{"title":"经导管自扩张支架封闭静脉窦房间隔缺损,自行确定合适的支架长度,无需先进的成像技术-单中心病例系列经验","authors":"Chinnasamy Sivaprakasam Muthukumaran , Rajaguru Ganesan , Raja Vijayender Reddy , Radha Priya Yalamanchi , Refai Showkathali , Kanthallu Narayanamoorthy Srinivasan","doi":"10.1016/j.ihjccr.2021.10.011","DOIUrl":null,"url":null,"abstract":"<div><p>Transcatheter closure of sinus venous Atrial Septal Defect (SVASD) was proving to be safe and effective with accumulating evidence. We report 3 cases of successful transcatheter closure of SVASD with indigenous reproducible technique without need for advanced imagining techniques with favourable outcomes. Three patients aged 27,42 and 44 years underwent successful closure of SVASD using self-expandable Endurant™ II stent graft (Medtronic) after undergoing pre-procedure successful balloon occlusion of Superior Vena Cava (SVC) without obstructing right upper pulmonary vein (RUPV) drainage. We used fluoroscopy, <em>Trans</em>-esophageal echocardiography (TEE) and marker pigtail for guiding measurements to choose appropriate stent size. One patient developed cardiac tamponade requiring surgical removal of stent with surgical closure of the defect.</p></div>","PeriodicalId":100653,"journal":{"name":"IHJ Cardiovascular Case Reports (CVCR)","volume":"6 1","pages":"Pages 8-12"},"PeriodicalIF":0.0000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2468600X21000463/pdfft?md5=d6563731994a97821dc8239baad0b538&pid=1-s2.0-S2468600X21000463-main.pdf","citationCount":"1","resultStr":"{\"title\":\"Transcatheter self-expanding stent closure of sinus venosus atrial septal defects with indigenous method of determining appropriate stent length without advanced imaging techniques -A case series of single centre experience\",\"authors\":\"Chinnasamy Sivaprakasam Muthukumaran , Rajaguru Ganesan , Raja Vijayender Reddy , Radha Priya Yalamanchi , Refai Showkathali , Kanthallu Narayanamoorthy Srinivasan\",\"doi\":\"10.1016/j.ihjccr.2021.10.011\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>Transcatheter closure of sinus venous Atrial Septal Defect (SVASD) was proving to be safe and effective with accumulating evidence. We report 3 cases of successful transcatheter closure of SVASD with indigenous reproducible technique without need for advanced imagining techniques with favourable outcomes. Three patients aged 27,42 and 44 years underwent successful closure of SVASD using self-expandable Endurant™ II stent graft (Medtronic) after undergoing pre-procedure successful balloon occlusion of Superior Vena Cava (SVC) without obstructing right upper pulmonary vein (RUPV) drainage. We used fluoroscopy, <em>Trans</em>-esophageal echocardiography (TEE) and marker pigtail for guiding measurements to choose appropriate stent size. One patient developed cardiac tamponade requiring surgical removal of stent with surgical closure of the defect.</p></div>\",\"PeriodicalId\":100653,\"journal\":{\"name\":\"IHJ Cardiovascular Case Reports (CVCR)\",\"volume\":\"6 1\",\"pages\":\"Pages 8-12\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2468600X21000463/pdfft?md5=d6563731994a97821dc8239baad0b538&pid=1-s2.0-S2468600X21000463-main.pdf\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IHJ Cardiovascular Case Reports (CVCR)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2468600X21000463\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IHJ Cardiovascular Case Reports (CVCR)","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2468600X21000463","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Transcatheter self-expanding stent closure of sinus venosus atrial septal defects with indigenous method of determining appropriate stent length without advanced imaging techniques -A case series of single centre experience
Transcatheter closure of sinus venous Atrial Septal Defect (SVASD) was proving to be safe and effective with accumulating evidence. We report 3 cases of successful transcatheter closure of SVASD with indigenous reproducible technique without need for advanced imagining techniques with favourable outcomes. Three patients aged 27,42 and 44 years underwent successful closure of SVASD using self-expandable Endurant™ II stent graft (Medtronic) after undergoing pre-procedure successful balloon occlusion of Superior Vena Cava (SVC) without obstructing right upper pulmonary vein (RUPV) drainage. We used fluoroscopy, Trans-esophageal echocardiography (TEE) and marker pigtail for guiding measurements to choose appropriate stent size. One patient developed cardiac tamponade requiring surgical removal of stent with surgical closure of the defect.