{"title":"二维阵列离子室验证Acuros BV计算的相对剂量分布的可行性研究","authors":"Qiang Zhao, Xiangyang Wu, Xiao-long Chang, Tao Feng, Di Yang, Kun Zhang, X. Qu, Xuemin Wang","doi":"10.3760/CMA.J.ISSN.0254-5098.2020.01.011","DOIUrl":null,"url":null,"abstract":"Objective \nTo study the feasibility of using the PTW729 2D array ion-chamber to verify the relative dose distribution calculated with the Acuros BV algorithm. Both advantages and disadvantages of the method were analyzed to provide reference for practical clinical practices. \n \n \nMethods \nBased on self-built measurement phantoms, the dose distribution on the same slice of the phantom was measured with PTW729 and film, respectively, under the same measurement condition and plan. The dose distributions obtained by the two method were compared with the result calculated with Acuros BV, separately, by using γ analytical tool. And the stability of the PTW729 was tested. \n \n \nResults \nThe γ comparison value was 95.9% between the film and Acuros BV, 98.9% between the PTW729 and Acuros BV and 88.0% between the film and PTW729, with 95.0%, 100.0%, and 100.0%, in their stability test respectively. \n \n \nConclusions \nPTW729 2D array ion-chamber can be applied to the rapid verification of Acuros BV algorithm-calculated relative dose distribution. \n \n \nKey words: \n2D array ion-chamber; Acuros BV; Brachy therapy","PeriodicalId":36403,"journal":{"name":"中华放射医学与防护杂志","volume":"11 1","pages":"59-63"},"PeriodicalIF":0.0000,"publicationDate":"2020-01-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Feasibility study of using two dimensional array ion-chamber to verifiy relative dose distribution calculated with Acuros BV\",\"authors\":\"Qiang Zhao, Xiangyang Wu, Xiao-long Chang, Tao Feng, Di Yang, Kun Zhang, X. Qu, Xuemin Wang\",\"doi\":\"10.3760/CMA.J.ISSN.0254-5098.2020.01.011\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo study the feasibility of using the PTW729 2D array ion-chamber to verify the relative dose distribution calculated with the Acuros BV algorithm. Both advantages and disadvantages of the method were analyzed to provide reference for practical clinical practices. \\n \\n \\nMethods \\nBased on self-built measurement phantoms, the dose distribution on the same slice of the phantom was measured with PTW729 and film, respectively, under the same measurement condition and plan. The dose distributions obtained by the two method were compared with the result calculated with Acuros BV, separately, by using γ analytical tool. And the stability of the PTW729 was tested. \\n \\n \\nResults \\nThe γ comparison value was 95.9% between the film and Acuros BV, 98.9% between the PTW729 and Acuros BV and 88.0% between the film and PTW729, with 95.0%, 100.0%, and 100.0%, in their stability test respectively. \\n \\n \\nConclusions \\nPTW729 2D array ion-chamber can be applied to the rapid verification of Acuros BV algorithm-calculated relative dose distribution. \\n \\n \\nKey words: \\n2D array ion-chamber; Acuros BV; Brachy therapy\",\"PeriodicalId\":36403,\"journal\":{\"name\":\"中华放射医学与防护杂志\",\"volume\":\"11 1\",\"pages\":\"59-63\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-01-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华放射医学与防护杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.01.011\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华放射医学与防护杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.0254-5098.2020.01.011","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的研究使用PTW729二维阵列离子室验证acros BV算法计算的相对剂量分布的可行性。分析了该方法的优缺点,为临床实践提供参考。方法在自建测量幻影的基础上,在相同的测量条件和计划下,分别用PTW729和薄膜测量幻影同一切片上的剂量分布。用γ分析工具将两种方法得到的剂量分布分别与acros BV计算的结果进行比较。对PTW729的稳定性进行了测试。结果薄膜与acros BV的γ比较值为95.9%,PTW729与acros BV的γ比较值为98.9%,PTW729与acros BV的γ比较值为88.0%,其稳定性测试值分别为95.0%、100.0%和100.0%。结论PTW729二维阵列离子室可用于Acuros BV算法计算的相对剂量分布的快速验证。关键词:二维阵列离子室;Acuros BV;短的治疗
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Feasibility study of using two dimensional array ion-chamber to verifiy relative dose distribution calculated with Acuros BV
Objective To study the feasibility of using the PTW729 2D array ion-chamber to verify the relative dose distribution calculated with the Acuros BV algorithm. Both advantages and disadvantages of the method were analyzed to provide reference for practical clinical practices. Methods Based on self-built measurement phantoms, the dose distribution on the same slice of the phantom was measured with PTW729 and film, respectively, under the same measurement condition and plan. The dose distributions obtained by the two method were compared with the result calculated with Acuros BV, separately, by using γ analytical tool. And the stability of the PTW729 was tested. Results The γ comparison value was 95.9% between the film and Acuros BV, 98.9% between the PTW729 and Acuros BV and 88.0% between the film and PTW729, with 95.0%, 100.0%, and 100.0%, in their stability test respectively. Conclusions PTW729 2D array ion-chamber can be applied to the rapid verification of Acuros BV algorithm-calculated relative dose distribution. Key words: 2D array ion-chamber; Acuros BV; Brachy therapy
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中华放射医学与防护杂志
中华放射医学与防护杂志 Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
0.60
自引率
0.00%
发文量
6377
期刊介绍:
期刊最新文献
Infection control status of novel coronavirus pneumonia and radiation safety at CT workplace Current status of the application of radiodiagnosis and radiotherapy resources in Huai'an City in 2020 The application value of low-dose CT scan in pregnant women with COVID-19 Eeffects of tube voltage on image quality and radiation dose of paranasal sinus CT examination Analyzing quality of life and influencing factors in patients with bone metastasis of hepatocellular carcinoma before and after radiotherapy by bone pain scale
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1