Adding an extension piece to the end of the purge side arm of the Impella device can prevent the incidence of the cassette breaking and decrease the Impella device failure rate: Impact of practice change on patient outcome.

IF 1.1 4区 医学 Q4 CARDIAC & CARDIOVASCULAR SYSTEMS Perfusion-Uk Pub Date : 2024-11-01 Epub Date: 2023-12-07 DOI:10.1177/02676591231220305
Parthkumar Satashia, Andrew White, Shahin Isha, Abby Hanson, Anna Jenkins, Jessica Blasavage, Nikki Matos, Amanda Tomlinson, Stephanie Zhang, Quintin Quinones, Nathan Waldron, Anirban Bhattacharyya, Sean Kiley, Pramod Guru, Sanjay Chaudhury, Anna Shapiro, Pablo Moreno Franco, Devang K Sanghavi
{"title":"Adding an extension piece to the end of the purge side arm of the Impella device can prevent the incidence of the cassette breaking and decrease the Impella device failure rate: Impact of practice change on patient outcome.","authors":"Parthkumar Satashia, Andrew White, Shahin Isha, Abby Hanson, Anna Jenkins, Jessica Blasavage, Nikki Matos, Amanda Tomlinson, Stephanie Zhang, Quintin Quinones, Nathan Waldron, Anirban Bhattacharyya, Sean Kiley, Pramod Guru, Sanjay Chaudhury, Anna Shapiro, Pablo Moreno Franco, Devang K Sanghavi","doi":"10.1177/02676591231220305","DOIUrl":null,"url":null,"abstract":"<p><p><i>Background:</i> Impella 5.5® with Smart Assist is a minimally invasive Left Ventricular Assist Devices (LVAD) approved by the Food and Drug Administration (FDA) for treating ongoing cardiogenic shock for up to 14 days. The Impella® intends to reduce ventricular workload and provide the circulatory support necessary for myocardial recovery.<i>Research Question:</i> Compared to standard practice, does adding an extension piece to the purge tube side arm of the Impella® Device decrease the incidence of device failure and positively impact the health outcome of adult patients receiving Impella® support?<i>Study Design and Methods:</i> A retrospective chart review of ICU patients was done at a tertiary care center from August 2018 to August 2022 to assess the differences in patient outcomes related to Impella® Device utilization before and after the implementation of the extension piece to the purge tube sidearm. Among patients reviewed, a total of 20 were included in our review, with seven not having the purge tube side arm extension added, while 13 patients had the extension.<i>Results:</i> The two study groups had no significant difference in patient health outcomes. Additionally, there were no instances of device failure requiring explanation without the extension tubing. However, there were no cases of the purge cassette cracking with the addition of the extension tubing.<i>Conclusion:</i> The addition of extension tubing to the purge cassette of the Impella® Device did not impact patient health outcomes or the incidence of device failure. There was a complete reduction in the incidence of the purge cassette cracking, which could reduce the potential for infection or device failure over a long period of mechanical support. There is a need for long-term prospective studies to confirm the results.</p>","PeriodicalId":49707,"journal":{"name":"Perfusion-Uk","volume":" ","pages":"1757-1760"},"PeriodicalIF":1.1000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Perfusion-Uk","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/02676591231220305","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/7 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Impella 5.5® with Smart Assist is a minimally invasive Left Ventricular Assist Devices (LVAD) approved by the Food and Drug Administration (FDA) for treating ongoing cardiogenic shock for up to 14 days. The Impella® intends to reduce ventricular workload and provide the circulatory support necessary for myocardial recovery.Research Question: Compared to standard practice, does adding an extension piece to the purge tube side arm of the Impella® Device decrease the incidence of device failure and positively impact the health outcome of adult patients receiving Impella® support?Study Design and Methods: A retrospective chart review of ICU patients was done at a tertiary care center from August 2018 to August 2022 to assess the differences in patient outcomes related to Impella® Device utilization before and after the implementation of the extension piece to the purge tube sidearm. Among patients reviewed, a total of 20 were included in our review, with seven not having the purge tube side arm extension added, while 13 patients had the extension.Results: The two study groups had no significant difference in patient health outcomes. Additionally, there were no instances of device failure requiring explanation without the extension tubing. However, there were no cases of the purge cassette cracking with the addition of the extension tubing.Conclusion: The addition of extension tubing to the purge cassette of the Impella® Device did not impact patient health outcomes or the incidence of device failure. There was a complete reduction in the incidence of the purge cassette cracking, which could reduce the potential for infection or device failure over a long period of mechanical support. There is a need for long-term prospective studies to confirm the results.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在Impella设备的吹吹侧臂末端增加一个延长片,可以防止卡壳断裂的发生,降低Impella设备的故障率:实践改变对患者预后的影响。
背景:Impella 5.5®与Smart Assist是一种微创左心室辅助装置(LVAD),经美国食品和药物管理局(FDA)批准,用于治疗持续14天的心源性休克。Impella®旨在减少心室负荷,并为心肌恢复提供必要的循环支持。研究问题:与标准实践相比,在Impella®设备的吹吹管侧臂上添加延长片是否可以降低设备故障的发生率,并对接受Impella®支持的成年患者的健康结果产生积极影响?研究设计与方法:2018年8月至2022年8月,对某三级医疗中心ICU患者进行回顾性图表分析,以评估在吹吹管侧臂安装延伸片前后患者使用Impella®设备的结果差异。在我们回顾的患者中,共有20例纳入我们的回顾,其中7例未添加吹扫管侧臂延长,而13例添加了延长。结果:两组患者健康状况无显著差异。此外,在没有延长油管的情况下,也没有设备故障需要解释的情况。然而,没有清洗盒开裂的情况下,增加了延长管。结论:在Impella®设备的清除盒中添加延长管不会影响患者的健康结果或设备故障的发生率。清洗盒破裂的发生率完全降低,这可以减少感染或设备在长时间机械支持期间故障的可能性。需要进行长期的前瞻性研究来证实这些结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Perfusion-Uk
Perfusion-Uk 医学-外周血管病
CiteScore
3.00
自引率
8.30%
发文量
203
审稿时长
6-12 weeks
期刊介绍: Perfusion is an ISI-ranked, peer-reviewed scholarly journal, which provides current information on all aspects of perfusion, oxygenation and biocompatibility and their use in modern cardiac surgery. The journal is at the forefront of international research and development and presents an appropriately multidisciplinary approach to perfusion science.
期刊最新文献
Ensuring quality information for patients tool to assess patient information on CABG websites: Systemic search and evaluation. Hemoadsorption during cardiopulmonary bypass to absorb plasma-free hemoglobin in patients with acute type a aortic dissection: A randomized controlled trial. Letter re: The COMICS trial: Randomization to MiECC significantly decreases serious adverse events. MiECC reloaded. Non-invasive cerebral autoregulation monitoring during paediatric cardiac surgery without cardiopulmonary bypass requiring intraoperative cross-clamping of one of the carotid arteries.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1