{"title":"[Dosage errors and their prevention in intraocular SF6 injection for endotamponade in pneumatic retinopexy and after vitrectomy].","authors":"W Schrader,&nbsp;K Rodemann,&nbsp;B Schrader","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Sulfur hexafluoride (SF6) is used for internal tamponade during retinal detachment surgery. It is usually injected into the eye by a plastic syringe either with a sharp needle or by using the infusion system during pars plana vitrectomy. Although several suggestions exist for the appropriate application of intraocular SF6 and other perfluocarbons, a sudden postoperative rise in tension or hypotonia has been observed repeatedly. We simulated the preparation and application of mixtures of sulfur hexafluoride and air and analyzed possible mistakes. SF6 was quantitatively analyzed by infrared spectroscopy. SF6 does not rapidly diffuse out of capped plastic syringes made of polypropylene, as commonly employed in the Federal Republic of Germany. SF6 and air mix completely during aspiration. If the volume of microporefilters or tubes being used for preparation and instillation of mixtures of SF6 and air is ignored, this may result in a 20% deviation in SF6 concentration. Silicone tubes, which are widely used for the influsion during pars plana vitrectomy, are not recommended for the application of SF6. SF6 and fluorcarbones diffuse through silicone tubes and absorb and desorbe in the tube. Infusion lines made of polyethylene or polyvinylchloride, e.g. butterflies, should be preferred.</p>","PeriodicalId":12437,"journal":{"name":"Fortschritte der Ophthalmologie : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft","volume":"88 6","pages":"633-6"},"PeriodicalIF":0.0000,"publicationDate":"1991-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Fortschritte der Ophthalmologie : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Sulfur hexafluoride (SF6) is used for internal tamponade during retinal detachment surgery. It is usually injected into the eye by a plastic syringe either with a sharp needle or by using the infusion system during pars plana vitrectomy. Although several suggestions exist for the appropriate application of intraocular SF6 and other perfluocarbons, a sudden postoperative rise in tension or hypotonia has been observed repeatedly. We simulated the preparation and application of mixtures of sulfur hexafluoride and air and analyzed possible mistakes. SF6 was quantitatively analyzed by infrared spectroscopy. SF6 does not rapidly diffuse out of capped plastic syringes made of polypropylene, as commonly employed in the Federal Republic of Germany. SF6 and air mix completely during aspiration. If the volume of microporefilters or tubes being used for preparation and instillation of mixtures of SF6 and air is ignored, this may result in a 20% deviation in SF6 concentration. Silicone tubes, which are widely used for the influsion during pars plana vitrectomy, are not recommended for the application of SF6. SF6 and fluorcarbones diffuse through silicone tubes and absorb and desorbe in the tube. Infusion lines made of polyethylene or polyvinylchloride, e.g. butterflies, should be preferred.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[气动视网膜固定术和玻璃体切除术后眼内注射SF6治疗内膜填塞的剂量错误及预防]。
在视网膜脱离手术中,六氟化硫(SF6)用于内部填塞。它通常通过塑料注射器注射到眼睛中,或者用锋利的针头注射,或者在玻璃体切割手术中使用输注系统注射。尽管存在一些建议,建议适当应用眼内SF6和其他全氟碳化合物,但反复观察到术后突然升高的紧张或低张力。模拟了六氟化硫与空气混合物的制备和应用过程,分析了可能出现的错误。利用红外光谱对SF6进行定量分析。SF6不会像德意志联邦共和国通常使用的那样,从聚丙烯制成的有盖塑料注射器中迅速扩散出去。吸入过程中SF6与空气完全混合。如果忽略用于制备和注入SF6与空气混合物的微孔过滤器或管的体积,则可能导致SF6浓度偏差20%。硅酮管在玻璃体切割手术中广泛应用,不推荐用于SF6。SF6和氟碳化合物通过硅胶管扩散并在管中吸收和解吸。应优先使用聚乙烯或聚氯乙烯制成的输液管,例如蝶形输液管。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Thermoradiotherapy for Malignant Choroidal Melanoma: The Development of a New Microwave Hyperthermia System [Planimetric follow-up of cytomegalovirus retinitis treated with ganciclovir in patients with AIDS]. [Detection of HCMV-DNA in aqueous humor for diagnostic confirmation of viral retinitis]. [Herpes simplex virus retinitis. Role of the immune system in the animal experiment]. [Electro-ophthalmologic studies in HIV patients].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1