Evaluation of 278 New Bluephase Style Light Curing Units Purchased Over Six Years.

IF 1.1 Q3 DENTISTRY, ORAL SURGERY & MEDICINE European Journal of Prosthodontics and Restorative Dentistry Pub Date : 2024-05-31 DOI:10.1922/EJPRD_2605Bila07
B Bila, C Maucoski, R Price, B Sullivan
{"title":"Evaluation of 278 New Bluephase Style Light Curing Units Purchased Over Six Years.","authors":"B Bila, C Maucoski, R Price, B Sullivan","doi":"10.1922/EJPRD_2605Bila07","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>This study measured 278 new curing lights (LCUs) to determine if they met the manufacturer's specifications.</p><p><strong>Materials and methods: </strong>Over a 6-year period, the light output from 278 Bluephase Style LCUs was measured before the LCUs were given to students (41 in 2017, 44 in 2018, 46 in 2019, 48 in 2020, 50 in 2021, and 49 in 2022). The radiant exitance (mW/cm²) from each year was compared using ANOVA and Tukey- Kramer multiple comparisons test with adjustment for the unequal sample sizes (α=0.05). The results were also compared against the manufacturer's stated radiant exitance of 1200 mW/cm².</p><p><strong>Results: </strong>The 278 LCUs delivered an average (± standard deviation) radiant exitance of 1203 ± 16.7 mW/cm². The lowest radiant exitance was 1149 mW/cm² and the highest was 1257 mW/cm². Although there were statistically significant differences in the LCUs from each year, these values were within the manufacturer's tolerance of values of ± 10%. Approximately 13% of the emitted light was below 420 nm, and 87% was above 420 nm.</p><p><strong>Conclusion: </strong>These new LCUs were all multiple peak LCUs that emitted an average radiant exitance of 1,203 mW/cm² (range 1149 to 1257 mW/cm². Purchasers of the Style LCU can be confident that their new LCU will meet the manufacturer's specifications.</p>","PeriodicalId":45686,"journal":{"name":"European Journal of Prosthodontics and Restorative Dentistry","volume":" ","pages":"168-174"},"PeriodicalIF":1.1000,"publicationDate":"2024-05-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Prosthodontics and Restorative Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1922/EJPRD_2605Bila07","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: This study measured 278 new curing lights (LCUs) to determine if they met the manufacturer's specifications.

Materials and methods: Over a 6-year period, the light output from 278 Bluephase Style LCUs was measured before the LCUs were given to students (41 in 2017, 44 in 2018, 46 in 2019, 48 in 2020, 50 in 2021, and 49 in 2022). The radiant exitance (mW/cm²) from each year was compared using ANOVA and Tukey- Kramer multiple comparisons test with adjustment for the unequal sample sizes (α=0.05). The results were also compared against the manufacturer's stated radiant exitance of 1200 mW/cm².

Results: The 278 LCUs delivered an average (± standard deviation) radiant exitance of 1203 ± 16.7 mW/cm². The lowest radiant exitance was 1149 mW/cm² and the highest was 1257 mW/cm². Although there were statistically significant differences in the LCUs from each year, these values were within the manufacturer's tolerance of values of ± 10%. Approximately 13% of the emitted light was below 420 nm, and 87% was above 420 nm.

Conclusion: These new LCUs were all multiple peak LCUs that emitted an average radiant exitance of 1,203 mW/cm² (range 1149 to 1257 mW/cm². Purchasers of the Style LCU can be confident that their new LCU will meet the manufacturer's specifications.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
对六年来购买的 278 台新 Bluephase 型光固化设备进行评估。
目标:本研究测量了 278 个新的固化灯(LCU),以确定它们是否符合制造商的规格要求:在 6 年时间里,在将 LCU 分配给学生之前,对 278 台 Bluephase Style LCU 的光输出进行了测量(2017 年 41 台,2018 年 44 台,2019 年 46 台,2020 年 48 台,2021 年 50 台,2022 年 49 台)。使用方差分析和 Tukey- Kramer 多重比较检验对每年的辐射输出(毫瓦/平方厘米)进行了比较,并对不等的样本量进行了调整(α=0.05)。此外,还将结果与制造商规定的 1200 mW/cm² 辐射输出进行了比较:278 台 LCU 的平均(± 标准偏差)辐射输出为 1203 ± 16.7 mW/cm²。最低辐射输出为 1149 mW/cm²,最高为 1257 mW/cm²。虽然每年的 LCU 在统计上都有显著差异,但这些数值都在制造商规定的 ± 10% 的误差范围内。大约 13% 的发射光低于 420 纳米,87% 的发射光高于 420 纳米:这些新型 LCU 都是多峰值 LCU,平均辐射输出功率为 1,203 mW/cm²(范围为 1149 至 1257 mW/cm²)。Style LCU 的购买者可以放心,他们的新 LCU 将符合制造商的规格要求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
2.30
自引率
7.70%
发文量
0
期刊介绍: The European Journal of Prosthodontics and Restorative Dentistry is published quarterly and includes clinical and research articles in subjects such as prosthodontics, operative dentistry, implantology, endodontics, periodontics and dental materials.
期刊最新文献
Influence of Surface Roughness on the Adhesion of Hydroxyapatite Coatings to Titanium and Titanium Alloy Surfaces: A Systematic Review of in vitro Experimental Studies. Performance of Fiberglass Posts Versus Fiber- Reinforced Resin Composites in Endodontically Treated Anterior Teeth Without Ferrule: A Systematic Review. Assessment of Efficacy of Implant Screwdrivers' During Multiple Tightening/ Untightening Cycles. An in vitro Study. Hybrid Full-Arch Rehabilitation Using Conventional And Zygomatic Implants: A Short-Term Retrospective Analysis. Clinical Survival of Indirect, Posterior Fiber-Reinforced Composite Fixed Dental Prosthesis: Up to 15 Years of Prospective Clinical Follow Up.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1