Brett S Mansfield, Belinda Stevens, Frederick J Raal, Farzahna Mohamed
{"title":"Point-of-care testing compared to gold-standard laboratory methods in the measurement of serum lipids.","authors":"Brett S Mansfield, Belinda Stevens, Frederick J Raal, Farzahna Mohamed","doi":"10.5830/CVJA-2023-039","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Cardiovascular disease is the leading cause of mortality worldwide, with dyslipidaemia being one of the major risk factors. Point-of-care testing (POCT) allows for the rapid measurement of serum lipids. The aim of this study was to assess the accuracy of serum lipid measurement by the Fujifilm<sup>™</sup> NX700 POCT compared to a gold-standard clinical laboratory method (Medpace, Leuven, Belgium).</p><p><strong>Methods: </strong>This was a prospective, observational study conducted at the Lipid Clinic at Charlotte Maxeke Johannesburg Academic Hospital from July to September 2022. Participants were known to have a lipid disorder, most commonly, familial hypercholesterolaemia. Samples sent for lipid measurement by standard laboratory methods were simultaneously measured by the Fujifilm<sup>™</sup> NX700 POCT.</p><p><strong>Results: </strong>Lipograms evaluating total cholesterol (TC), triglycerides (TG), high-density lipoprotein cholesterol (HDL-C) and calculated low-density lipoprotein cholesterol (LDL-C) were obtained from 115 participants. No statistically significant difference was noted between the parameters tested on the different platforms. The Fujifilm<sup>™</sup> NX700 POCT correctly identified > 91% of serum lipid results as normal or abnormal, as defined by NCEP-ATP III criteria, and exhibited good sensitivity and specificity for each parameter. Lin's concordance correlation coefficient demonstrated a strong correlation for all parameters; TC (ρ<sub>c</sub> = 0.9861), HDL-C (ρ<sub>c</sub> = 0.95919), LDL-C (ρ<sub>c</sub> = 0.98134) and TG (ρ<sub>c</sub> = 0.92775). Bland-Altman plots identified low bias and a good level of agreement between the two test methods.</p><p><strong>Conclusion: </strong>The Fujifilm<sup>™</sup> NX700 POCT compared favourably with gold-standard laboratory methods in the determination of serum lipid measurements, allowing for rapid screening at the primary healthcare level.</p>","PeriodicalId":9434,"journal":{"name":"Cardiovascular Journal of Africa","volume":null,"pages":null},"PeriodicalIF":0.7000,"publicationDate":"2023-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Journal of Africa","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5830/CVJA-2023-039","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Cardiovascular disease is the leading cause of mortality worldwide, with dyslipidaemia being one of the major risk factors. Point-of-care testing (POCT) allows for the rapid measurement of serum lipids. The aim of this study was to assess the accuracy of serum lipid measurement by the Fujifilm™ NX700 POCT compared to a gold-standard clinical laboratory method (Medpace, Leuven, Belgium).
Methods: This was a prospective, observational study conducted at the Lipid Clinic at Charlotte Maxeke Johannesburg Academic Hospital from July to September 2022. Participants were known to have a lipid disorder, most commonly, familial hypercholesterolaemia. Samples sent for lipid measurement by standard laboratory methods were simultaneously measured by the Fujifilm™ NX700 POCT.
Results: Lipograms evaluating total cholesterol (TC), triglycerides (TG), high-density lipoprotein cholesterol (HDL-C) and calculated low-density lipoprotein cholesterol (LDL-C) were obtained from 115 participants. No statistically significant difference was noted between the parameters tested on the different platforms. The Fujifilm™ NX700 POCT correctly identified > 91% of serum lipid results as normal or abnormal, as defined by NCEP-ATP III criteria, and exhibited good sensitivity and specificity for each parameter. Lin's concordance correlation coefficient demonstrated a strong correlation for all parameters; TC (ρc = 0.9861), HDL-C (ρc = 0.95919), LDL-C (ρc = 0.98134) and TG (ρc = 0.92775). Bland-Altman plots identified low bias and a good level of agreement between the two test methods.
Conclusion: The Fujifilm™ NX700 POCT compared favourably with gold-standard laboratory methods in the determination of serum lipid measurements, allowing for rapid screening at the primary healthcare level.
期刊介绍:
The Cardiovascular Journal of Africa (CVJA) is an international peer-reviewed journal that keeps cardiologists up to date with advances in the diagnosis and treatment of cardiovascular disease. Topics covered include coronary disease, electrophysiology, valve disease, imaging techniques, congenital heart disease (fetal, paediatric and adult), heart failure, surgery, and basic science.