防紫外线隐形眼镜紫外线透射率的测定方法

Hermann Faubl, Michael H. Quinn
{"title":"防紫外线隐形眼镜紫外线透射率的测定方法","authors":"Hermann Faubl,&nbsp;Michael H. Quinn","doi":"10.1016/S0892-8967(98)00037-6","DOIUrl":null,"url":null,"abstract":"<div><p>Ultraviolet (UV) transmission spectra of UV-absorbing hydrogel contact lenses have been reported in the literature. Few of these studies were done following ANSI or ISO specified procedures. These published procedures do specify an integrating sphere in the detector system and a beam size, but not other instrument settings such as slit width, spectral smoothing functions, and scan speed. This work, done with Acuvue<sup>R</sup> and Precision UV<sup>TM</sup> contact lenses, shows that an integrating sphere is needed to ensure the accuracy of the data. A smaller beam size than specified was required to obtain accurate spectra. A slit width larger than recommended by the spectrophotometer’s manufacturer was needed to increase precision. Spectra results were unaffected by scan speed and spectral smoothing settings.</p></div>","PeriodicalId":80286,"journal":{"name":"International contact lens clinic (New York, N.Y.)","volume":"25 5","pages":"Pages 142-148"},"PeriodicalIF":0.0000,"publicationDate":"1998-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S0892-8967(98)00037-6","citationCount":"12","resultStr":"{\"title\":\"Methods for determining ultraviolet transmission of UV-blocking contact lenses\",\"authors\":\"Hermann Faubl,&nbsp;Michael H. Quinn\",\"doi\":\"10.1016/S0892-8967(98)00037-6\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>Ultraviolet (UV) transmission spectra of UV-absorbing hydrogel contact lenses have been reported in the literature. Few of these studies were done following ANSI or ISO specified procedures. These published procedures do specify an integrating sphere in the detector system and a beam size, but not other instrument settings such as slit width, spectral smoothing functions, and scan speed. This work, done with Acuvue<sup>R</sup> and Precision UV<sup>TM</sup> contact lenses, shows that an integrating sphere is needed to ensure the accuracy of the data. A smaller beam size than specified was required to obtain accurate spectra. A slit width larger than recommended by the spectrophotometer’s manufacturer was needed to increase precision. Spectra results were unaffected by scan speed and spectral smoothing settings.</p></div>\",\"PeriodicalId\":80286,\"journal\":{\"name\":\"International contact lens clinic (New York, N.Y.)\",\"volume\":\"25 5\",\"pages\":\"Pages 142-148\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/S0892-8967(98)00037-6\",\"citationCount\":\"12\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International contact lens clinic (New York, N.Y.)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0892896798000376\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International contact lens clinic (New York, N.Y.)","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0892896798000376","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 12

摘要

文献报道了吸收紫外线的水凝胶隐形眼镜的紫外透射光谱。这些研究很少是按照ANSI或ISO规定的程序进行的。这些公布的程序确实规定了探测器系统中的积分球和光束大小,但没有规定其他仪器设置,如狭缝宽度、光谱平滑功能和扫描速度。这项工作是用AcuvueR和Precision UVTM隐形眼镜完成的,表明需要一个积分球来确保数据的准确性。为了获得准确的光谱,需要比规定的光束尺寸更小的光束。需要比分光光度计制造商推荐的狭缝宽度更大的狭缝宽度来提高精度。光谱结果不受扫描速度和光谱平滑设置的影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Methods for determining ultraviolet transmission of UV-blocking contact lenses

Ultraviolet (UV) transmission spectra of UV-absorbing hydrogel contact lenses have been reported in the literature. Few of these studies were done following ANSI or ISO specified procedures. These published procedures do specify an integrating sphere in the detector system and a beam size, but not other instrument settings such as slit width, spectral smoothing functions, and scan speed. This work, done with AcuvueR and Precision UVTM contact lenses, shows that an integrating sphere is needed to ensure the accuracy of the data. A smaller beam size than specified was required to obtain accurate spectra. A slit width larger than recommended by the spectrophotometer’s manufacturer was needed to increase precision. Spectra results were unaffected by scan speed and spectral smoothing settings.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Prognostic Value of Flow-Mediated Vasodilation in Brachial Artery and Fingertip Artery for Cardiovascular Events: A Systematic Review and Meta-Analysis. Corneal pannus resolved with silicone hydrogel contact lenses: a case series Patient and practitioner perspectives on PureVision continuous wear lenses Cumulative clinical results and projected incident rates of microbial keratitis with PureVision™ silicone hydrogel lenses “Two steps forward, one step back”
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1