{"title":"构筑区Roos平行板电离室极性效应评价","authors":"A. Al-Aghbari","doi":"10.2478/pjmpe-2022-0015","DOIUrl":null,"url":null,"abstract":"Abstract Purpose: Despite widespread studying of the polarity effect of Roos parallel plate ion chamber in electron beams as mentioned in several protocols, no investigations have up till now studied this effect in photon beams in the build-up region. It is important to examine its polarity effect in the build-up region for photon beams, so this is the first work that focuses in to evaluate the polarity effect of the Roos chamber in the surface and build-up region and comparing its effect with other chambers. Methods: In this study, the Roos chamber was irradiated by a Theratron 780E 60Co beam to a known polarity effect. The Polarity effects of 5×5 up to 35×35 cm2 field sizes at positive and negative polarizing voltages were measured in the build-up region from surface to 0.7 cm in a solid water phantom. Results: The polarity ratios (PRs) were obtained at 1.020 ± 0.00 and 1.015 ± 0.00 for field sizes 5 × 5 up to 35 × 35 cm2, respectively. For the same fields, the percentage of polarity effects (%PEs) was obtained at 1.99% ± 0.00% and 1.47% ± 0.02%, respectively. The results found that the %PEs decrease with increased field sizes and depths. Moreover, the %PEs exhibited a decrease with an increased percentage surface dose (%SD). The uncertainty of %PE was estimated as 0.01% for all measurements in this study. Conclusions: As a result, the average %PE of the Roos chamber described here is equal to 0.756% ± 0.013% for all depths and field sizes for the 60Co γ-ray beam. It has introduced a less percentage of polarity effect than other chambers.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":"29 1","pages":"127 - 132"},"PeriodicalIF":0.7000,"publicationDate":"2022-08-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Evaluation of the polarity effect of Roos parallel plate ionization chamber in build-up region\",\"authors\":\"A. Al-Aghbari\",\"doi\":\"10.2478/pjmpe-2022-0015\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Purpose: Despite widespread studying of the polarity effect of Roos parallel plate ion chamber in electron beams as mentioned in several protocols, no investigations have up till now studied this effect in photon beams in the build-up region. It is important to examine its polarity effect in the build-up region for photon beams, so this is the first work that focuses in to evaluate the polarity effect of the Roos chamber in the surface and build-up region and comparing its effect with other chambers. Methods: In this study, the Roos chamber was irradiated by a Theratron 780E 60Co beam to a known polarity effect. The Polarity effects of 5×5 up to 35×35 cm2 field sizes at positive and negative polarizing voltages were measured in the build-up region from surface to 0.7 cm in a solid water phantom. Results: The polarity ratios (PRs) were obtained at 1.020 ± 0.00 and 1.015 ± 0.00 for field sizes 5 × 5 up to 35 × 35 cm2, respectively. For the same fields, the percentage of polarity effects (%PEs) was obtained at 1.99% ± 0.00% and 1.47% ± 0.02%, respectively. The results found that the %PEs decrease with increased field sizes and depths. Moreover, the %PEs exhibited a decrease with an increased percentage surface dose (%SD). The uncertainty of %PE was estimated as 0.01% for all measurements in this study. Conclusions: As a result, the average %PE of the Roos chamber described here is equal to 0.756% ± 0.013% for all depths and field sizes for the 60Co γ-ray beam. It has introduced a less percentage of polarity effect than other chambers.\",\"PeriodicalId\":53955,\"journal\":{\"name\":\"Polish Journal of Medical Physics and Engineering\",\"volume\":\"29 1\",\"pages\":\"127 - 132\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2022-08-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Polish Journal of Medical Physics and Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2478/pjmpe-2022-0015\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2022-0015","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 1

摘要

摘要:目的:虽然有一些文献对Roos平行板离子室在电子束中的极性效应进行了广泛的研究,但迄今为止还没有研究在积聚区光子束中的极性效应。研究其在光子束积聚区域的极性效应是非常重要的,因此本文首次着重评估了Roos腔在表面和积聚区域的极性效应,并将其与其他腔的效应进行了比较。方法:在本研究中,用Theratron 780e60co光束照射Roos腔,产生已知的极性效应。在固体水模体中,从表面到0.7 cm的积累区域,测量了在正、负极化电压下5×5至35×35 cm2范围内电场大小的极性效应。结果:在5 × 5 ~ 35 × 35 cm2范围内,极性比(pr)分别为1.020±0.00和1.015±0.00。在相同电场下,极性效应百分比(%PEs)分别为1.99%±0.00%和1.47%±0.02%。结果表明,pe随井深和井径的增大而减小。此外,pe随表面剂量百分比(%SD)的增加而降低。本研究中所有测量结果的不确定度估计为0.01%。结论:对于60Co γ射线光束,在所有深度和视场尺寸下,Roos腔的平均%PE为0.756%±0.013%。它引入的极性效应百分比比其他腔室小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Evaluation of the polarity effect of Roos parallel plate ionization chamber in build-up region
Abstract Purpose: Despite widespread studying of the polarity effect of Roos parallel plate ion chamber in electron beams as mentioned in several protocols, no investigations have up till now studied this effect in photon beams in the build-up region. It is important to examine its polarity effect in the build-up region for photon beams, so this is the first work that focuses in to evaluate the polarity effect of the Roos chamber in the surface and build-up region and comparing its effect with other chambers. Methods: In this study, the Roos chamber was irradiated by a Theratron 780E 60Co beam to a known polarity effect. The Polarity effects of 5×5 up to 35×35 cm2 field sizes at positive and negative polarizing voltages were measured in the build-up region from surface to 0.7 cm in a solid water phantom. Results: The polarity ratios (PRs) were obtained at 1.020 ± 0.00 and 1.015 ± 0.00 for field sizes 5 × 5 up to 35 × 35 cm2, respectively. For the same fields, the percentage of polarity effects (%PEs) was obtained at 1.99% ± 0.00% and 1.47% ± 0.02%, respectively. The results found that the %PEs decrease with increased field sizes and depths. Moreover, the %PEs exhibited a decrease with an increased percentage surface dose (%SD). The uncertainty of %PE was estimated as 0.01% for all measurements in this study. Conclusions: As a result, the average %PE of the Roos chamber described here is equal to 0.756% ± 0.013% for all depths and field sizes for the 60Co γ-ray beam. It has introduced a less percentage of polarity effect than other chambers.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Polish Journal of Medical Physics and Engineering
Polish Journal of Medical Physics and Engineering RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.30
自引率
0.00%
发文量
19
期刊介绍: Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.
期刊最新文献
Parametrization of subsegmental infarcts using high spatial resolution 2DSTE and synthetic ultrasonic data Automated differential diagnostics of respiratory diseases using an electronic stethoscope Evaluating the impact of anatomical changes on dose distributions in head and neck cancer Comparing eDQE and eNEQ metrics – is there an alternative approach to assessing image quality in digital mammography? Vascular stiffness in cold pressor test hyper-reactors
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1