{"title":"Improving the Performances of a Reverberation Chamber: a Real Case","authors":"M. Pocai, I. Dotto, D. Festa, G. D'Abreu","doi":"10.1109/EMCZUR.2009.4783388","DOIUrl":null,"url":null,"abstract":"In a previous measurement campaign [1] it was found that the reverberation chamber (RC) of CISAM EMC Department evidenced anomalous values of the Quality factor (Q) - too low to be \"right\" for a fully aluminum chamber - and moreover, after some use it became absolutely evident that it was too easy to improve - lowering it, of course - the value of the Lowest Usable Frequency (LUF) of this chamber. After some investigations on the chamber materials/components and on the assembly procedures, the authors developed the idea that this strange behavior had to be attributed to the tuners - being the original ones made with somehow absorbing material - and therefore they decided to replace them with new, all metal tuners. In this paper, the results of a calibration campaign on the reverberation chamber after tuner replacement are presented, as well as the results of an emission test on a standard reference source before and after the replacement. The calibration has been done in the frequency range 80 MHz ÷ 3 GHz, and the emission tests have been performed in the range 100 MHz to 1 GHz, which is exactly the frequency range of the reference radiator used to generate a standard spectrum. The comparison between the results obtained, under the same conditions, with the old and with the new tuners is presented, and the goodness of the assumptions demonstrated: with the new tuners the performances of the chamber have been considerably improved.","PeriodicalId":192851,"journal":{"name":"2009 20th International Zurich Symposium on Electromagnetic Compatibility","volume":"49 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2009-02-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2009 20th International Zurich Symposium on Electromagnetic Compatibility","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/EMCZUR.2009.4783388","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5
Abstract
In a previous measurement campaign [1] it was found that the reverberation chamber (RC) of CISAM EMC Department evidenced anomalous values of the Quality factor (Q) - too low to be "right" for a fully aluminum chamber - and moreover, after some use it became absolutely evident that it was too easy to improve - lowering it, of course - the value of the Lowest Usable Frequency (LUF) of this chamber. After some investigations on the chamber materials/components and on the assembly procedures, the authors developed the idea that this strange behavior had to be attributed to the tuners - being the original ones made with somehow absorbing material - and therefore they decided to replace them with new, all metal tuners. In this paper, the results of a calibration campaign on the reverberation chamber after tuner replacement are presented, as well as the results of an emission test on a standard reference source before and after the replacement. The calibration has been done in the frequency range 80 MHz ÷ 3 GHz, and the emission tests have been performed in the range 100 MHz to 1 GHz, which is exactly the frequency range of the reference radiator used to generate a standard spectrum. The comparison between the results obtained, under the same conditions, with the old and with the new tuners is presented, and the goodness of the assumptions demonstrated: with the new tuners the performances of the chamber have been considerably improved.