No clinical difference at mid-term follow-up between TiN-coated versus uncoated cemented mobile-bearing total knee arthroplasty: a matched cohort study.
{"title":"No clinical difference at mid-term follow-up between TiN-coated versus uncoated cemented mobile-bearing total knee arthroplasty: a matched cohort study.","authors":"Etienne Deroche, Cécile Batailler, Jobe Shatrov, Stanislas Gunst, Elvire Servien, Sébastien Lustig","doi":"10.1051/sicotj/2023001","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Nitride-based ceramic coating was introduced into surgical implants to improve hardness, reduce abrasion, and decrease the risk of metal-induced adverse reactions, especially for patients with suspected or identified metal hypersensitivity. The study aimed to evaluate the effectiveness and safety of a titanium nitride (TiN) coated prosthesis with a mobile bearing design.</p><p><strong>Methods: </strong>This was a retrospective matched-cohort study from a single center, comparing clinical outcomes between patients receiving either a TiN-coated versus an uncoated cobalt-chromium-molybdenum (CoCrMo) prostheses for primary total knee replacement. Seventeen patients received the TiN prosthesis between 2015 and 2019. These were matched 1:2 with patients receiving uncoated mobile-bearing knee prostheses with the same design manufacturer.</p><p><strong>Results: </strong>Fourteen patients in the TiN group had complete 5-year follow-up data and were compared with 34 patients from the CoCrMo group. The Knee Society Score was 170.6 ± 28.0 (Function subscore 83.7 ± 17.5 and Knee subscore 86.9 ± 13.8) in the TiN group and 180.7 ± 49.4 (Function subscore 87.5 ± 14.3 and Knee subscore 93.2 ± 9.6) in CoCrMo group, with no statistically significant difference (p = 0.19). One patient underwent a revision for instability requiring the removal of the implant in the TiN group and none in the CoCrMo group. The survival rates were 92.9% (CI95% 77.3-100.0) and 100.0% in the TiN group and CoCrMo group respectively (p = 1.0).</p><p><strong>Discussion: </strong>TiN-coated TKA with mobile bearing resulted in satisfactory clinical outcomes, and a low revision rate, and there was no complication related to the coated implant. The use of TiN-coated prostheses in case of confirmed or suspected metal allergy provides satisfactory short-term clinic outcomes.</p>","PeriodicalId":46378,"journal":{"name":"SICOT-J","volume":"9 ","pages":"5"},"PeriodicalIF":1.8000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9910165/pdf/","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"SICOT-J","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1051/sicotj/2023001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 1
Abstract
Introduction: Nitride-based ceramic coating was introduced into surgical implants to improve hardness, reduce abrasion, and decrease the risk of metal-induced adverse reactions, especially for patients with suspected or identified metal hypersensitivity. The study aimed to evaluate the effectiveness and safety of a titanium nitride (TiN) coated prosthesis with a mobile bearing design.
Methods: This was a retrospective matched-cohort study from a single center, comparing clinical outcomes between patients receiving either a TiN-coated versus an uncoated cobalt-chromium-molybdenum (CoCrMo) prostheses for primary total knee replacement. Seventeen patients received the TiN prosthesis between 2015 and 2019. These were matched 1:2 with patients receiving uncoated mobile-bearing knee prostheses with the same design manufacturer.
Results: Fourteen patients in the TiN group had complete 5-year follow-up data and were compared with 34 patients from the CoCrMo group. The Knee Society Score was 170.6 ± 28.0 (Function subscore 83.7 ± 17.5 and Knee subscore 86.9 ± 13.8) in the TiN group and 180.7 ± 49.4 (Function subscore 87.5 ± 14.3 and Knee subscore 93.2 ± 9.6) in CoCrMo group, with no statistically significant difference (p = 0.19). One patient underwent a revision for instability requiring the removal of the implant in the TiN group and none in the CoCrMo group. The survival rates were 92.9% (CI95% 77.3-100.0) and 100.0% in the TiN group and CoCrMo group respectively (p = 1.0).
Discussion: TiN-coated TKA with mobile bearing resulted in satisfactory clinical outcomes, and a low revision rate, and there was no complication related to the coated implant. The use of TiN-coated prostheses in case of confirmed or suspected metal allergy provides satisfactory short-term clinic outcomes.