{"title":"Metrological characteristics of measuring converters based on oscillistor effect","authors":"A. Cheredov, A. Shchelkanov","doi":"10.1109/DYNAMICS.2016.7818993","DOIUrl":null,"url":null,"abstract":"The characteristics of measuring converters with frequency output signal based on oscillistor effect in n-type germanium are described. We define the bounds of the linearity of static conversion characteristic. It has been found that the converters sensitivity increases at the temperature decrease. Methods of expanding of the conversion range are described. The results of the experimental studies of voltage and magnetic induction measuring converters at room and cryogenic temperature are presented in the paper.","PeriodicalId":293543,"journal":{"name":"2016 Dynamics of Systems, Mechanisms and Machines (Dynamics)","volume":"25 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2016-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2016 Dynamics of Systems, Mechanisms and Machines (Dynamics)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/DYNAMICS.2016.7818993","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
Abstract
The characteristics of measuring converters with frequency output signal based on oscillistor effect in n-type germanium are described. We define the bounds of the linearity of static conversion characteristic. It has been found that the converters sensitivity increases at the temperature decrease. Methods of expanding of the conversion range are described. The results of the experimental studies of voltage and magnetic induction measuring converters at room and cryogenic temperature are presented in the paper.