F. Breithuth, J. Schiefer, M. Arn, S. Peters, W. Seelentag
{"title":"Determination of the source dwell position of an afterloading device with a detector array","authors":"F. Breithuth, J. Schiefer, M. Arn, S. Peters, W. Seelentag","doi":"10.5166/JROI-3-2-18","DOIUrl":null,"url":null,"abstract":"It was the aim to develop and test a measurement technique for the source position of an afterloading system using an electronic two-dimensional detector array (2D array). A \"GammaMed plus IX\" high dose rate afterloading device (Varian Medical Systems, Palo Alto, USA), and a Seven29 2D detector array (PTW Freiburg GmbH, Freiburg, Germany) have been used. A hollow needle has been connected to the afterloading device. Its outer diameter is 3.0 mm and its length about 220mm. A 14 mm thick Perspex slab was fixed in a reproducible position on the 2D array. Above the 14th detector row, a groove was cut in the slab which accommodates the hollow needle in a fixed position relative to the 2D array. In order to define the position of the source, the signals of the detectors in the 14th detector row have been evaluated. A theoretical curve depending on the source position and strength, has been fitted to the acquired detector signals. It has been shown that the reproducibility of the measurement technique - including the reproducibility of the source placement - was within 0.15mm. This method can not only replace film measurements, it is also more exact and less time consuming.\n","PeriodicalId":426862,"journal":{"name":"Journal of Radiation Oncology Informatics","volume":"14 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2011-02-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiation Oncology Informatics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5166/JROI-3-2-18","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
Abstract
It was the aim to develop and test a measurement technique for the source position of an afterloading system using an electronic two-dimensional detector array (2D array). A "GammaMed plus IX" high dose rate afterloading device (Varian Medical Systems, Palo Alto, USA), and a Seven29 2D detector array (PTW Freiburg GmbH, Freiburg, Germany) have been used. A hollow needle has been connected to the afterloading device. Its outer diameter is 3.0 mm and its length about 220mm. A 14 mm thick Perspex slab was fixed in a reproducible position on the 2D array. Above the 14th detector row, a groove was cut in the slab which accommodates the hollow needle in a fixed position relative to the 2D array. In order to define the position of the source, the signals of the detectors in the 14th detector row have been evaluated. A theoretical curve depending on the source position and strength, has been fitted to the acquired detector signals. It has been shown that the reproducibility of the measurement technique - including the reproducibility of the source placement - was within 0.15mm. This method can not only replace film measurements, it is also more exact and less time consuming.