Effect of pump characteristic (H-Q) curves on the efficiency of Total Artificial Heart

M. R. Islam, Meah Imtiaz Zulkarnain
{"title":"Effect of pump characteristic (H-Q) curves on the efficiency of Total Artificial Heart","authors":"M. R. Islam, Meah Imtiaz Zulkarnain","doi":"10.1109/ICAEE.2015.7506795","DOIUrl":null,"url":null,"abstract":"Nowadays, heart failure of human being is a very common occurrence all over the world. Total Artificial Heart (TAH) is an outstanding achievement of biomedical engineering where the patient's natural heart is removed and replaced with two mechanical pumps that provide perfusion to the systemic and pulmonary circulations. This paper focuses on the effect of pump characteristic (H-Q) curves on the efficiency of TAH. To show the effect of pump parameters, a simplified model of human cardiovascular system (CVS) is developed with an electrical circuit. A numerical simulation is conducted in order to find out the different physiological pressures and pump flows in response to change of parameters related to pump performance curves. Finally, the effect of H-Q curves on the efficiency of TAH is analyzed.","PeriodicalId":123939,"journal":{"name":"2015 International Conference on Advances in Electrical Engineering (ICAEE)","volume":"60 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2015-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2015 International Conference on Advances in Electrical Engineering (ICAEE)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ICAEE.2015.7506795","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Nowadays, heart failure of human being is a very common occurrence all over the world. Total Artificial Heart (TAH) is an outstanding achievement of biomedical engineering where the patient's natural heart is removed and replaced with two mechanical pumps that provide perfusion to the systemic and pulmonary circulations. This paper focuses on the effect of pump characteristic (H-Q) curves on the efficiency of TAH. To show the effect of pump parameters, a simplified model of human cardiovascular system (CVS) is developed with an electrical circuit. A numerical simulation is conducted in order to find out the different physiological pressures and pump flows in response to change of parameters related to pump performance curves. Finally, the effect of H-Q curves on the efficiency of TAH is analyzed.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
泵特性(H-Q)曲线对全人工心脏效率的影响
如今,人类心力衰竭在世界各地都很常见。全人工心脏(TAH)是生物医学工程的一项杰出成就,它将患者的天然心脏移除,并用两个机械泵代替,为体循环和肺循环提供灌注。本文重点研究了泵特性(H-Q)曲线对TAH效率的影响。为了说明泵参数的影响,建立了一个带有电路的人体心血管系统(CVS)简化模型。为了了解不同生理压力和泵流量对泵性能曲线相关参数变化的响应,进行了数值模拟。最后,分析了H-Q曲线对TAH效率的影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Problem analysis and probable solution of existing interconnected grid system of Chittagong region Development of an LCL filter for a residential Microgrid system Hardware implementation of grid connected transformer-less semi-Z-source inverter topology to mitigate common mode leakage current and THD Earthquake monitoring and warning system The next generation of irrigation
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1