不同质量测量中眼部超声和光学生物测量装置的比较。

IF 2.2 Q2 OPHTHALMOLOGY Journal of Optometry Pub Date : 2023-10-01 DOI:10.1016/j.optom.2023.05.001
Masoud Khorrami-Nejad , Ahmed Mohammed Khodair , Mehdi Khodaparast , Farshid Babapour Mofrad , Farzaneh Dehghanian Nasrabadi
{"title":"不同质量测量中眼部超声和光学生物测量装置的比较。","authors":"Masoud Khorrami-Nejad ,&nbsp;Ahmed Mohammed Khodair ,&nbsp;Mehdi Khodaparast ,&nbsp;Farshid Babapour Mofrad ,&nbsp;Farzaneh Dehghanian Nasrabadi","doi":"10.1016/j.optom.2023.05.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>To compare the reliability and agreement of axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) measurements obtained with optical biometry based on swept-source optical coherence tomography (IOLMaster 700; Carl Zeiss, Germany) and an ultrasound biometry device (Nidek; US-4000 Echoscan, Japan) in different qualities of AL measurement.</p></div><div><h3>Methods</h3><p>A total of 239 consecutive eyes of 239 cataract surgery candidates with a mean age of 56 ± 14 years were included. The quality measurements were grouped according to the quartiles of SD of the measured AL by IOLMaster 700. The first and fourth quartile's SD are defined as high and low-quality measurement, respectively, and the second and third quartiles’ SD is defined as moderate-quality.</p></div><div><h3>Results</h3><p>The reliability of AL and ACD between the two devices in all patients and in different quality measurement groups was excellent with highly statistically significant (AL: all ICC=0.999 and <em>P</em>&lt;0.001, ACD: all ICC&gt;0.920 and <em>P</em>&lt;0.001). AL and ACD in all quality measurements showed a very strong correlation between devices with highly statistically significant. However, there was poor (ICC=0.305), moderate (ICC=0.742), and good (ICC=0.843) reliability in measuring LT in low-, moderate-, and high-quality measurements, respectively. LT showed a very strong correlation (<em>r</em> = 0.854) with highly statistically significant (<em>P</em>&lt;0.001) between devices only in patients with high-quality measurements.</p></div><div><h3>Conclusions</h3><p>AL and ACD of the IOLMaster700 had outstanding agreements with the US-4000 ultrasound in different quality measurements of AL and can be used interchangeably. But LT should be used interchangeably cautiously only in the high-quality measurements group.</p></div>","PeriodicalId":46407,"journal":{"name":"Journal of Optometry","volume":"16 4","pages":"Pages 284-295"},"PeriodicalIF":2.2000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10518768/pdf/","citationCount":"0","resultStr":"{\"title\":\"Comparison of the ocular ultrasonic and optical biometry devices in the different quality measurements\",\"authors\":\"Masoud Khorrami-Nejad ,&nbsp;Ahmed Mohammed Khodair ,&nbsp;Mehdi Khodaparast ,&nbsp;Farshid Babapour Mofrad ,&nbsp;Farzaneh Dehghanian Nasrabadi\",\"doi\":\"10.1016/j.optom.2023.05.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><p>To compare the reliability and agreement of axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) measurements obtained with optical biometry based on swept-source optical coherence tomography (IOLMaster 700; Carl Zeiss, Germany) and an ultrasound biometry device (Nidek; US-4000 Echoscan, Japan) in different qualities of AL measurement.</p></div><div><h3>Methods</h3><p>A total of 239 consecutive eyes of 239 cataract surgery candidates with a mean age of 56 ± 14 years were included. The quality measurements were grouped according to the quartiles of SD of the measured AL by IOLMaster 700. The first and fourth quartile's SD are defined as high and low-quality measurement, respectively, and the second and third quartiles’ SD is defined as moderate-quality.</p></div><div><h3>Results</h3><p>The reliability of AL and ACD between the two devices in all patients and in different quality measurement groups was excellent with highly statistically significant (AL: all ICC=0.999 and <em>P</em>&lt;0.001, ACD: all ICC&gt;0.920 and <em>P</em>&lt;0.001). AL and ACD in all quality measurements showed a very strong correlation between devices with highly statistically significant. However, there was poor (ICC=0.305), moderate (ICC=0.742), and good (ICC=0.843) reliability in measuring LT in low-, moderate-, and high-quality measurements, respectively. LT showed a very strong correlation (<em>r</em> = 0.854) with highly statistically significant (<em>P</em>&lt;0.001) between devices only in patients with high-quality measurements.</p></div><div><h3>Conclusions</h3><p>AL and ACD of the IOLMaster700 had outstanding agreements with the US-4000 ultrasound in different quality measurements of AL and can be used interchangeably. But LT should be used interchangeably cautiously only in the high-quality measurements group.</p></div>\",\"PeriodicalId\":46407,\"journal\":{\"name\":\"Journal of Optometry\",\"volume\":\"16 4\",\"pages\":\"Pages 284-295\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2023-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10518768/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Optometry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1888429623000171\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Optometry","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1888429623000171","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:比较基于扫描源光学相干断层扫描(IOLMaster 700;Carl Zeiss,德国)和超声生物测量设备(Nidek;US-4000 Echoscan,日本)的光学生物测量在不同AL测量质量下获得的轴向长度(AL)、前房深度(ACD)和晶状体厚度(LT)测量的可靠性和一致性。方法:纳入239名白内障手术候选人的239只连续眼睛,平均年龄为56±14岁。根据IOLMaster 700测量的AL的SD四分位数对质量测量进行分组。第一和第四四分位数SD分别被定义为高质量和低质量测量,结果:在所有患者和不同质量测量组中,两种设备之间的AL和ACD的可靠性都非常好,具有高度统计学意义(AL:所有ICC=0.999、P0.920和PC结论:IOLMaster700的AL和ACD在AL的不同质量测量中与US-4000超声有显著的一致性,可以互换使用。但LT应仅在高质量测量组中谨慎地互换使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

摘要图片

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Comparison of the ocular ultrasonic and optical biometry devices in the different quality measurements

Purpose

To compare the reliability and agreement of axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) measurements obtained with optical biometry based on swept-source optical coherence tomography (IOLMaster 700; Carl Zeiss, Germany) and an ultrasound biometry device (Nidek; US-4000 Echoscan, Japan) in different qualities of AL measurement.

Methods

A total of 239 consecutive eyes of 239 cataract surgery candidates with a mean age of 56 ± 14 years were included. The quality measurements were grouped according to the quartiles of SD of the measured AL by IOLMaster 700. The first and fourth quartile's SD are defined as high and low-quality measurement, respectively, and the second and third quartiles’ SD is defined as moderate-quality.

Results

The reliability of AL and ACD between the two devices in all patients and in different quality measurement groups was excellent with highly statistically significant (AL: all ICC=0.999 and P<0.001, ACD: all ICC>0.920 and P<0.001). AL and ACD in all quality measurements showed a very strong correlation between devices with highly statistically significant. However, there was poor (ICC=0.305), moderate (ICC=0.742), and good (ICC=0.843) reliability in measuring LT in low-, moderate-, and high-quality measurements, respectively. LT showed a very strong correlation (r = 0.854) with highly statistically significant (P<0.001) between devices only in patients with high-quality measurements.

Conclusions

AL and ACD of the IOLMaster700 had outstanding agreements with the US-4000 ultrasound in different quality measurements of AL and can be used interchangeably. But LT should be used interchangeably cautiously only in the high-quality measurements group.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Optometry
Journal of Optometry OPHTHALMOLOGY-
CiteScore
5.20
自引率
0.00%
发文量
60
审稿时长
66 days
期刊最新文献
Blue light stimulation of the optic nerve head reduces melatonin levels in rabbit posterior segment Effect of a vergence-accommodation conflict induced during a 30-minute Virtual Reality game on vergence-accommodation parameters and related symptoms Visual performance of a new trifocal intraocular lens design evaluated with a clinical adaptative optics visual simulator Analysis of patient referrals from primary care to ophthalmology. The role of the optometrist Corneal higher-order aberrations in different types of irregular cornea
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1