得分(OR 0.19,p=0.019)相关,与血糖控制、性别、年龄、糖尿病持续时间、胰岛素泵使用者持续时间和每日胰岛素剂量无关。结论:AHCL系统是用户满意的治疗方法,无论血糖控制效果如何,每个系统之间都存在潜在的差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Patient satisfaction in three advanced hybrid closed-loop systems at 6 months of treatment in adults with type 1 diabetes mellitus: a follow-up study

Introduction

Advanced hybrid closed-loop (AHCL) systems have demonstrated improved glycemic control in individuals with Type 1 Diabetes Mellitus. The aim of this study is to compare patient satisfaction among three available AHCL systems (Medtronic Minimed780 G, Roche Diabeloop DBLG1, and Tandem t:slim X2 Control IQ) after six months of treatment and to determine if it is related to glycemic control.

Methods

The data of 75 individuals were analyzed, including 15 using the DBLG1 system, 9 using Control IQ, and 51 using MM780 G. Patient satisfaction was assessed using the Diabetes Treatment Satisfaction Questionnaire for Diabetes Mellitus (DTSQc), a validated instrument.

Results

All systems demonstrated treatment satisfaction. The DBLG-1 system scored 14 (−15–21) points, while Control IQ scored 21 (9–24) and M780 G scored 19 (11–24) (p = 0.004). The multivariate analysis revealed that the DBLG-1 system is associated with a lower DTSQc score (OR 0.19, p = 0.019) independent of glycemic control, sex, age, duration of diabetes, duration as an insulin pump user, and daily insulin dose.

Conclusion

AHCL systems are satisfactory treatments for users, with potential variations observed between each system regardless of the achieved glycemic control.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.10
自引率
10.50%
发文量
99
期刊介绍: Endocrinología, Diabetes y Nutrición is the official journal of the Spanish Society of Endocrinology and Nutrition (Sociedad Española de Endocrinología y Nutrición, SEEN) and the Spanish Society of Diabetes (Sociedad Española de Diabetes, SED), and was founded in 1954. The aim of the journal is to improve knowledge and be a useful tool in practice for clinical and laboratory specialists, trainee physicians, researchers, and nurses interested in endocrinology, diabetes, nutrition and related disciplines. It is an international journal published in Spanish (print and online) and English (online), covering different fields of endocrinology and metabolism, including diabetes, obesity, and nutrition disorders, as well as the most relevant research produced mainly in Spanish language territories. The quality of the contents is ensured by a prestigious national and international board, and by a selected panel of specialists involved in a rigorous peer review. The result is that only manuscripts containing high quality research and with utmost interest for clinicians and professionals related in the field are published. The Journal publishes Original clinical and research articles, Reviews, Special articles, Clinical Guidelines, Position Statements from both societies and Letters to the editor. Endocrinología, Diabetes y Nutrición can be found at Science Citation Index Expanded, Medline/PubMed and SCOPUS.
期刊最新文献
Are we underdiagnosing cognitive impairment in patients with type 2 diabetes mellitus? Automated insulin delivery systems in the treatment of diabetes: Benefits, challenges, and practical considerations in pediatric patients. Phase angle: Differences between disease-related malnutrition, obesity and healthy people. Early thyroid carcinoma in PTEN syndrome. Importance of immediate ultrasound screening. Minimally invasive techniques for the management of benign thyroid nodules: Efficacy and impact on quality of life of percutaneous ethanol injection. Results from a high-resolution thyroid nodule unit.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1