Comparative Evaluation of the Irradiance Emitted by Different Methods of Photopolymerization

Diogo de Azevedo Miranda, Anderson Quirino Prenassi, Gabriel Furtado Corrêa, Juliana Abdallah Mendes Costa
{"title":"Comparative Evaluation of the Irradiance Emitted by Different Methods of Photopolymerization","authors":"Diogo de Azevedo Miranda, Anderson Quirino Prenassi, Gabriel Furtado Corrêa, Juliana Abdallah Mendes Costa","doi":"10.32474/MADOHC.2018.03.000160","DOIUrl":null,"url":null,"abstract":"Introduction: To obtain success in the polymerization of restorative materials, obtaining a quality, it is necessary a light curing device and this in turn must be in proper operation. It is to be known that some other factors will influence these restorations as the emission and intensity of the light, adequate time of exposure and correct wavelength. Most of the three methods of light curing are used: High intensity in continuous light (High), increasing intensity (Soft-Start) and low (Low) intensity. The alteration between the above methods according to the time of exposure may cause different effects and should be chosen according to the need of the restorative procedure to be performed. Objective: To compare the photoactivation modes (High and Soft Start) emitted by two different devices (Dabi and Gnatus). All photo activators come from the Dental Clinic of Maternities Terezinha de Jesus Hospital. Method: An analysis of variance (ANOVA) was applied, considering the two device brands (Dabi and Gnatus) in the statistical model, so that the two photoactivation modes (High and Soft start) were evaluated in the same apparatus. All analyzes were performed in the R * program considering the level of significance of 5%. Results: It was observed that the Gnatus device presented higher mean irradiance (energy density), p <0.05, independent of the photoactivation mode. It is also observed that, for both devices, higher irradiance averages were observed with the high photoactivation mode (p <0.05). Conclusion: The two methods tested reached the minimum value indicated for polymerization of resinous materials, however, the Gnatus showed higher mean irradiance, and in both appliances the High method presented values statistically higher than the Soft Start method.","PeriodicalId":121604,"journal":{"name":"Modern Approaches in Dentistry and Oral Health Care","volume":"3 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2018-10-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Modern Approaches in Dentistry and Oral Health Care","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.32474/MADOHC.2018.03.000160","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction: To obtain success in the polymerization of restorative materials, obtaining a quality, it is necessary a light curing device and this in turn must be in proper operation. It is to be known that some other factors will influence these restorations as the emission and intensity of the light, adequate time of exposure and correct wavelength. Most of the three methods of light curing are used: High intensity in continuous light (High), increasing intensity (Soft-Start) and low (Low) intensity. The alteration between the above methods according to the time of exposure may cause different effects and should be chosen according to the need of the restorative procedure to be performed. Objective: To compare the photoactivation modes (High and Soft Start) emitted by two different devices (Dabi and Gnatus). All photo activators come from the Dental Clinic of Maternities Terezinha de Jesus Hospital. Method: An analysis of variance (ANOVA) was applied, considering the two device brands (Dabi and Gnatus) in the statistical model, so that the two photoactivation modes (High and Soft start) were evaluated in the same apparatus. All analyzes were performed in the R * program considering the level of significance of 5%. Results: It was observed that the Gnatus device presented higher mean irradiance (energy density), p <0.05, independent of the photoactivation mode. It is also observed that, for both devices, higher irradiance averages were observed with the high photoactivation mode (p <0.05). Conclusion: The two methods tested reached the minimum value indicated for polymerization of resinous materials, however, the Gnatus showed higher mean irradiance, and in both appliances the High method presented values statistically higher than the Soft Start method.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
不同光聚合方法辐照度的比较评价
简介:为了获得聚合修复材料的成功,获得质量,需要光固化装置,而光固化装置又必须正确操作。要知道,其他一些因素也会影响这些修复,如光的发射和强度、适当的曝光时间和正确的波长。光固化的三种方法大多采用:连续光照下的高强度(High)、增加强度(Soft-Start)和低强度(low)。根据暴露时间的不同,上述方法之间的变化可能会产生不同的效果,应根据所要进行的修复程序的需要来选择。目的:比较Dabi和Gnatus两种不同器件发出的高启动和软启动光激活模式。所有的光激活剂来自Terezinha de Jesus医院牙科诊所。方法:采用方差分析(ANOVA),在统计模型中考虑两个设备品牌(Dabi和Gnatus),在同一设备上评估两种光激活模式(高启动和软启动)。所有分析均在R *程序中进行,考虑显著性水平为5%。结果:与光激活方式无关,Gnatus器件的平均辐照度(能量密度)较高,p <0.05。还观察到,对于两种器件,高光激活模式下观察到更高的平均辐照度(p <0.05)。结论:两种方法均达到树脂材料聚合的最小值,但Gnatus的平均辐照度较高,且High方法的辐照度均高于Soft Start方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The contribution of fatigue on the transition from acute to chronic painful temporomandibular disorders, and its persistence: a prospective 3-month cohort study Perception of Cameroonian Parents about Orthodontic Tooth Positioner Advantages and Precautions of Universal Shade Composite Resin Epidemiological Aspects of Dental Anomalies in Moroccan Patients with Rare Diseases. Cultural Practice of Dental Health Among South Asians
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1