[Performance evaluation of the X-ray computed tomography "TCT-700 S" (Part 1)].

Shika gakuho. Dental science reports Pub Date : 1989-05-01
K Yamamoto, K Nishikawa, M Wakoh, Y Hayakawa, K Enomoto, K Tamura, C Hagiwara, H Fujimori, Y Kohsuge, N Kobayashi
{"title":"[Performance evaluation of the X-ray computed tomography \"TCT-700 S\" (Part 1)].","authors":"K Yamamoto,&nbsp;K Nishikawa,&nbsp;M Wakoh,&nbsp;Y Hayakawa,&nbsp;K Enomoto,&nbsp;K Tamura,&nbsp;C Hagiwara,&nbsp;H Fujimori,&nbsp;Y Kohsuge,&nbsp;N Kobayashi","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The performance of the Rotate/Rotate CT scanner \"TCT-700 S\" (Toshiba) was evaluated on the basis of the item entitled \"User's initial acceptance tests (phase II),\" set forth in \"Standard of performance evaluation for X-ray computed tomography (second recommendation)\" proposed by Takenaka et al. (1982). This report describes measurements for noise, contrast scale, spatial resolution, and high- and low-contrast resolution. Recommended phantoms were used to test performance. Results 1. Noise edepended on slice thickness and mAs. It increased as slice thickness decreased. No matter what the slice thickness, noise was great when mAs settings were small. Furthermore, short scan time caused considerable noise for each slice thickness. Reduction of mAs setting caused remarkable increased in noise when slice thickness were 1 mm or 2 mm Under routine examination, when thickness was 5 mm, at 120 kV, with an mAs setting of 440, scan field S(diameter 240 mm), and convolution filter FC-2, noise was 0.31%. When slice thickness was 10 mm at 120kv, 800mAs, with a scan field S (diameter 240 mm), and convolution filter FC-2, contrast scale was 1.84 x 10(-4) (cm-1/CT number). 2. When slice thickness was 5 mm, at 120 kV, with a mAs setting of 1200, scan field S(diameter 240 mm), and a convolution filter of FC-2, spatial resolution was 0.5LP/mm. 3. When slice thickness was 5 mm, at 120kV, with a mAs setting of 1200, scan field S (diameter 240 mm), and convolution filter of FC-2, minimum high-contrast resolution was 0.5 mm.(ABSTRACT TRUNCATED AT 250 WORDS)</p>","PeriodicalId":76540,"journal":{"name":"Shika gakuho. Dental science reports","volume":"89 5","pages":"1031-8"},"PeriodicalIF":0.0000,"publicationDate":"1989-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Shika gakuho. Dental science reports","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The performance of the Rotate/Rotate CT scanner "TCT-700 S" (Toshiba) was evaluated on the basis of the item entitled "User's initial acceptance tests (phase II)," set forth in "Standard of performance evaluation for X-ray computed tomography (second recommendation)" proposed by Takenaka et al. (1982). This report describes measurements for noise, contrast scale, spatial resolution, and high- and low-contrast resolution. Recommended phantoms were used to test performance. Results 1. Noise edepended on slice thickness and mAs. It increased as slice thickness decreased. No matter what the slice thickness, noise was great when mAs settings were small. Furthermore, short scan time caused considerable noise for each slice thickness. Reduction of mAs setting caused remarkable increased in noise when slice thickness were 1 mm or 2 mm Under routine examination, when thickness was 5 mm, at 120 kV, with an mAs setting of 440, scan field S(diameter 240 mm), and convolution filter FC-2, noise was 0.31%. When slice thickness was 10 mm at 120kv, 800mAs, with a scan field S (diameter 240 mm), and convolution filter FC-2, contrast scale was 1.84 x 10(-4) (cm-1/CT number). 2. When slice thickness was 5 mm, at 120 kV, with a mAs setting of 1200, scan field S(diameter 240 mm), and a convolution filter of FC-2, spatial resolution was 0.5LP/mm. 3. When slice thickness was 5 mm, at 120kV, with a mAs setting of 1200, scan field S (diameter 240 mm), and convolution filter of FC-2, minimum high-contrast resolution was 0.5 mm.(ABSTRACT TRUNCATED AT 250 WORDS)

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[x射线计算机断层扫描“tct - 700s”的性能评估(第一部分)]。
旋转/旋转CT扫描仪“tct - 700s”(东芝)的性能根据Takenaka等人(1982年)提出的“x射线计算机断层扫描性能评估标准(第二次建议)”中提出的题为“用户初始验收测试(第二阶段)”的项目进行了评估。本报告描述了噪声、对比度尺度、空间分辨率以及高低对比度分辨率的测量方法。使用推荐的幻影来测试性能。结果1。噪声取决于切片厚度和mAs。随切片厚度的减小而增大。无论切片厚度如何,当mAs设置较小时,噪声都很大。此外,较短的扫描时间对每个切片厚度造成了相当大的噪声。在常规检查中,当切片厚度为1 mm或2 mm时,厚度为5 mm, 120 kV, mAs设置为440,扫描场S(直径240 mm),卷积滤波器FC-2,噪声为0.31%。在120kv、800ma、扫描场S(直径240 mm)、卷积滤波器FC-2下,切片厚度为10 mm时,对比尺度为1.84 × 10(-4) (cm-1/CT数)。2. 当层厚为5 mm,电压为120 kV, mAs设置为1200,扫描场S(直径240 mm),卷积滤波器FC-2时,空间分辨率为0.5LP/mm。3.当切片厚度为5 mm,电压为120kV, mAs设置为1200,扫描场S(直径240 mm),卷积滤波器为FC-2时,最小高对比度分辨率为0.5 mm。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
[Study of structural changes in canine maxillary trabecular bone after tooth extraction]. [A study of the internal structure of the Japanese edentulous mandible]. [Oriental psychosomatic medical therapy for vertigo]. [Experimental analytical electron microscopic studies on the quantitative analysis of elemental concentrations in biological thin specimens and its application to dental science]. [Experimental studies on healing process of plication technique at postdiscal connective tissue of TMJ in Macaca fuscata].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1