Kyung-Soo Kim, Seung-Hwan Lee, Won Sang Yoo, Cheol-Young Park
{"title":"Accuracy and Safety of the 15-Day CareSens Air Continuous Glucose Monitoring System.","authors":"Kyung-Soo Kim, Seung-Hwan Lee, Won Sang Yoo, Cheol-Young Park","doi":"10.1089/dia.2023.0468","DOIUrl":null,"url":null,"abstract":"<p><p><b><i>Background:</i></b> We evaluated the accuracy and safety of the CareSens Air, a novel real-time continuous glucose monitoring system (CGMS), during 15 days of use in adults with diabetes. <b><i>Methods:</i></b> Adults with either type 1 diabetes or type 2 diabetes requiring intensive insulin therapy participated at four sites in South Korea. All participants wore the sensor for 15 days. Participants were scheduled for four 8-h clinic sessions on Day 1, 5 ± 1, 10 ± 1, and 15. Accuracy was evaluated based on the proportion of continuous glucose monitoring (CGM) values within 15% of YSI values ≥100 mg/dL or within 15 mg/dL of YSI values <100 mg/dL (%15/15), along with the %20/20, %30/30, and %40/40 agreement rates. The mean absolute relative difference (MARD) between the CGM and YSI values was calculated. <b><i>Results:</i></b> Data from 83 participants (83 sensors, 10,029 CGM-YSI matched pairs) were analyzed. The overall MARD was 10.42%, and the overall %15/15, %20/20, %30/30, and %40/40 accuracy were 78.55%, 89.04%, 96.47%, and 98.87%, respectively. The consensus error grid analysis showed that 99.92% of CGM values fell into Zone A or B (Zone A: 89.83%, Zone B: 10.09%). The %20/20 accuracy of CGMS was 88.11% on Day 1, 90.11% on Day 3-5, 92.09% on Day 8-10, and 85.63% on Day 15. No serious adverse events were reported. <b><i>Conclusions:</i></b> The CareSens Air demonstrated accurate performance across the wide glycemic range and was well tolerated during the 15-day sensor use period.</p>","PeriodicalId":11159,"journal":{"name":"Diabetes technology & therapeutics","volume":null,"pages":null},"PeriodicalIF":5.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10979678/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Diabetes technology & therapeutics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1089/dia.2023.0468","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/13 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
Abstract
Background: We evaluated the accuracy and safety of the CareSens Air, a novel real-time continuous glucose monitoring system (CGMS), during 15 days of use in adults with diabetes. Methods: Adults with either type 1 diabetes or type 2 diabetes requiring intensive insulin therapy participated at four sites in South Korea. All participants wore the sensor for 15 days. Participants were scheduled for four 8-h clinic sessions on Day 1, 5 ± 1, 10 ± 1, and 15. Accuracy was evaluated based on the proportion of continuous glucose monitoring (CGM) values within 15% of YSI values ≥100 mg/dL or within 15 mg/dL of YSI values <100 mg/dL (%15/15), along with the %20/20, %30/30, and %40/40 agreement rates. The mean absolute relative difference (MARD) between the CGM and YSI values was calculated. Results: Data from 83 participants (83 sensors, 10,029 CGM-YSI matched pairs) were analyzed. The overall MARD was 10.42%, and the overall %15/15, %20/20, %30/30, and %40/40 accuracy were 78.55%, 89.04%, 96.47%, and 98.87%, respectively. The consensus error grid analysis showed that 99.92% of CGM values fell into Zone A or B (Zone A: 89.83%, Zone B: 10.09%). The %20/20 accuracy of CGMS was 88.11% on Day 1, 90.11% on Day 3-5, 92.09% on Day 8-10, and 85.63% on Day 15. No serious adverse events were reported. Conclusions: The CareSens Air demonstrated accurate performance across the wide glycemic range and was well tolerated during the 15-day sensor use period.
期刊介绍:
Diabetes Technology & Therapeutics is the only peer-reviewed journal providing healthcare professionals with information on new devices, drugs, drug delivery systems, and software for managing patients with diabetes. This leading international journal delivers practical information and comprehensive coverage of cutting-edge technologies and therapeutics in the field, and each issue highlights new pharmacological and device developments to optimize patient care.