{"title":"注意在二氧化碳样品容器中使用Viton®或FETFE®o型环进行δ180分析","authors":"Kinga Revesz, Tyler B. Coplen","doi":"10.1016/0168-9622(91)90054-Z","DOIUrl":null,"url":null,"abstract":"<div><p>After 10 days, 3-μmol CO<sub>2</sub> samples in containers having glass stopcocks with Viton<sup>®</sup> or FETFE O-rings were enriched in <sup>18</sup>O by 1.5% as a result of absorption by the elastomer; this amount of enrichment is ∼20 times greater than the precision of <em>δ</em><sup>18</sup>O measurements of most laboratories. No change in <sup>13</sup>C content was observed. Increasing the sample size to 100 μmol resulted in an <sup>18</sup>O enrichment of 0.2% and did not affect the <sup>13</sup>C content.</p><p>Caution needs to be exercised in selecting sample containers for CO<sub>2</sub> isotope-ratio samples of < 200 μmol. If stopcocks are used in construction of containers for such samples, the use of all-glass stopcocks with Apiezon N<sup>®</sup> hydrocarbon-based grease will eliminate the fractionation of oxygen isotopes.</p></div>","PeriodicalId":100231,"journal":{"name":"Chemical Geology: Isotope Geoscience section","volume":"86 3","pages":"Pages 259-261"},"PeriodicalIF":0.0000,"publicationDate":"1991-02-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/0168-9622(91)90054-Z","citationCount":"6","resultStr":"{\"title\":\"Caution on the use of Viton® or FETFE® O-rings in carbon dioxide sample containers for δ180 analysis\",\"authors\":\"Kinga Revesz, Tyler B. Coplen\",\"doi\":\"10.1016/0168-9622(91)90054-Z\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>After 10 days, 3-μmol CO<sub>2</sub> samples in containers having glass stopcocks with Viton<sup>®</sup> or FETFE O-rings were enriched in <sup>18</sup>O by 1.5% as a result of absorption by the elastomer; this amount of enrichment is ∼20 times greater than the precision of <em>δ</em><sup>18</sup>O measurements of most laboratories. No change in <sup>13</sup>C content was observed. Increasing the sample size to 100 μmol resulted in an <sup>18</sup>O enrichment of 0.2% and did not affect the <sup>13</sup>C content.</p><p>Caution needs to be exercised in selecting sample containers for CO<sub>2</sub> isotope-ratio samples of < 200 μmol. If stopcocks are used in construction of containers for such samples, the use of all-glass stopcocks with Apiezon N<sup>®</sup> hydrocarbon-based grease will eliminate the fractionation of oxygen isotopes.</p></div>\",\"PeriodicalId\":100231,\"journal\":{\"name\":\"Chemical Geology: Isotope Geoscience section\",\"volume\":\"86 3\",\"pages\":\"Pages 259-261\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1991-02-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/0168-9622(91)90054-Z\",\"citationCount\":\"6\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Chemical Geology: Isotope Geoscience section\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/016896229190054Z\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Chemical Geology: Isotope Geoscience section","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/016896229190054Z","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6
摘要
10天后,装有Viton®或fefe o -环的玻璃塞的容器中的3 μmol CO2样品由于弹性体的吸收而富集了1.5%的18O;这个富集量是大多数实验室δ18O测量精度的20倍。13C含量未见变化。将样品量增加到100 μmol时,18O的富集量为0.2%,对13C的含量没有影响。在为<的CO2同位素比样品选择样品容器时需要谨慎;200μ摩尔。如果在构建此类样品的容器时使用stopcock,则使用带有Apiezon N®碳氢化合物基润滑脂的全玻璃stopcock将消除氧同位素的分异。
Caution on the use of Viton® or FETFE® O-rings in carbon dioxide sample containers for δ180 analysis
After 10 days, 3-μmol CO2 samples in containers having glass stopcocks with Viton® or FETFE O-rings were enriched in 18O by 1.5% as a result of absorption by the elastomer; this amount of enrichment is ∼20 times greater than the precision of δ18O measurements of most laboratories. No change in 13C content was observed. Increasing the sample size to 100 μmol resulted in an 18O enrichment of 0.2% and did not affect the 13C content.
Caution needs to be exercised in selecting sample containers for CO2 isotope-ratio samples of < 200 μmol. If stopcocks are used in construction of containers for such samples, the use of all-glass stopcocks with Apiezon N® hydrocarbon-based grease will eliminate the fractionation of oxygen isotopes.