{"title":"低频示波器的相位校正","authors":"Joo-Gwang Lee, Chihyun Cho","doi":"10.1109/IRMMW-THZ.2015.7327761","DOIUrl":null,"url":null,"abstract":"This paper describes how to calibrate phase response of oscilloscopes, especially at low frequencies from 10 MHz to 1 GHz. The method uses the phase response of a standard mixer calibrated by three-mixer method as the reference phase. We describe basic principles and show results obtained using a balanced mixer and a real-time oscilloscope (1 GHz BW).","PeriodicalId":6577,"journal":{"name":"2015 40th International Conference on Infrared, Millimeter, and Terahertz waves (IRMMW-THz)","volume":"22 1","pages":"1-2"},"PeriodicalIF":0.0000,"publicationDate":"2015-11-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Phase calibration of oscilloscopes at low frequencies\",\"authors\":\"Joo-Gwang Lee, Chihyun Cho\",\"doi\":\"10.1109/IRMMW-THZ.2015.7327761\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper describes how to calibrate phase response of oscilloscopes, especially at low frequencies from 10 MHz to 1 GHz. The method uses the phase response of a standard mixer calibrated by three-mixer method as the reference phase. We describe basic principles and show results obtained using a balanced mixer and a real-time oscilloscope (1 GHz BW).\",\"PeriodicalId\":6577,\"journal\":{\"name\":\"2015 40th International Conference on Infrared, Millimeter, and Terahertz waves (IRMMW-THz)\",\"volume\":\"22 1\",\"pages\":\"1-2\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2015-11-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2015 40th International Conference on Infrared, Millimeter, and Terahertz waves (IRMMW-THz)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/IRMMW-THZ.2015.7327761\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2015 40th International Conference on Infrared, Millimeter, and Terahertz waves (IRMMW-THz)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IRMMW-THZ.2015.7327761","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Phase calibration of oscilloscopes at low frequencies
This paper describes how to calibrate phase response of oscilloscopes, especially at low frequencies from 10 MHz to 1 GHz. The method uses the phase response of a standard mixer calibrated by three-mixer method as the reference phase. We describe basic principles and show results obtained using a balanced mixer and a real-time oscilloscope (1 GHz BW).