A comparative assessment of intra- and inter- observer repeatability of three widely used software packages for the quantification of defect size in stress myocardial perfusion scintigraphy.
{"title":"A comparative assessment of intra- and inter- observer repeatability of three widely used software packages for the quantification of defect size in stress myocardial perfusion scintigraphy.","authors":"Christos Sachpekidis, Annette Kopp-Schneider, Vasileios Sachpekidis, Efstratios Moralidis","doi":"10.1967/s002449912516","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To assess the intra- and inter-observer repeatability of popular software packages for the quantitative determination of abnormality size in stress myocardial perfusion scintigraphy.</p><p><strong>Subjects and methods: </strong>A total of 182 tomographic stress myocardial perfusion scans were processed in duplicate by an experienced and trainee observer to assess SSSext (summed stress score multiplied by 100/68) and total defect extent (TDE), as % of the left ventricle, with 4 dimension-myocardial (4DM), emory cardiac toolbox (ECTb) and quantitative perfusion SPECT (QPS) packages. The Bland-Altman (B-A) analysis and Lin's concordance correlation coefficient (CCC) were used to assess agreement.</p><p><strong>Results: </strong>In SSSext's intra-observer repeatability, CCC showed substantial agreement for 4DM and QPS, and moderate agreement for ECTb for both observers. In inter-observer repeatability, CCC revealed substantial agreement for 4DM and QPS, and poor agreement for ECTb. Regarding TDE, CCC showed substantial intra-observer repeatability for both operators using all packages, while the inter-observer repeatability was substantial for 4DM and QPS, and moderate for ECTb.In SSSext's intra-observer repeatability for 4DM, ECTb and QPS, the B-A analysis provided (mean±1.96SD of paired measurements) 0.0±4.3, 0.2±7.8, -0.6±7.6 for the experienced physician and 0.2±5.9, 0.0±7.5, -0.5±5.4 for the trainee, respectively; in inter-observer repeatability it provided 0.2±5.4, 0.1±9.6, 0.2±8.1, respectively. Regarding TDE, the B-A values for intra-observer repeatability were 0.1±5.2, 0.1±7.9, 0.1±2.8 for the experienced reader and 0.3±6.6, -0.1±6.4, -0.1±2.4 for the trainee, respectively; in inter-observer agreement the B-A provided 0.6±6.4, -0.2±10.3, -0.1±4.3, respectively.</p><p><strong>Conclusion: </strong>Considerable differences in intra- and inter- observer agreement were noted for the quantitative determination of defect size using widely employed software packages, suggesting limitations in the clinical use of these measurements. Quantitative perfusion SPECT appears preferable, but with no significant advantage over 4DM. There were no significant differences between the observers.</p>","PeriodicalId":12871,"journal":{"name":"Hellenic journal of nuclear medicine","volume":"25 3","pages":"274-284"},"PeriodicalIF":0.9000,"publicationDate":"2022-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hellenic journal of nuclear medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1967/s002449912516","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: To assess the intra- and inter-observer repeatability of popular software packages for the quantitative determination of abnormality size in stress myocardial perfusion scintigraphy.
Subjects and methods: A total of 182 tomographic stress myocardial perfusion scans were processed in duplicate by an experienced and trainee observer to assess SSSext (summed stress score multiplied by 100/68) and total defect extent (TDE), as % of the left ventricle, with 4 dimension-myocardial (4DM), emory cardiac toolbox (ECTb) and quantitative perfusion SPECT (QPS) packages. The Bland-Altman (B-A) analysis and Lin's concordance correlation coefficient (CCC) were used to assess agreement.
Results: In SSSext's intra-observer repeatability, CCC showed substantial agreement for 4DM and QPS, and moderate agreement for ECTb for both observers. In inter-observer repeatability, CCC revealed substantial agreement for 4DM and QPS, and poor agreement for ECTb. Regarding TDE, CCC showed substantial intra-observer repeatability for both operators using all packages, while the inter-observer repeatability was substantial for 4DM and QPS, and moderate for ECTb.In SSSext's intra-observer repeatability for 4DM, ECTb and QPS, the B-A analysis provided (mean±1.96SD of paired measurements) 0.0±4.3, 0.2±7.8, -0.6±7.6 for the experienced physician and 0.2±5.9, 0.0±7.5, -0.5±5.4 for the trainee, respectively; in inter-observer repeatability it provided 0.2±5.4, 0.1±9.6, 0.2±8.1, respectively. Regarding TDE, the B-A values for intra-observer repeatability were 0.1±5.2, 0.1±7.9, 0.1±2.8 for the experienced reader and 0.3±6.6, -0.1±6.4, -0.1±2.4 for the trainee, respectively; in inter-observer agreement the B-A provided 0.6±6.4, -0.2±10.3, -0.1±4.3, respectively.
Conclusion: Considerable differences in intra- and inter- observer agreement were noted for the quantitative determination of defect size using widely employed software packages, suggesting limitations in the clinical use of these measurements. Quantitative perfusion SPECT appears preferable, but with no significant advantage over 4DM. There were no significant differences between the observers.
期刊介绍:
The Hellenic Journal of Nuclear Medicine published by the Hellenic Society of
Nuclear Medicine in Thessaloniki, aims to contribute to research, to education and
cover the scientific and professional interests of physicians, in the field of nuclear
medicine and in medicine in general. The journal may publish papers of nuclear
medicine and also papers that refer to related subjects as dosimetry, computer science,
targeting of gene expression, radioimmunoassay, radiation protection, biology, cell
trafficking, related historical brief reviews and other related subjects. Original papers
are preferred. The journal may after special agreement publish supplements covering
important subjects, dully reviewed and subscripted separately.