Distortion verification of helical computed tomography for image-guided radiotherapy

IF 0.3 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Journal of Radiotherapy in Practice Pub Date : 2023-01-01 DOI:10.1017/s1460396923000316
Takayuki Harada, Akihiro Takemura
{"title":"Distortion verification of helical computed tomography for image-guided radiotherapy","authors":"Takayuki Harada, Akihiro Takemura","doi":"10.1017/s1460396923000316","DOIUrl":null,"url":null,"abstract":"Abstract Introduction: In image-guided radiotherapy (IGRT), the imaging conditions of computed tomography (CT) may impact the positioning uncertainty, but verification methods are currently unavailable. This study aimed to propose a validation method for the imaging conditions of helical CT for IGRT. Predicting the impact of image distortion on image guidance may reduce uncertainty in radiotherapy planning. Methods: Image guidance was performed on the reference images of four Duracon balls by changing the imaging conditions and the positions on the CT images by helical scanning. The predictors of image guidance error and those of the contour mismatch between the reference and cone-beam CT (CBCT) images were analysed. Results: The image guidance error exceeded 1 mm when the contour centre of the ball was shifted by more than 1 mm. The mismatch between the contours of the reference and CBCT images occurred with the imaging conditions wherein the first slice of the ball was distorted. Conclusions: Mismatch can be predicted by the coefficient of variation of the radii in the first and centre slices of the ball. Moreover, the image guidance error can be predicted by the contour centre shift of the ball.","PeriodicalId":44597,"journal":{"name":"Journal of Radiotherapy in Practice","volume":null,"pages":null},"PeriodicalIF":0.3000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiotherapy in Practice","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1017/s1460396923000316","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Abstract Introduction: In image-guided radiotherapy (IGRT), the imaging conditions of computed tomography (CT) may impact the positioning uncertainty, but verification methods are currently unavailable. This study aimed to propose a validation method for the imaging conditions of helical CT for IGRT. Predicting the impact of image distortion on image guidance may reduce uncertainty in radiotherapy planning. Methods: Image guidance was performed on the reference images of four Duracon balls by changing the imaging conditions and the positions on the CT images by helical scanning. The predictors of image guidance error and those of the contour mismatch between the reference and cone-beam CT (CBCT) images were analysed. Results: The image guidance error exceeded 1 mm when the contour centre of the ball was shifted by more than 1 mm. The mismatch between the contours of the reference and CBCT images occurred with the imaging conditions wherein the first slice of the ball was distorted. Conclusions: Mismatch can be predicted by the coefficient of variation of the radii in the first and centre slices of the ball. Moreover, the image guidance error can be predicted by the contour centre shift of the ball.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
图像引导放射治疗的螺旋计算机断层畸变验证
摘要简介:在图像引导放射治疗(IGRT)中,计算机断层扫描(CT)的成像条件可能会影响定位不确定性,但目前尚无验证方法。本研究旨在提出一种螺旋CT用于IGRT成像条件的验证方法。预测图像畸变对图像引导的影响可以减少放疗计划的不确定性。方法:通过改变成像条件和螺旋扫描CT图像上的位置,对4颗Duracon球的参考图像进行图像引导。分析了图像制导误差和参考图像与锥束CT (CBCT)图像轮廓不匹配的预测因子。结果:当球的轮廓中心移动超过1mm时,图像引导误差超过1mm。在第一层球被扭曲的成像条件下,参考图像和CBCT图像的轮廓不匹配。结论:失配可以通过球的第一和中心片半径的变异系数来预测。此外,还可以利用球的轮廓中心位移来预测图像制导误差。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Radiotherapy in Practice
Journal of Radiotherapy in Practice RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
0.80
自引率
0.00%
发文量
36
期刊介绍: Journal of Radiotherapy in Practice is a peer-reviewed journal covering all of the current modalities specific to clinical oncology and radiotherapy. The journal aims to publish research from a wide range of styles and encourage debate and the exchange of information and opinion from within the field of radiotherapy practice and clinical oncology. The journal also aims to encourage technical evaluations and case studies as well as equipment reviews that will be of interest to an international radiotherapy audience.
期刊最新文献
Wobbling nature of gamma passing rate as a function of calibration field sizes in patient-specific quality assurance Secondary fragmentation and relative biological effectiveness (RBE) study using Bridge SOI microdosimeter: Monte Carlo simulation Keratin-based topical cream for radiation dermatitis during head and neck radiotherapy: a randomised, open-label pilot study. Single catheter 3d volume based hybrid inverse planning optimization in IVBT can improve organ sparing – CORRIGENDUM Dosimetric case study of 3-D FiF vs. VMAT techniques in the treatment of H/N tumour
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1