Weili Pan, Wenjuan Wei, Yumeng Hu, Li Feng, Yongkui Ren, Xinsheng Li, Changling Li, Jun Jiang, Jianping Xiang, Xiaochang Leng, Da Yin
{"title":"Diagnostic accuracy of a novel optical coherence tomography-based fractional flow reserve algorithm for assessment of coronary stenosis significance.","authors":"Weili Pan, Wenjuan Wei, Yumeng Hu, Li Feng, Yongkui Ren, Xinsheng Li, Changling Li, Jun Jiang, Jianping Xiang, Xiaochang Leng, Da Yin","doi":"10.5603/cj.90744","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study aimed to introduce a novel optical coherence tomography-derived fractional flow reserve (FFR) computational approach and assess the diagnostic performance of the algorithm for assessing physiological function.</p><p><strong>Methods: </strong>The fusion of coronary optical coherence tomography and angiography was used to generate a novel FFR algorithm (AccuFFRoct) to evaluate functional ischemia of coronary stenosis. In the current study, a total of 34 consecutive patients were included, and AccuFFRoct was used to calculate the FFR for these patients. With the wire-measured FFR as the reference standard, we evaluated the performance of our approach by accuracy, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV).</p><p><strong>Results: </strong>Per vessel accuracy, sensitivity, specificity, PPV, and NPV for AccuFFRoct in identifying hemodynamically significant coronary stenosis were 93.8%, 94.7%, 92.3%, 94.7%, and 92.3%, respectively, were found. Good correlation (Pearson's correlation coefficient r = 0.80, p < 0.001) between AccuFFRoct and FFR was observed. The Bland-Altman analysis showed a mean difference value of -0.037 (limits of agreement: -0.189 to 0.115). The area under the receiver-operating characteristic curve (AUC) of AccuFFRoct in identifying physiologically significant stenosis was 0.94, which was higher than the minimum lumen area (MLA, AUC = 0.91) and significantly higher than the diameter stenosis (%DS, AUC = 0.78).</p><p><strong>Conclusions: </strong>This clinical study shows the efficiency and accuracy of AccuFFRoct for clinical implementation when using invasive FFR measurement as a reference. It could provide important insights into coronary imaging superior to current methods based on the degree of coronary artery stenosis.</p>","PeriodicalId":93923,"journal":{"name":"Cardiology journal","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11229798/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiology journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5603/cj.90744","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/11/15 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: This study aimed to introduce a novel optical coherence tomography-derived fractional flow reserve (FFR) computational approach and assess the diagnostic performance of the algorithm for assessing physiological function.
Methods: The fusion of coronary optical coherence tomography and angiography was used to generate a novel FFR algorithm (AccuFFRoct) to evaluate functional ischemia of coronary stenosis. In the current study, a total of 34 consecutive patients were included, and AccuFFRoct was used to calculate the FFR for these patients. With the wire-measured FFR as the reference standard, we evaluated the performance of our approach by accuracy, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV).
Results: Per vessel accuracy, sensitivity, specificity, PPV, and NPV for AccuFFRoct in identifying hemodynamically significant coronary stenosis were 93.8%, 94.7%, 92.3%, 94.7%, and 92.3%, respectively, were found. Good correlation (Pearson's correlation coefficient r = 0.80, p < 0.001) between AccuFFRoct and FFR was observed. The Bland-Altman analysis showed a mean difference value of -0.037 (limits of agreement: -0.189 to 0.115). The area under the receiver-operating characteristic curve (AUC) of AccuFFRoct in identifying physiologically significant stenosis was 0.94, which was higher than the minimum lumen area (MLA, AUC = 0.91) and significantly higher than the diameter stenosis (%DS, AUC = 0.78).
Conclusions: This clinical study shows the efficiency and accuracy of AccuFFRoct for clinical implementation when using invasive FFR measurement as a reference. It could provide important insights into coronary imaging superior to current methods based on the degree of coronary artery stenosis.