Harriet S Julian, Göksu Kandemir, James Redfern, Thomas J Joyce, Daniel J Brown
{"title":"Impingement-related osteolysis in Motec total wrist arthroplasty: an explant analysis and review of the literature.","authors":"Harriet S Julian, Göksu Kandemir, James Redfern, Thomas J Joyce, Daniel J Brown","doi":"10.1177/17531934231194287","DOIUrl":null,"url":null,"abstract":"<p><p>We present five cases of osteolysis in the Motec total wrist prosthesis, three around the radial implant, one around the metacarpal implant and one around both. Three of these were progressive and required revision, and biomechanical explant analyses of these revised prostheses were performed. Ex vivo testing of the contact points of the Motec implants was also performed at maximum extension. Here, impingement occurs between the metacarpal screw and the dorsal rim of the cup (non-articulating surfaces) with the short-necked prosthesis, leading to metacarpal screw damage, titanium debris formation and osteolysis. An analysis of three previously published cases suggests that this may have been the likely mode of failure in those cases. This complication is preventable by avoiding use of the short-neck prosthesis.<b>Level of evidence:</b> IV.</p>","PeriodicalId":54813,"journal":{"name":"Journal of Hand Surgery-European Volume","volume":null,"pages":null},"PeriodicalIF":2.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Hand Surgery-European Volume","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/17531934231194287","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/9/4 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
We present five cases of osteolysis in the Motec total wrist prosthesis, three around the radial implant, one around the metacarpal implant and one around both. Three of these were progressive and required revision, and biomechanical explant analyses of these revised prostheses were performed. Ex vivo testing of the contact points of the Motec implants was also performed at maximum extension. Here, impingement occurs between the metacarpal screw and the dorsal rim of the cup (non-articulating surfaces) with the short-necked prosthesis, leading to metacarpal screw damage, titanium debris formation and osteolysis. An analysis of three previously published cases suggests that this may have been the likely mode of failure in those cases. This complication is preventable by avoiding use of the short-neck prosthesis.Level of evidence: IV.
期刊介绍:
The Journal of Hand Surgery (European Volume) is essential reading for everyone involved in restoring the function to the hand and upper limb. Dedicated to the needs of hand, plastic, reconstructive and orthopaedic surgeons, it publishes the best selection of current papers on hand surgery.
The journal regularly highlights key developments in a range of original, authoritative and highly informative articles written by distinguished experts from around the world. Rapid coverage of the latest research, techniques, trends and ideas to make sure you stay up-to-date.
Subjects covered include:
joint replacement
fracture and joint injuries
microvascular surgery
congenital hand problems
You will also find editorials, book reviews, conference dates, case studies, detailed coverage of new techniques and full-length reviews articles.