长轴与短轴视野免疫PET/CT:89Zr曲妥珠单抗的半定量评价。

IF 9.1 1区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Journal of Nuclear Medicine Pub Date : 2023-11-01 Epub Date: 2023-08-03 DOI:10.2967/jnumed.123.265621
Philipp Mohr, Joyce van Sluis, Laura Providência, Johannes H van Snick, Marjolijn N Lub-de Hooge, Antoon T Willemsen, Andor W J M Glaudemans, Ronald Boellaard, Adriaan A Lammertsma, Adrienne H Brouwers, Charalampos Tsoumpas
{"title":"长轴与短轴视野免疫PET/CT:89Zr曲妥珠单抗的半定量评价。","authors":"Philipp Mohr,&nbsp;Joyce van Sluis,&nbsp;Laura Providência,&nbsp;Johannes H van Snick,&nbsp;Marjolijn N Lub-de Hooge,&nbsp;Antoon T Willemsen,&nbsp;Andor W J M Glaudemans,&nbsp;Ronald Boellaard,&nbsp;Adriaan A Lammertsma,&nbsp;Adrienne H Brouwers,&nbsp;Charalampos Tsoumpas","doi":"10.2967/jnumed.123.265621","DOIUrl":null,"url":null,"abstract":"<p><p>The purpose of this study was to quantify any differences between the SUVs of <sup>89</sup>Zr immuno-PET scans obtained using a PET/CT system with a long axial field of view (LAFOV; Biograph Vision Quadra) compared to a PET/CT system with a short axial field of view (SAFOV; Biograph Vision) and to evaluate how LAFOV PET scan duration affects image noise and SUV metrics. <b>Methods:</b> Five metastatic breast cancer patients were scanned consecutively on SAFOV and LAFOV PET/CT scanners. Four additional patients were scanned using only LAFOV PET/CT. Scans on both systems lasted approximately 30 min and were acquired 4 d after injection of 37 MBq of <sup>89</sup>Zr-trastuzumab. LAFOV list-mode data were reprocessed to obtain images acquired using shorter scan durations (15, 10, 7.5, 5, and 3 min). Volumes of interest were placed in healthy tissues, and tumors were segmented semiautomatically to compare coefficients of variation and to perform Bland-Altman analysis on SUV metrics (SUV<sub>max</sub>, SUV<sub>peak</sub>, and SUV<sub>mean</sub>). <b>Results:</b> Using 30-min images, 2 commonly used lesion SUV metrics were higher for SAFOV than for LAFOV PET (SUV<sub>max</sub>, 16.2% ± 13.4%, and SUV<sub>peak</sub>, 10.1% ± 7.2%), whereas the SUV<sub>mean</sub> of healthy tissues showed minimal differences (0.7% ± 5.8%). Coefficients of variation in the liver derived from 30-min SAFOV PET were between those of 3- and 5-min LAFOV PET. The smallest SUV<sub>max</sub> and SUV<sub>peak</sub> differences between SAFOV and LAFOV were found for 3-min LAFOV PET. <b>Conclusion:</b> LAFOV <sup>89</sup>Zr immuno-PET showed a lower SUV<sub>max</sub> and SUV<sub>peak</sub> than SAFOV because of lower image noise. LAFOV PET scan duration may be reduced at the expense of increasing image noise and bias in SUV metrics. Nevertheless, SUV<sub>peak</sub> showed only minimal bias when reducing scan duration from 30 to 10 min.</p>","PeriodicalId":16758,"journal":{"name":"Journal of Nuclear Medicine","volume":" ","pages":"1815-1820"},"PeriodicalIF":9.1000,"publicationDate":"2023-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":"{\"title\":\"Long Versus Short Axial Field of View Immuno-PET/CT: Semiquantitative Evaluation for <sup>89</sup>Zr-Trastuzumab.\",\"authors\":\"Philipp Mohr,&nbsp;Joyce van Sluis,&nbsp;Laura Providência,&nbsp;Johannes H van Snick,&nbsp;Marjolijn N Lub-de Hooge,&nbsp;Antoon T Willemsen,&nbsp;Andor W J M Glaudemans,&nbsp;Ronald Boellaard,&nbsp;Adriaan A Lammertsma,&nbsp;Adrienne H Brouwers,&nbsp;Charalampos Tsoumpas\",\"doi\":\"10.2967/jnumed.123.265621\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The purpose of this study was to quantify any differences between the SUVs of <sup>89</sup>Zr immuno-PET scans obtained using a PET/CT system with a long axial field of view (LAFOV; Biograph Vision Quadra) compared to a PET/CT system with a short axial field of view (SAFOV; Biograph Vision) and to evaluate how LAFOV PET scan duration affects image noise and SUV metrics. <b>Methods:</b> Five metastatic breast cancer patients were scanned consecutively on SAFOV and LAFOV PET/CT scanners. Four additional patients were scanned using only LAFOV PET/CT. Scans on both systems lasted approximately 30 min and were acquired 4 d after injection of 37 MBq of <sup>89</sup>Zr-trastuzumab. LAFOV list-mode data were reprocessed to obtain images acquired using shorter scan durations (15, 10, 7.5, 5, and 3 min). Volumes of interest were placed in healthy tissues, and tumors were segmented semiautomatically to compare coefficients of variation and to perform Bland-Altman analysis on SUV metrics (SUV<sub>max</sub>, SUV<sub>peak</sub>, and SUV<sub>mean</sub>). <b>Results:</b> Using 30-min images, 2 commonly used lesion SUV metrics were higher for SAFOV than for LAFOV PET (SUV<sub>max</sub>, 16.2% ± 13.4%, and SUV<sub>peak</sub>, 10.1% ± 7.2%), whereas the SUV<sub>mean</sub> of healthy tissues showed minimal differences (0.7% ± 5.8%). Coefficients of variation in the liver derived from 30-min SAFOV PET were between those of 3- and 5-min LAFOV PET. The smallest SUV<sub>max</sub> and SUV<sub>peak</sub> differences between SAFOV and LAFOV were found for 3-min LAFOV PET. <b>Conclusion:</b> LAFOV <sup>89</sup>Zr immuno-PET showed a lower SUV<sub>max</sub> and SUV<sub>peak</sub> than SAFOV because of lower image noise. LAFOV PET scan duration may be reduced at the expense of increasing image noise and bias in SUV metrics. Nevertheless, SUV<sub>peak</sub> showed only minimal bias when reducing scan duration from 30 to 10 min.</p>\",\"PeriodicalId\":16758,\"journal\":{\"name\":\"Journal of Nuclear Medicine\",\"volume\":\" \",\"pages\":\"1815-1820\"},\"PeriodicalIF\":9.1000,\"publicationDate\":\"2023-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Nuclear Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.2967/jnumed.123.265621\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/8/3 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Nuclear Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2967/jnumed.123.265621","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/8/3 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 2

摘要

本研究的目的是量化使用具有长轴视场的PET/CT系统(LAFOV;Biograph-Vision Quadra)获得的89Zr免疫PET扫描的SUV与具有短轴视场的PET/COT系统(SAFOV;Biograph-Vision)之间的任何差异,并评估LAFOV PET扫描持续时间如何影响图像噪声和SUV指标。方法:对5例转移性癌症患者在SAFOV和LAFOV PET/CT扫描仪上连续扫描。另外四名患者仅使用LAFOV PET/CT进行扫描。两个系统的扫描持续了大约30次 分钟,并在注射37MBq的89Zr曲妥珠单抗后4天获得。对LAFOV列表模式数据进行再处理,以获得使用较短扫描持续时间(15、10、7.5、5和3 分钟)。将感兴趣的体积放置在健康组织中,并半自动分割肿瘤,以比较变异系数,并对SUV指标(SUVmax、SUVpeak和SUVmean)进行Bland-Altman分析。结果:使用30分钟图像,SAFOV的2个常用病变SUV指标高于LAFOV PET(SUVmax,16.2%±13.4%,SUVpeak,10.1%±7.2%),而健康组织的SUVmean差异最小(0.7%±5.8%)。30分钟SAFOV PET得出的肝脏变异系数介于3分钟和5分钟LAFOV PET。对于3分钟的LAFOV PET,发现SAFOV和LAFOV之间的SUVmax和SUVpeak差异最小。结论:LAFOV 89Zr免疫PET显示出比SAFOV更低的SUVmax和SUVpeak,因为图像噪声更低。LAFOV PET扫描持续时间可以以增加图像噪声和SUV度量中的偏差为代价来减少。然而,当扫描持续时间从30减少到10时,SUVpeak只显示出最小的偏差 最小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Long Versus Short Axial Field of View Immuno-PET/CT: Semiquantitative Evaluation for 89Zr-Trastuzumab.

The purpose of this study was to quantify any differences between the SUVs of 89Zr immuno-PET scans obtained using a PET/CT system with a long axial field of view (LAFOV; Biograph Vision Quadra) compared to a PET/CT system with a short axial field of view (SAFOV; Biograph Vision) and to evaluate how LAFOV PET scan duration affects image noise and SUV metrics. Methods: Five metastatic breast cancer patients were scanned consecutively on SAFOV and LAFOV PET/CT scanners. Four additional patients were scanned using only LAFOV PET/CT. Scans on both systems lasted approximately 30 min and were acquired 4 d after injection of 37 MBq of 89Zr-trastuzumab. LAFOV list-mode data were reprocessed to obtain images acquired using shorter scan durations (15, 10, 7.5, 5, and 3 min). Volumes of interest were placed in healthy tissues, and tumors were segmented semiautomatically to compare coefficients of variation and to perform Bland-Altman analysis on SUV metrics (SUVmax, SUVpeak, and SUVmean). Results: Using 30-min images, 2 commonly used lesion SUV metrics were higher for SAFOV than for LAFOV PET (SUVmax, 16.2% ± 13.4%, and SUVpeak, 10.1% ± 7.2%), whereas the SUVmean of healthy tissues showed minimal differences (0.7% ± 5.8%). Coefficients of variation in the liver derived from 30-min SAFOV PET were between those of 3- and 5-min LAFOV PET. The smallest SUVmax and SUVpeak differences between SAFOV and LAFOV were found for 3-min LAFOV PET. Conclusion: LAFOV 89Zr immuno-PET showed a lower SUVmax and SUVpeak than SAFOV because of lower image noise. LAFOV PET scan duration may be reduced at the expense of increasing image noise and bias in SUV metrics. Nevertheless, SUVpeak showed only minimal bias when reducing scan duration from 30 to 10 min.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Nuclear Medicine
Journal of Nuclear Medicine 医学-核医学
CiteScore
13.00
自引率
8.60%
发文量
340
审稿时长
1 months
期刊介绍: The Journal of Nuclear Medicine (JNM), self-published by the Society of Nuclear Medicine and Molecular Imaging (SNMMI), provides readers worldwide with clinical and basic science investigations, continuing education articles, reviews, employment opportunities, and updates on practice and research. In the 2022 Journal Citation Reports (released in June 2023), JNM ranked sixth in impact among 203 medical journals worldwide in the radiology, nuclear medicine, and medical imaging category.
期刊最新文献
C-X-C Motif Chemokine Receptor 4-Directed Scintigraphy of Multiple Myeloma Using [99mTc]Tc-PentixaTec. Debating the Future of Nuclear Medicine: The Greek Experience. Peptide Receptor Radionuclide Therapy Is Effective for Clinical Control of Symptomatic Metastatic Insulinoma: A Long-Term Retrospective Analysis Identification of (R)-[18F]YH134 for Monoacylglycerol Lipase Neuroimaging and Exploration of Its Use for Central Nervous System and Peripheral Drug Development Value of68Ga-FAPI-04 and18F-FDG PET/CT in Early Prediction of Pathologic Response to Neoadjuvant Chemotherapy in Locally Advanced Gastric Cancer
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1