Fabio Mancino, Haley Tornberg, Christopher W Jones, Thomas A Bucher, Michael-Alexander Malahias
{"title":"The exeter cemented stem provides outstanding long-term fixation and bone load at 15 years follow-up: A systematic review and meta-analysis.","authors":"Fabio Mancino, Haley Tornberg, Christopher W Jones, Thomas A Bucher, Michael-Alexander Malahias","doi":"10.1177/10225536231153232","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Since its introduction in 1988, the double-tapered polished Exeter cemented stem has been widely adopted in primary total hip arthroplasty (THA). Despite the results coming from the arthroplasty registries have proven great survivorship, the aim of this study was to dig deeper and describe the modes of failure of the Exeter stem at 15 years follow-up while reporting the clinical and radiographic outcomes.</p><p><strong>Methods: </strong>A search of PubMed, MEDLINE, and Embase was performed using the Preferred Reporting Items for Systematic Review and Meta-Analyses since inception of database to January 2022. A meta-analysis was performed on stem's failure rates and clinical outcomes using random effects models. Publication bias was assessed with funnel plots.</p><p><strong>Results: </strong>Overall, ten studies met the inclusion criteria with 2167 hips at mean 14.8 ± 4.1 years follow-up. The meta-effect estimate for revision rate for stem-related reasons was 3.8% (CI 95% 2.1-5.6, <i>p</i> < 0.01). The meta-effect for revision rate for stem aseptic loosening (AL) was 0.22% (CI 95% 0-0.4, <i>p</i> = 0.048) and for periprosthetic fracture was 0.6% (CI95% 0.3-0.9, <i>p</i> < 0.001). The meta effect estimate for Oxford Hip Score (OHS) at final follow-up was 32.4 (moderate; CI 95% 23.2-41.6, <i>p</i> <0.001) with and heterogeneity among the studies of <i>I</i><sup>2</sup> 0%. Radiolucent lines were reported in 5.5% of cases, with 1.0% of cases (21 hips) reported to be progressive.</p><p><strong>Conclusion: </strong>Current evidence suggests that the Exeter cemented stem not only has proven long-term outstanding reliability with a revision rate of 3.8%, but also incredibly low revision rates for AL (0.22%) and periprosthetic fracture (0.6%). It is suitable for a variety of indications, and the consistent radiological appearances indicate durable fixation and load transmission while being associated with a remarkably low stem-related complication rate.</p>","PeriodicalId":48794,"journal":{"name":"Journal of Orthopaedic Surgery","volume":"31 1","pages":"10225536231153232"},"PeriodicalIF":1.3000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedic Surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/10225536231153232","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Since its introduction in 1988, the double-tapered polished Exeter cemented stem has been widely adopted in primary total hip arthroplasty (THA). Despite the results coming from the arthroplasty registries have proven great survivorship, the aim of this study was to dig deeper and describe the modes of failure of the Exeter stem at 15 years follow-up while reporting the clinical and radiographic outcomes.
Methods: A search of PubMed, MEDLINE, and Embase was performed using the Preferred Reporting Items for Systematic Review and Meta-Analyses since inception of database to January 2022. A meta-analysis was performed on stem's failure rates and clinical outcomes using random effects models. Publication bias was assessed with funnel plots.
Results: Overall, ten studies met the inclusion criteria with 2167 hips at mean 14.8 ± 4.1 years follow-up. The meta-effect estimate for revision rate for stem-related reasons was 3.8% (CI 95% 2.1-5.6, p < 0.01). The meta-effect for revision rate for stem aseptic loosening (AL) was 0.22% (CI 95% 0-0.4, p = 0.048) and for periprosthetic fracture was 0.6% (CI95% 0.3-0.9, p < 0.001). The meta effect estimate for Oxford Hip Score (OHS) at final follow-up was 32.4 (moderate; CI 95% 23.2-41.6, p <0.001) with and heterogeneity among the studies of I2 0%. Radiolucent lines were reported in 5.5% of cases, with 1.0% of cases (21 hips) reported to be progressive.
Conclusion: Current evidence suggests that the Exeter cemented stem not only has proven long-term outstanding reliability with a revision rate of 3.8%, but also incredibly low revision rates for AL (0.22%) and periprosthetic fracture (0.6%). It is suitable for a variety of indications, and the consistent radiological appearances indicate durable fixation and load transmission while being associated with a remarkably low stem-related complication rate.
期刊介绍:
Journal of Orthopaedic Surgery is an open access peer-reviewed journal publishing original reviews and research articles on all aspects of orthopaedic surgery. It is the official journal of the Asia Pacific Orthopaedic Association.
The journal welcomes and will publish materials of a diverse nature, from basic science research to clinical trials and surgical techniques. The journal encourages contributions from all parts of the world, but special emphasis is given to research of particular relevance to the Asia Pacific region.