IF 0.3 4区 医学Q3 MEDICINE, GENERAL & INTERNALHippokratiaPub Date : 2022-04-01
T Lirong, X Minmin, Y Manqiu, L Yanping
{"title":"通过化学发光分析仪对血栓分子标记物血栓调节蛋白、凝血酶-抗凝血酶复合物、纤溶酶-α2-纤溶酶抑制剂复合物和组织纤溶酶原激活物抑制剂复合物的性能评估。","authors":"T Lirong, X Minmin, Y Manqiu, L Yanping","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To comprehensively evaluate the primary performance regarding the task of detecting thrombomodulin (TM), thrombin-antithrombin complex (TAT), plasmin-α2-plasmin inhibitor complex (PIC), and tissue plasminogen activator/plasminogen activator inhibitor-1 complex (t-PAIC) by Sysmex HisCL5000 high sensitivity chemiluminescence analyzer.</p><p><strong>Methods: </strong>The performance of the chemiluminescence analyzer was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) documents for in-batch precision, daytime precision, carryover rate, linearity, and reference range.</p><p><strong>Results: </strong>The intra-batch and inter-day variation coefficients of the test items were all less than 5 %, and the contamination rate of each index was less than 10 %. The linear verification analysis showed that the correlation coefficients of TM, TAT, PIC, and t-PAIC were 0.9968, 0.9988, 0.9981, and 0.9930, respectively. The project recommended reference range was applicable to our laboratory.</p><p><strong>Conclusion: </strong>The high-sensitivity chemiluminescence analyzer has good performance in the detection of TM, TAT, PIC, and t-PAIC and is suitable for the detection of clinical specimens. HIPPOKRATIA 2022, 26 (2):78-82.</p>","PeriodicalId":50405,"journal":{"name":"Hippokratia","volume":"26 2","pages":"78-82"},"PeriodicalIF":0.3000,"publicationDate":"2022-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10177852/pdf/hippokratia-26-78.pdf","citationCount":"0","resultStr":"{\"title\":\"Performance evaluation of thrombus molecular markers thrombomodulin, thrombin-antithrombin complex, plasmin-α2-plasmin inhibitor complex, and tissue plasminogen activator-inhibitor complex by a chemiluminescence analyzer.\",\"authors\":\"T Lirong, X Minmin, Y Manqiu, L Yanping\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To comprehensively evaluate the primary performance regarding the task of detecting thrombomodulin (TM), thrombin-antithrombin complex (TAT), plasmin-α2-plasmin inhibitor complex (PIC), and tissue plasminogen activator/plasminogen activator inhibitor-1 complex (t-PAIC) by Sysmex HisCL5000 high sensitivity chemiluminescence analyzer.</p><p><strong>Methods: </strong>The performance of the chemiluminescence analyzer was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) documents for in-batch precision, daytime precision, carryover rate, linearity, and reference range.</p><p><strong>Results: </strong>The intra-batch and inter-day variation coefficients of the test items were all less than 5 %, and the contamination rate of each index was less than 10 %. The linear verification analysis showed that the correlation coefficients of TM, TAT, PIC, and t-PAIC were 0.9968, 0.9988, 0.9981, and 0.9930, respectively. The project recommended reference range was applicable to our laboratory.</p><p><strong>Conclusion: </strong>The high-sensitivity chemiluminescence analyzer has good performance in the detection of TM, TAT, PIC, and t-PAIC and is suitable for the detection of clinical specimens. HIPPOKRATIA 2022, 26 (2):78-82.</p>\",\"PeriodicalId\":50405,\"journal\":{\"name\":\"Hippokratia\",\"volume\":\"26 2\",\"pages\":\"78-82\"},\"PeriodicalIF\":0.3000,\"publicationDate\":\"2022-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10177852/pdf/hippokratia-26-78.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Hippokratia\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICINE, GENERAL & INTERNAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hippokratia","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
Performance evaluation of thrombus molecular markers thrombomodulin, thrombin-antithrombin complex, plasmin-α2-plasmin inhibitor complex, and tissue plasminogen activator-inhibitor complex by a chemiluminescence analyzer.
Objective: To comprehensively evaluate the primary performance regarding the task of detecting thrombomodulin (TM), thrombin-antithrombin complex (TAT), plasmin-α2-plasmin inhibitor complex (PIC), and tissue plasminogen activator/plasminogen activator inhibitor-1 complex (t-PAIC) by Sysmex HisCL5000 high sensitivity chemiluminescence analyzer.
Methods: The performance of the chemiluminescence analyzer was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) documents for in-batch precision, daytime precision, carryover rate, linearity, and reference range.
Results: The intra-batch and inter-day variation coefficients of the test items were all less than 5 %, and the contamination rate of each index was less than 10 %. The linear verification analysis showed that the correlation coefficients of TM, TAT, PIC, and t-PAIC were 0.9968, 0.9988, 0.9981, and 0.9930, respectively. The project recommended reference range was applicable to our laboratory.
Conclusion: The high-sensitivity chemiluminescence analyzer has good performance in the detection of TM, TAT, PIC, and t-PAIC and is suitable for the detection of clinical specimens. HIPPOKRATIA 2022, 26 (2):78-82.
期刊介绍:
Hippokratia journal is a quarterly issued, open access, peer reviewed, general medical journal, published in Thessaloniki, Greece. It is a forum for all medical specialties. The journal is published continuously since 1997, its official language is English and all submitted manuscripts undergo peer review by two independent reviewers, assigned by the Editor (double blinded review process).
Hippokratia journal is managed by its Editorial Board and has an International Advisory Committee and over 500 expert Reviewers covering all medical specialties and additionally Technical Reviewers, Statisticians, Image processing Experts and a journal Secretary. The Society “Friends of Hippokratia Journal” has the financial management of both the printed and electronic edition of the journal.