Efektifitas Penambahan Source to Image Distance (SID) terhadap Penurunan Dosis Radiasi pada Pemeriksaan Radiografi Cranium

{"title":"Efektifitas Penambahan Source to Image Distance (SID) terhadap Penurunan Dosis Radiasi pada Pemeriksaan Radiografi Cranium","authors":"Angga Yosainto Bequet, Panji Wibowo Nurcahyo, A. H. Sulistiyadi","doi":"10.31983/jimed.v8i1.8043","DOIUrl":null,"url":null,"abstract":"Background: Cranial radiographs are routinely made in 2 projections, namely Antero Posterior (AP) and Lateral with a minimum distance of 100 cm from the radiation source to the image receptor. In addition, it has been shown that increasing the SID from 40 to 48 inches reduces the skin dose even when the requirement for an increase in mAs is considered. A 44-inch or 48-inch SID is recommended where departmental equipment and protocols allow. This study aims to determine the effectiveness of the addition of SID to decrease the radiation dose on cranium radiographs.Methods: This type of research is a pre-experimental quantitative. The study was conducted at the radiology laboratory of Diploma Three Program of Radiology Purwokerto, Semarang Health Politechnic of Ministry of Health. The research subject is a Radiographic Xray Cranial Phantom. At the time of exposure, the radiation dose was measured and the resulting image was measured CNR value.Results: The results of radiation dose measurements at each SID setting for AP cranium radiographs are shown in the table above. setting SID 100 cm produces a radiation dose of 5.58 mSv, SID 110 cm produces a radiation dose of 5.16 mSv, SID 120 cm produces a radiation dose of 5.02 mSv, SID 130 cm produces a radiation dose of 4.84 mSv, SID 140 cm produces a radiation dose of 4.64 mSv and SID 150 cm produces a radiation dose of 4.36 mSv. There are differences in radiation dose values between SID settings of 100 cm, 110 cm, 120 cm, 130 cm, 140 cm and 150 cm with a statistical test p-value 0.001. The results of statistical tests on the Contrast to Noise Ratio value on changes in the SID value of the AP cranium radiography examination showed no significant difference. The p-value of the statistical test is 0.274 which means there is no difference in the CNR value between the SID settings of 100 cm, 110 cm, 120 cm, 130 cm, 140 cm and 150 cm.Conclusions: An increase in SID causes a decrease in radiation dose. The increase in SID did not cause a significant difference to the quality of the Contrast to Noise Ratio on the AP cranium radiograph image.","PeriodicalId":205622,"journal":{"name":"Jurnal Imejing Diagnostik (JImeD)","volume":"20 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-02-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jurnal Imejing Diagnostik (JImeD)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31983/jimed.v8i1.8043","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Cranial radiographs are routinely made in 2 projections, namely Antero Posterior (AP) and Lateral with a minimum distance of 100 cm from the radiation source to the image receptor. In addition, it has been shown that increasing the SID from 40 to 48 inches reduces the skin dose even when the requirement for an increase in mAs is considered. A 44-inch or 48-inch SID is recommended where departmental equipment and protocols allow. This study aims to determine the effectiveness of the addition of SID to decrease the radiation dose on cranium radiographs.Methods: This type of research is a pre-experimental quantitative. The study was conducted at the radiology laboratory of Diploma Three Program of Radiology Purwokerto, Semarang Health Politechnic of Ministry of Health. The research subject is a Radiographic Xray Cranial Phantom. At the time of exposure, the radiation dose was measured and the resulting image was measured CNR value.Results: The results of radiation dose measurements at each SID setting for AP cranium radiographs are shown in the table above. setting SID 100 cm produces a radiation dose of 5.58 mSv, SID 110 cm produces a radiation dose of 5.16 mSv, SID 120 cm produces a radiation dose of 5.02 mSv, SID 130 cm produces a radiation dose of 4.84 mSv, SID 140 cm produces a radiation dose of 4.64 mSv and SID 150 cm produces a radiation dose of 4.36 mSv. There are differences in radiation dose values between SID settings of 100 cm, 110 cm, 120 cm, 130 cm, 140 cm and 150 cm with a statistical test p-value 0.001. The results of statistical tests on the Contrast to Noise Ratio value on changes in the SID value of the AP cranium radiography examination showed no significant difference. The p-value of the statistical test is 0.274 which means there is no difference in the CNR value between the SID settings of 100 cm, 110 cm, 120 cm, 130 cm, 140 cm and 150 cm.Conclusions: An increase in SID causes a decrease in radiation dose. The increase in SID did not cause a significant difference to the quality of the Contrast to Noise Ratio on the AP cranium radiograph image.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在颅骨x光检查中,向面散射的有效性增加
背景:头颅x线片通常为前后侧(AP)和侧位,从辐射源到图像受体的最小距离为100cm。此外,研究表明,即使考虑到增加mAs的要求,将SID从40英寸增加到48英寸也会减少皮肤剂量。如果部门设备和协议允许,建议使用44英寸或48英寸的SID。本研究旨在确定添加SID降低颅骨x线片辐射剂量的有效性。方法:本研究采用实验前定量法。这项研究是在卫生部三宝垄卫生技术学院放射学三级课程放射学实验室进行的。本研究对象为x线影像颅影。在照射时,测量辐射剂量,得到的图像是测量CNR值。结果:AP头颅x线片在不同SID设置下的辐射剂量测量结果见上表。设置SID 100 cm产生的辐射剂量为5.58 mSv, SID 110 cm产生的辐射剂量为5.16 mSv, SID 120 cm产生的辐射剂量为5.02 mSv, SID 130 cm产生的辐射剂量为4.84 mSv, SID 140 cm产生的辐射剂量为4.64 mSv, SID 150 cm产生的辐射剂量为4.36 mSv。在100 cm、110 cm、120 cm、130 cm、140 cm和150 cm的SID设置中,辐射剂量值存在差异,p值为0.001。对比噪声比(Contrast to Noise Ratio)值对AP头颅x线检查SID值变化的统计检验结果无显著差异。统计检验的p值为0.274,即100 cm、110 cm、120 cm、130 cm、140 cm和150 cm的SID设置之间的CNR值无差异。结论:SID升高引起辐射剂量降低。SID的增加并未对AP颅骨x线片图像的噪比质量造成显著影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Evaluasi Pemeriksaan MRI Brain dengan Klinis Vertigo di Rumah Sakit Otak dr. Drs. M. Hatta Bukittinggi Implementasi Pemanfaatan Sumber Radiasi Pengion pada Ruang CT Scan ditinjau berdasarkan Peraturan BAPETEN No. 4 tahun 2020 di Instalasi Radiologi Rumah Sakit type B di Wilayah Kabupaten Sumedang Nilai Entrance Skin Dose Tiroid antara menggunakan Thyroid Shield dan tanpa Thyroid Shield pada Pemeriksaan CT Scan Kepala Analisis Informasi Anatomi Pemeriksaan MRI Ankle Joint pada Penggunaan Foot Ankle Coil dan Flex Coil Proton Density Fat Saturation Irisan Sagital Evaluasi Kualitas Radiograf Periapikal Teknik Bisektris : Kesalahan Penempatan Sudut Penyinaran dan Film
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1