用解析和蒙特卡罗方法评估4兆电子伏特放射治疗掩体的光子束照射剂量当量率

Q3 Health Professions Frontiers in Biomedical Technologies Pub Date : 2023-03-14 DOI:10.18502/fbt.v10i2.12220
Reza Eghdam-Zamiri, Hosein Ghiasi, S. Keshavarz
{"title":"用解析和蒙特卡罗方法评估4兆电子伏特放射治疗掩体的光子束照射剂量当量率","authors":"Reza Eghdam-Zamiri, Hosein Ghiasi, S. Keshavarz","doi":"10.18502/fbt.v10i2.12220","DOIUrl":null,"url":null,"abstract":"Purpose: Skyshine radiation dose equivalent dose rate is known as scattered radiation by the room above air to points at the ground level points outside the Linear Accelerator (LINAC) room. Our aim was to estimate skyshine around the LINAC-based radiotherapy by a 4MV LINAC photon beam. \nMaterials and Methods: Monte Carlo (MC) MCNP code calculation was conducted to skyshine at the control room, 60Co treatment room, physics, and simulator rooms. National Council on Radiation Protection and Measurements (NCRP) 151 was also used and it reported analytical formulation methods for photon beam calculation. A Flattering Filter (FF) equipped and Flattening Filter-Free (FFF) LINACs photon beams were derived and differences and agreements were discussed. \nResults: The results showed high skyshine for FF equipped relative to FFF LINACs. This effect may be attributed to photon beam hardening by FF in the LINAC head and higher transmission through the ceiling shield and more presence of photons on the roof above the air. NCRP 151 method results were higher than MC simulated photon beam skyshine dose equivalent dose rate and it may be the cause of the inflexible analytical method in contrast to MC simulation. Finally, FFF and FF-equipped LINACs result in skyshine compared and they compared to NCRP 151 report. MC simulation performed reasonably in estimation in different conditions. \nConclusion: Our results showed that FF-equipped skyshine is higher than FFF LINAC and NCRP 151 is an inflexible method that does not take some effective parameters into account and calculates skyshine higher.","PeriodicalId":34203,"journal":{"name":"Frontiers in Biomedical Technologies","volume":"68 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-03-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Assessment of Photon Beam Skyshine Dose Equivalent Rate of a 4 MeV Radiation Therapy Bunker Using Analytical and Monte Carlo Methods\",\"authors\":\"Reza Eghdam-Zamiri, Hosein Ghiasi, S. Keshavarz\",\"doi\":\"10.18502/fbt.v10i2.12220\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Purpose: Skyshine radiation dose equivalent dose rate is known as scattered radiation by the room above air to points at the ground level points outside the Linear Accelerator (LINAC) room. Our aim was to estimate skyshine around the LINAC-based radiotherapy by a 4MV LINAC photon beam. \\nMaterials and Methods: Monte Carlo (MC) MCNP code calculation was conducted to skyshine at the control room, 60Co treatment room, physics, and simulator rooms. National Council on Radiation Protection and Measurements (NCRP) 151 was also used and it reported analytical formulation methods for photon beam calculation. A Flattering Filter (FF) equipped and Flattening Filter-Free (FFF) LINACs photon beams were derived and differences and agreements were discussed. \\nResults: The results showed high skyshine for FF equipped relative to FFF LINACs. This effect may be attributed to photon beam hardening by FF in the LINAC head and higher transmission through the ceiling shield and more presence of photons on the roof above the air. NCRP 151 method results were higher than MC simulated photon beam skyshine dose equivalent dose rate and it may be the cause of the inflexible analytical method in contrast to MC simulation. Finally, FFF and FF-equipped LINACs result in skyshine compared and they compared to NCRP 151 report. MC simulation performed reasonably in estimation in different conditions. \\nConclusion: Our results showed that FF-equipped skyshine is higher than FFF LINAC and NCRP 151 is an inflexible method that does not take some effective parameters into account and calculates skyshine higher.\",\"PeriodicalId\":34203,\"journal\":{\"name\":\"Frontiers in Biomedical Technologies\",\"volume\":\"68 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-03-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Frontiers in Biomedical Technologies\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18502/fbt.v10i2.12220\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in Biomedical Technologies","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18502/fbt.v10i2.12220","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

摘要

用途:天光辐射剂量当量剂量率被称为散射辐射,由房间上方的空气点到直线加速器(LINAC)房间外的地面点。我们的目的是通过4MV的LINAC光子束来估计基于LINAC的放射治疗周围的天光。材料与方法:对控制室、60Co治疗室、物理室、模拟室进行蒙特卡洛(MC) MCNP代码计算。国家辐射防护与测量委员会(NCRP)第151号报告了光子束计算的解析公式方法。推导了带奉承滤波器(FF)和不带平坦滤波器(FFF)的LINACs光子光束,并讨论了两者的差异和一致之处。结果:实验结果显示,相对于FFF LINACs,配备FFF的天效较高。这种效应可能是由于光子束在LINAC头部被FF硬化,通过天花板屏蔽的透射率更高,并且在空气上方的屋顶上存在更多的光子。NCRP - 151方法的计算结果高于MC模拟的光子束天射剂量等效剂量率,这可能是与MC模拟相比分析方法不够灵活的原因。最后,FFF和配备FFF的LINACs导致了天光的比较,他们比较了NCRP 151报告。MC模拟在不同条件下的估计效果都比较合理。结论:我们的研究结果表明,配备FFF的天光照高于FFF LINAC, NCRP 151是一种不灵活的方法,没有考虑一些有效参数,计算出更高的天光照。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Assessment of Photon Beam Skyshine Dose Equivalent Rate of a 4 MeV Radiation Therapy Bunker Using Analytical and Monte Carlo Methods
Purpose: Skyshine radiation dose equivalent dose rate is known as scattered radiation by the room above air to points at the ground level points outside the Linear Accelerator (LINAC) room. Our aim was to estimate skyshine around the LINAC-based radiotherapy by a 4MV LINAC photon beam. Materials and Methods: Monte Carlo (MC) MCNP code calculation was conducted to skyshine at the control room, 60Co treatment room, physics, and simulator rooms. National Council on Radiation Protection and Measurements (NCRP) 151 was also used and it reported analytical formulation methods for photon beam calculation. A Flattering Filter (FF) equipped and Flattening Filter-Free (FFF) LINACs photon beams were derived and differences and agreements were discussed. Results: The results showed high skyshine for FF equipped relative to FFF LINACs. This effect may be attributed to photon beam hardening by FF in the LINAC head and higher transmission through the ceiling shield and more presence of photons on the roof above the air. NCRP 151 method results were higher than MC simulated photon beam skyshine dose equivalent dose rate and it may be the cause of the inflexible analytical method in contrast to MC simulation. Finally, FFF and FF-equipped LINACs result in skyshine compared and they compared to NCRP 151 report. MC simulation performed reasonably in estimation in different conditions. Conclusion: Our results showed that FF-equipped skyshine is higher than FFF LINAC and NCRP 151 is an inflexible method that does not take some effective parameters into account and calculates skyshine higher.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Frontiers in Biomedical Technologies
Frontiers in Biomedical Technologies Health Professions-Medical Laboratory Technology
CiteScore
0.80
自引率
0.00%
发文量
34
审稿时长
12 weeks
期刊最新文献
AI in Nuclear Medical Applications: Challenges and Opportunities Evaluation of Eye-Blinking Dynamics in Human Emotion Recognition Using Weighted Visibility Graph Assessment of SPECT Image Reconstruction in Liver Scanning Using 99mTc/ EDDA/ HYNIC-TOCAssessment of SPECT Image Reconstruction in Liver Scanning Using 99mTc/ EDDA/ HYNIC-TOC Analysis of the Prevalence of Lumbar Annular Tears in Adult Patients Using Magnetic Resonance Imaging Data Grading the Dominant Pathological Indices in Liver Diseases from Pathological Images Using Radiomics Methods
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1