{"title":"Ocular biometry characteristics in cataract surgery candidates: a cross-sectional study.","authors":"Alireza Jamali, Taghi Naghdi, Mohsen Hosseini Abardeh, Mojgan Jamalzehi, Maedeh Khalajzadeh, Moslem Kamangar, Niloofar Tehranchi, Payam Nabovati","doi":"10.51329/mehdiophthal1416","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study was conducted to investigate ocular biometry parameters in cataract surgery candidates in northern Tehran, Iran using OA-2000 biometry device.</p><p><strong>Methods: </strong>In this cross-sectional study, values of ocular biometry parameters, including axial length (AL), anterior chamber depth (ACD), mean corneal curvature (mean K), lens thickness (LT), corneal astigmatism (CA), and white-to-white (WTW) of 818 eyes with cataracts, were measured using the OA-2000 biometry device (Tomey, Nagoya, Japan). The participants were divided into si age subgroups, in 10-year intervals. Finally, the values of the biometry parameters were calculated, and the trend of changes was examined for both age and sex subgroups.</p><p><strong>Results: </strong>The mean (standard deviation [SD]) of age of the participants was 63.82 (13.25) years. Mean (SD) of biometry parameters were as follows: AL, 23.36 (1.55) mm; ACD, 3.09 (0.40) mm; LT, 4.45 (0.55) mm; mean K, 44.51 (1.72) D; CA, 1.06 (0.94) D; and WTW, 11.81 (0.45) mm. Most of the parameters showed significant age-related changes in the total population. There was an increase in LT (<i>P</i> < 0.001) and mean K (<i>P</i> = 0.001), as well as a decrease in AL (<i>P</i> < 0.001) and ACD (<i>P</i> < 0.001) with age. Moreover, AL had a negative negligible correlation with LT (r = - 0.24; <i>P</i> < 0.001) and mean K (r = - 026; <i>P</i> < 0.001), as well as a weak positive correlation with ACD (r = + 0.44; <i>P</i> < 0.001).</p><p><strong>Conclusions: </strong>Our study revealed that the mean values of most biometric parameters varied across age and sex subgroups. Moreover, most of the parameters showed significant age-related changes in the total population.</p>","PeriodicalId":36524,"journal":{"name":"Medical Hypothesis, Discovery, and Innovation in Ophthalmology","volume":"10 1","pages":"11-17"},"PeriodicalIF":0.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/ce/55/mehdiophth-10-011.PMC10460215.pdf","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Hypothesis, Discovery, and Innovation in Ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.51329/mehdiophthal1416","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 1
Abstract
Background: This study was conducted to investigate ocular biometry parameters in cataract surgery candidates in northern Tehran, Iran using OA-2000 biometry device.
Methods: In this cross-sectional study, values of ocular biometry parameters, including axial length (AL), anterior chamber depth (ACD), mean corneal curvature (mean K), lens thickness (LT), corneal astigmatism (CA), and white-to-white (WTW) of 818 eyes with cataracts, were measured using the OA-2000 biometry device (Tomey, Nagoya, Japan). The participants were divided into si age subgroups, in 10-year intervals. Finally, the values of the biometry parameters were calculated, and the trend of changes was examined for both age and sex subgroups.
Results: The mean (standard deviation [SD]) of age of the participants was 63.82 (13.25) years. Mean (SD) of biometry parameters were as follows: AL, 23.36 (1.55) mm; ACD, 3.09 (0.40) mm; LT, 4.45 (0.55) mm; mean K, 44.51 (1.72) D; CA, 1.06 (0.94) D; and WTW, 11.81 (0.45) mm. Most of the parameters showed significant age-related changes in the total population. There was an increase in LT (P < 0.001) and mean K (P = 0.001), as well as a decrease in AL (P < 0.001) and ACD (P < 0.001) with age. Moreover, AL had a negative negligible correlation with LT (r = - 0.24; P < 0.001) and mean K (r = - 026; P < 0.001), as well as a weak positive correlation with ACD (r = + 0.44; P < 0.001).
Conclusions: Our study revealed that the mean values of most biometric parameters varied across age and sex subgroups. Moreover, most of the parameters showed significant age-related changes in the total population.