{"title":"Longitudinal and transverse uniformity in clinical whole body counting by an improved technique.","authors":"M M Rehani, M M Gupta, A K Gasu","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A new technique using a single NaI(Tl) detector (9 cm x 6.3 cm) in two positions and based on the response curves of the detector has been described. Also a different approach to select the energy band was adopted which is helpful in reducing the dependence of counting efficiency on source depth. Various portions of the compton continuum were taken rather than confining to valley and photo-peak. For transverse uniformity it was found that the energy band which gave 50% normalized count rate at half depth would give a resultant flat response with least variation. For longitudinal uniformity the separation between the two positions of the detector need to be nearly equal to the full width at half maximum of the response curve. The response curve coupling technique described here is different from the commonly used technique of iso-response curves.</p>","PeriodicalId":19324,"journal":{"name":"Nuclear-Medizin","volume":"13 4","pages":"379-88"},"PeriodicalIF":0.0000,"publicationDate":"1975-01-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear-Medizin","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
A new technique using a single NaI(Tl) detector (9 cm x 6.3 cm) in two positions and based on the response curves of the detector has been described. Also a different approach to select the energy band was adopted which is helpful in reducing the dependence of counting efficiency on source depth. Various portions of the compton continuum were taken rather than confining to valley and photo-peak. For transverse uniformity it was found that the energy band which gave 50% normalized count rate at half depth would give a resultant flat response with least variation. For longitudinal uniformity the separation between the two positions of the detector need to be nearly equal to the full width at half maximum of the response curve. The response curve coupling technique described here is different from the commonly used technique of iso-response curves.
本文描述了一种基于检测器的响应曲线,在两个位置使用单个NaI(Tl)探测器(9 cm x 6.3 cm)的新技术。同时采用了一种不同的能带选择方法,有助于降低计数效率对源深度的依赖。康普顿连续体的不同部分,而不是局限于谷和光峰。对于横向均匀性,发现在半深度处给予50%归一化计数率的能带将得到最小变化的平坦响应。为了保证纵向均匀性,探测器两个位置之间的距离需要接近于响应曲线半最大值处的全宽度。本文描述的响应曲线耦合技术不同于常用的等响应曲线技术。