Correlation between the electrocardiogram amplitude detected by an implantable cardiac monitor and the implantation depth

IF 1.1 4区 医学 Q4 CARDIAC & CARDIOVASCULAR SYSTEMS Annals of Noninvasive Electrocardiology Pub Date : 2023-12-13 DOI:10.1111/anec.13102
Yohei Kawatani MD, PhD, LLB, Takaki Hori MD, PhD
{"title":"Correlation between the electrocardiogram amplitude detected by an implantable cardiac monitor and the implantation depth","authors":"Yohei Kawatani MD, PhD, LLB,&nbsp;Takaki Hori MD, PhD","doi":"10.1111/anec.13102","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>Implantable cardiac monitors (ICMs) primarily use R-R intervals in subcutaneous electrocardiograms (ECGs) to detect arrhythmias. Therefore, reliable detection of R-wave amplitude by an ICM is vital. Since ICMs detect subcutaneous ECGs, the impact of the implantation depth should be assessed.</p>\n </section>\n \n <section>\n \n <h3> Methods and Results</h3>\n \n <p>This study investigated the influence of ICM depth on R-wave (ICM-R) amplitude on an ECG generated by an ICM (JOT Dx; Abbott). Overall, 58 patients who underwent ICM implantation at Kamagaya General Hospital from May 2022 to April 2023 were retrospectively reviewed. The depth-position was measured using ultrasound imaging after implantation. The depth of the ICM did not show any correlation with ICM-R amplitude (<i>r</i> = −.0141, <i>p</i> = .294). However, the distance between the ICM and the heart surface showed a significant correlation with ICM-R amplitude (<i>r</i> = −.581, <i>p</i> &lt; .001). Body weight (<i>r</i> = −.0283, <i>p</i> = .033) and body mass index (<i>r</i> = −.0342, <i>p</i> = .009) were associated with ICM-R amplitude. S wave in the V<sub>1</sub>-lead was also associated with ICM-R amplitude (<i>r</i> = .481, <i>p</i> &lt; .001). After multivariate analysis, the distance between the ICM and heart surface and the S wave in V<sub>1</sub> were independent determinants for the ICM-R amplitude.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>The ICM-R amplitude may be higher with the ICM implanted deeper.</p>\n </section>\n </div>","PeriodicalId":8074,"journal":{"name":"Annals of Noninvasive Electrocardiology","volume":"29 1","pages":""},"PeriodicalIF":1.1000,"publicationDate":"2023-12-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/anec.13102","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Noninvasive Electrocardiology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/anec.13102","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction

Implantable cardiac monitors (ICMs) primarily use R-R intervals in subcutaneous electrocardiograms (ECGs) to detect arrhythmias. Therefore, reliable detection of R-wave amplitude by an ICM is vital. Since ICMs detect subcutaneous ECGs, the impact of the implantation depth should be assessed.

Methods and Results

This study investigated the influence of ICM depth on R-wave (ICM-R) amplitude on an ECG generated by an ICM (JOT Dx; Abbott). Overall, 58 patients who underwent ICM implantation at Kamagaya General Hospital from May 2022 to April 2023 were retrospectively reviewed. The depth-position was measured using ultrasound imaging after implantation. The depth of the ICM did not show any correlation with ICM-R amplitude (r = −.0141, p = .294). However, the distance between the ICM and the heart surface showed a significant correlation with ICM-R amplitude (r = −.581, p < .001). Body weight (r = −.0283, p = .033) and body mass index (r = −.0342, p = .009) were associated with ICM-R amplitude. S wave in the V1-lead was also associated with ICM-R amplitude (r = .481, p < .001). After multivariate analysis, the distance between the ICM and heart surface and the S wave in V1 were independent determinants for the ICM-R amplitude.

Conclusion

The ICM-R amplitude may be higher with the ICM implanted deeper.

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
植入式心脏监测仪检测到的心电图振幅与植入深度之间的相关性
植入式心脏监护仪(ICMs)主要使用皮下心电图(ECGs)的R-R间隔来检测心律失常。因此,通过ICM可靠地检测r波振幅是至关重要的。由于ICMs检测皮下心电图,应评估植入深度的影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.40
自引率
0.00%
发文量
88
审稿时长
6-12 weeks
期刊介绍: The ANNALS OF NONINVASIVE ELECTROCARDIOLOGY (A.N.E) is an online only journal that incorporates ongoing advances in the clinical application and technology of traditional and new ECG-based techniques in the diagnosis and treatment of cardiac patients. ANE is the first journal in an evolving subspecialty that incorporates ongoing advances in the clinical application and technology of traditional and new ECG-based techniques in the diagnosis and treatment of cardiac patients. The publication includes topics related to 12-lead, exercise and high-resolution electrocardiography, arrhythmias, ischemia, repolarization phenomena, heart rate variability, circadian rhythms, bioengineering technology, signal-averaged ECGs, T-wave alternans and automatic external defibrillation. ANE publishes peer-reviewed articles of interest to clinicians and researchers in the field of noninvasive electrocardiology. Original research, clinical studies, state-of-the-art reviews, case reports, technical notes, and letters to the editors will be published to meet future demands in this field.
期刊最新文献
Association Between Different Insulin Resistance Indices and Heart Failure in US Adults With Diabetes Mellitus Coronary Angiographic Features of de Winter Syndrome: More Than Just Occlusion of the Left Anterior Descending Artery Intracardiac Echocardiography and Ablation of Atrial Fibrillation in Dextrocardia Introducing a Novel Pacemaker-Mediated Arrhythmia: The Pseudo-RNRVAS Arising From Atrial Capture Challenges. Catheter Ablation of Ventricular Premature Contractions Originating From RVOT With Interruption of the Inferior Vena Cava-A Case Report.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1