A Structural Analysis of 3D Printed Pediatric Space Maintainers.

IF 1.5 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE Pediatric dentistry Pub Date : 2023-07-15
Lydia Watson, Brent Danley, Antheunis Versluis, Daranee Tantbirojn, Jeffrey Brooks, Martha H Wells
{"title":"A Structural Analysis of 3D Printed Pediatric Space Maintainers.","authors":"Lydia Watson,&nbsp;Brent Danley,&nbsp;Antheunis Versluis,&nbsp;Daranee Tantbirojn,&nbsp;Jeffrey Brooks,&nbsp;Martha H Wells","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p><b>Purpose:</b> To design, fabricate, and evaluate in vitro 3D printed space maintainers (SMs) and compare their retentive capabilities to tradi- tional stainless-steel (SS) space maintainers. <b>Methods:</b> E-Guard was selected as the printing material based on flexural strength and elastic modulus. SMs with a claw design were printed, cemented to testing blocks, and vertically loaded to determine failure strength and flexure (n equals 10). The intaglio surfaces of SM bands, printed with plain, crosshatched, or horizontal ridges, were cemented to extracted primary teeth. The force needed to dislodge the bands was compared to cemented traditional SS bands (n equals 14). <b>Results:</b> Flexural strength (mean±standard deviation [SD]) of E-Guide, E-Dent, and E-Guard materials was 65±12, 90±13, 134±25 MPa, respectively. Elastic modulus (mean±SD) was 1.54±0.40, 2.49±0.14, 2.65±0.89 GPa, respectively. When subjected to vertical loading, the mean failure load of 3D printed SMs (E-Guard) was 124 N and the mean deflection at fracture was 1.73 mm. Retention strengths (mean±SD) of 3D printed bands were 32±13, 43±13, 43±16 N for the plain printed, cross- hatched, and horizontal ridges designs, respectively. The retention strength of traditional SS bands was 126±27 N. <b>Conclusions:</b> E-Guard had superior mechanical properties among tested printing resins. Strength and deflection under the vertical load of claw-design 3D printed space maintainers may be adequate as a viable alternative to traditional SMs. Retention of 3D printed SMs was significantly lower than for traditional SS bands. Textured intaglio surfaces did not significantly improve the retention of 3D printed bands.</p>","PeriodicalId":19863,"journal":{"name":"Pediatric dentistry","volume":"45 4","pages":"342-347"},"PeriodicalIF":1.5000,"publicationDate":"2023-07-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pediatric dentistry","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To design, fabricate, and evaluate in vitro 3D printed space maintainers (SMs) and compare their retentive capabilities to tradi- tional stainless-steel (SS) space maintainers. Methods: E-Guard was selected as the printing material based on flexural strength and elastic modulus. SMs with a claw design were printed, cemented to testing blocks, and vertically loaded to determine failure strength and flexure (n equals 10). The intaglio surfaces of SM bands, printed with plain, crosshatched, or horizontal ridges, were cemented to extracted primary teeth. The force needed to dislodge the bands was compared to cemented traditional SS bands (n equals 14). Results: Flexural strength (mean±standard deviation [SD]) of E-Guide, E-Dent, and E-Guard materials was 65±12, 90±13, 134±25 MPa, respectively. Elastic modulus (mean±SD) was 1.54±0.40, 2.49±0.14, 2.65±0.89 GPa, respectively. When subjected to vertical loading, the mean failure load of 3D printed SMs (E-Guard) was 124 N and the mean deflection at fracture was 1.73 mm. Retention strengths (mean±SD) of 3D printed bands were 32±13, 43±13, 43±16 N for the plain printed, cross- hatched, and horizontal ridges designs, respectively. The retention strength of traditional SS bands was 126±27 N. Conclusions: E-Guard had superior mechanical properties among tested printing resins. Strength and deflection under the vertical load of claw-design 3D printed space maintainers may be adequate as a viable alternative to traditional SMs. Retention of 3D printed SMs was significantly lower than for traditional SS bands. Textured intaglio surfaces did not significantly improve the retention of 3D printed bands.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
3D打印儿童空间维护器的结构分析。
目的:设计、制造和评估体外3D打印空间维持器(SMs),并将其保持能力与传统不锈钢空间维持器(SS)进行比较。方法:根据弯曲强度和弹性模量选择E-Guard作为打印材料。打印带有爪形设计的SMs,将其粘合到测试块上,并垂直加载以确定破坏强度和挠度(n = 10)。SM带的凹版表面印有平纹、交叉纹或水平纹,与拔出的乳牙粘接。将拆除腕带所需的力与胶结的传统SS腕带(n = 14)进行比较。结果:E-Guide、E-Dent、E-Guard材料的抗弯强度(均数±标准差[SD])分别为65±12、90±13、134±25 MPa。弹性模量(平均±SD)分别为1.54±0.40、2.49±0.14、2.65±0.89 GPa。在垂直载荷作用下,3D打印SMs (E-Guard)的平均破坏载荷为124 N,断裂时的平均挠度为1.73 mm。平面打印、交叉孵化和水平脊设计的3D打印条带的保持强度(平均值±SD)分别为32±13、43±13和43±16 N。结论:E-Guard在印刷树脂中具有较好的力学性能。爪式3D打印空间维护者在垂直载荷下的强度和挠度可能足以作为传统SMs的可行替代方案。3D打印SMs的保留率明显低于传统SS波段。纹理凹印表面并没有显著提高3D打印腕带的保留率。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Pediatric dentistry
Pediatric dentistry DENTISTRY, ORAL SURGERY & MEDICINEPEDIATRI-PEDIATRICS
CiteScore
1.20
自引率
6.20%
发文量
74
审稿时长
>12 weeks
期刊介绍: Pediatric Dentistry is the official publication of the American Academy of Pediatric Dentistry, the American Board of Pediatric Dentistry and the College of Diplomates of the American Board of Pediatric Dentistry. It is published bi-monthly and is internationally recognized as the leading journal in the area of pediatric dentistry. The journal promotes the practice, education and research specifically related to the specialty of pediatric dentistry. This peer-reviewed journal features scientific articles, case reports and abstracts of current pediatric dental research.
期刊最新文献
A Randomized Clinical Trial of Short Fiber Reinforced Composite and Glass Hybrid Restoration for Molars Affected by Molar Hypomineralization. Oral Health in the Down Syndrome Population: Parental Perceptions on Dental Care in the United States. Early Introduction of Total Sugar Foods and Early Childhood Caries. Articaine Infiltration Versus Lidocaine Inferior Alveolar Nerve Block for Primary Mandibular Molars: A Double-blind Randomized Controlled Clinical Trial. Pre-, Peri-, and Postnatal Risk for the Development of Enamel Defects in Permanent Dentition: A Birth Cohort in Southern Brazil.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1