Could the unexpected excessive airway pressure be caused by water in the ventilator circuit?: a case report

Taehee Pyeon, J. Hwang, Sungmin Kim, H. Bae, Joungmin Kim
{"title":"Could the unexpected excessive airway pressure be caused by water in the ventilator circuit?: a case report","authors":"Taehee Pyeon, J. Hwang, Sungmin Kim, H. Bae, Joungmin Kim","doi":"10.30579/MBSE.2019.2.1.22","DOIUrl":null,"url":null,"abstract":"Humidification of inhaled gases has been standard of care in mechanical ventilation recently. Heat and moisture exchange is one of the most important functions of the upper respiratory system. The warming and humidification function through the upper airway disappears when mechanical ventilation with endotracheal intubation. An active humidification system is often used to preserve of warming and humidification function. But, we experienced of unexpected high inspiratory pressure and auto-positive end-expiratory pressure (PEEP) during mechanical ventilation with heated wire humidifier that one of active humidification system. We did not find any factor to raise airway pressure other than the water in the inspiratory limb of circuit. After some of sterile water in the circuit removed, high inspiratory pressure and auto-PEEP reduced. So we think that inspiratory limb water is a rare cause of airway pressure rise. Therefore, if sudden high peak airway pressure and auto-PEEP is encountered upon ventilation, the ventilator circuit status should be checked for the safety of patients.","PeriodicalId":259565,"journal":{"name":"Medical Biological Science and Engineering","volume":"8 5 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2019-01-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Biological Science and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.30579/MBSE.2019.2.1.22","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Humidification of inhaled gases has been standard of care in mechanical ventilation recently. Heat and moisture exchange is one of the most important functions of the upper respiratory system. The warming and humidification function through the upper airway disappears when mechanical ventilation with endotracheal intubation. An active humidification system is often used to preserve of warming and humidification function. But, we experienced of unexpected high inspiratory pressure and auto-positive end-expiratory pressure (PEEP) during mechanical ventilation with heated wire humidifier that one of active humidification system. We did not find any factor to raise airway pressure other than the water in the inspiratory limb of circuit. After some of sterile water in the circuit removed, high inspiratory pressure and auto-PEEP reduced. So we think that inspiratory limb water is a rare cause of airway pressure rise. Therefore, if sudden high peak airway pressure and auto-PEEP is encountered upon ventilation, the ventilator circuit status should be checked for the safety of patients.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
意外的气道压力过高是否可能是由呼吸机回路中的水引起的?一份病例报告
近来,吸入气体的加湿已成为机械通气的标准操作。热量和水分的交换是上呼吸道最重要的功能之一。机械通气加气管插管后,上呼吸道的增温加湿功能消失。主动加湿系统通常用于保持加温加湿功能。但是,在采用主动加湿系统之一的热丝加湿器进行机械通气时,我们遇到了意想不到的高吸气压力和自动呼气末正压(PEEP)。除呼吸道水外,未发现其他因素引起气道压力升高。除去回路中的部分无菌水后,高吸气压力和自动peep降低。因此,我们认为吸入肢体水是气道压力升高的罕见原因。因此,如果在通气过程中遇到突然的气道压力峰值和自动peep,为了患者的安全,应检查呼吸机回路状态。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Pinball-like free-floating right heart thrombus presenting as unexplained hypoxemia Management of sodium thiosulfate on subcutaneous and vascular calcification of right hand in end-stage renal disease: a case report Therapeutic experiences for herpes zoster and risk factors for zoster-associated pain: a retrospective single-center observational study Polymyositis, presenting with acute respiratory failure, misdiagnosed as Guillain-Barre syndrome Preparation and evaluation of histone H1.4 C-terminal peptide for cellular delivery of oligonucleotides
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1