{"title":"尼日利亚阿布贾大学教学医院数字计算机放射照相(cr) x线照片的拒绝重复分析。","authors":"Nemi Joshua, M. Sidi, Isaac Udugh","doi":"10.48153/jrrs.v35i2.224928","DOIUrl":null,"url":null,"abstract":"Background: Repeat-Reject Analysis (RRA) is a quality indicator and a critical tool for dose and image quality optimization in Radiology Departments. Aim: This study was aimed at evaluating the reject-repeat analysis of radiographs in digital computed radiography (CR) at the University of Abuja Teaching Hospital, Nigeria. Materials and Methods: The study was a prospective, cross-sectional design. The study was conducted at the Department of Radiology, University of Abuja st st Teaching Hospital (UATH), Gwagwalada, from 1 September 2020 to 31 March 2021. All the rejected-repeated images were used during the study period; 377 images were rejected. An adapted data capture sheet was used from the quality assurance workbook for Radiographers and Radiological Technologists. The obtained data was analyzed using micro soft excel and statistical package for social sciences (SPSS) version 22.0. Result: Out of the total of 1362 images that were acquired during the study period, 377 images were rejected. The overall repeat-reject rate was 27.67%. The highest examination was the chest 560 (41.375), followed by lower extremities 41 (10.87%). The lowest examinations were the paranasal sinuses 1 (0.26%) and Thoracic spine 2 (0.52%). The pelvic/hips had the highest repeat-reject rate 28 (52.82%), followed by the lumbosacral spine 29 (49.15%) and then post-nasal space 16 (47.05%) respectively. Conclusion: In this study, the overall repeat-reject rate was higher than the acceptable limit. The chest has the highest number of examinations. The pelvic/hips had the highest repeat-reject rate, followed by the lumbosacral spine and post-nasal space respectively.","PeriodicalId":16919,"journal":{"name":"Journal of Radiography and Radiation Sciences","volume":"108 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-11-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"REJECT-REPEAT ANALYSIS OF RADIOGRAPHS IN DIGITAL COMPUTED RADIOGRAPHY (CR) UNIVERSITY OF ABUJA TEACHING HOSPITAL, NIGERIA.\",\"authors\":\"Nemi Joshua, M. Sidi, Isaac Udugh\",\"doi\":\"10.48153/jrrs.v35i2.224928\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Background: Repeat-Reject Analysis (RRA) is a quality indicator and a critical tool for dose and image quality optimization in Radiology Departments. Aim: This study was aimed at evaluating the reject-repeat analysis of radiographs in digital computed radiography (CR) at the University of Abuja Teaching Hospital, Nigeria. Materials and Methods: The study was a prospective, cross-sectional design. The study was conducted at the Department of Radiology, University of Abuja st st Teaching Hospital (UATH), Gwagwalada, from 1 September 2020 to 31 March 2021. All the rejected-repeated images were used during the study period; 377 images were rejected. An adapted data capture sheet was used from the quality assurance workbook for Radiographers and Radiological Technologists. The obtained data was analyzed using micro soft excel and statistical package for social sciences (SPSS) version 22.0. Result: Out of the total of 1362 images that were acquired during the study period, 377 images were rejected. The overall repeat-reject rate was 27.67%. The highest examination was the chest 560 (41.375), followed by lower extremities 41 (10.87%). The lowest examinations were the paranasal sinuses 1 (0.26%) and Thoracic spine 2 (0.52%). The pelvic/hips had the highest repeat-reject rate 28 (52.82%), followed by the lumbosacral spine 29 (49.15%) and then post-nasal space 16 (47.05%) respectively. Conclusion: In this study, the overall repeat-reject rate was higher than the acceptable limit. The chest has the highest number of examinations. The pelvic/hips had the highest repeat-reject rate, followed by the lumbosacral spine and post-nasal space respectively.\",\"PeriodicalId\":16919,\"journal\":{\"name\":\"Journal of Radiography and Radiation Sciences\",\"volume\":\"108 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-11-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Radiography and Radiation Sciences\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.48153/jrrs.v35i2.224928\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiography and Radiation Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.48153/jrrs.v35i2.224928","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

背景:重复拒绝分析(RRA)是一种质量指标,也是放射科剂量和图像质量优化的重要工具。目的:本研究旨在评估尼日利亚阿布贾大学教学医院数字计算机放射照相(CR) x线片的拒绝重复分析。材料与方法:本研究为前瞻性横断面设计。该研究于2020年9月1日至2021年3月31日在瓜瓦拉达阿布贾大学附属教学医院放射科进行。所有被拒绝重复的图像在研究期间被使用;377张图片被拒绝。从放射技师和放射技师的质量保证工作手册中使用了一份经过修改的数据采集表。所得数据采用microsoftexcel和SPSS 22.0版本进行分析。结果:在研究期间获得的1362张图像中,有377张被拒绝。总重复拒绝率为27.67%。检查最多的是胸部560例(41.375例),其次是下肢41例(10.87%)。鼻窦1例(0.26%)和胸椎2例(0.52%)检查率最低。重复拒绝率最高的部位为盆腔/髋部28例(52.82%),其次为腰骶部29例(49.15%),鼻后间隙16例(47.05%)。结论:本研究总体重复拒绝率高于可接受限度。胸部检查次数最多。骨盆/髋部的重复排斥率最高,其次是腰骶脊柱和后鼻间隙。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
REJECT-REPEAT ANALYSIS OF RADIOGRAPHS IN DIGITAL COMPUTED RADIOGRAPHY (CR) UNIVERSITY OF ABUJA TEACHING HOSPITAL, NIGERIA.
Background: Repeat-Reject Analysis (RRA) is a quality indicator and a critical tool for dose and image quality optimization in Radiology Departments. Aim: This study was aimed at evaluating the reject-repeat analysis of radiographs in digital computed radiography (CR) at the University of Abuja Teaching Hospital, Nigeria. Materials and Methods: The study was a prospective, cross-sectional design. The study was conducted at the Department of Radiology, University of Abuja st st Teaching Hospital (UATH), Gwagwalada, from 1 September 2020 to 31 March 2021. All the rejected-repeated images were used during the study period; 377 images were rejected. An adapted data capture sheet was used from the quality assurance workbook for Radiographers and Radiological Technologists. The obtained data was analyzed using micro soft excel and statistical package for social sciences (SPSS) version 22.0. Result: Out of the total of 1362 images that were acquired during the study period, 377 images were rejected. The overall repeat-reject rate was 27.67%. The highest examination was the chest 560 (41.375), followed by lower extremities 41 (10.87%). The lowest examinations were the paranasal sinuses 1 (0.26%) and Thoracic spine 2 (0.52%). The pelvic/hips had the highest repeat-reject rate 28 (52.82%), followed by the lumbosacral spine 29 (49.15%) and then post-nasal space 16 (47.05%) respectively. Conclusion: In this study, the overall repeat-reject rate was higher than the acceptable limit. The chest has the highest number of examinations. The pelvic/hips had the highest repeat-reject rate, followed by the lumbosacral spine and post-nasal space respectively.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
CT features, Pathological findings and Clinical data of Follicular Dendritic Cell Sarcoma in the Mediastinum: Case Report and Literature Review KNOWLEDGE AND PRACTICE OF IMAGING OF CHILDREN WITH SUSPECTED CASES OF NON-ACCIDENTAL INJURIES AMONG RADIOGRAPHERS WITHIN MAIDUGURI METROPOLIS, NORTHEASTERN NIGERIA Effective doses in chest and abdominal radiography following the ICRP recommendations of 1991 and 2007 in a regional hospital. Investigation of the Effect of Solar Panel Temperature on Power Output Efficiency in Brass, Nigeria. An Assessment of the Lead rubber aprons in Radiodiagnostic centres in South-South Nigeria
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1