Andreas Goldschmied, Juergen Schreieck, Michal Droppa
{"title":"Pacemaker Lead Entanglement during Interventional PFO Occlusion: Salvage Using a Sizing Balloon.","authors":"Andreas Goldschmied, Juergen Schreieck, Michal Droppa","doi":"10.1155/2023/5586197","DOIUrl":null,"url":null,"abstract":"<p><p>We present a case of a patient with a transient ischaemic attack (TIA) likely due to paradoxical embolism through a patent foramen ovale (PFO). Her medical history included 2<sup>nd</sup>-degree heart block Mobitz II, which manifested with recurrent syncopes and was treated with a dual chamber pacemaker. During the interventional PFO closure procedure, we noted entrapment of the atrial pacemaker lead between the right-sided occluder disc and the interatrial septum. We were able to successfully move the lead aside using a 24 mm sizing balloon and subsequently developed the right-sided occluder disc in the correct position. In conclusion, pacemaker-lead entrapment between a PFO occluder disc and the interatrial septum can be prevented using a sizing balloon.</p>","PeriodicalId":51760,"journal":{"name":"Case Reports in Cardiology","volume":"2023 ","pages":"5586197"},"PeriodicalIF":0.6000,"publicationDate":"2023-12-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10733004/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Case Reports in Cardiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2023/5586197","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
We present a case of a patient with a transient ischaemic attack (TIA) likely due to paradoxical embolism through a patent foramen ovale (PFO). Her medical history included 2nd-degree heart block Mobitz II, which manifested with recurrent syncopes and was treated with a dual chamber pacemaker. During the interventional PFO closure procedure, we noted entrapment of the atrial pacemaker lead between the right-sided occluder disc and the interatrial septum. We were able to successfully move the lead aside using a 24 mm sizing balloon and subsequently developed the right-sided occluder disc in the correct position. In conclusion, pacemaker-lead entrapment between a PFO occluder disc and the interatrial septum can be prevented using a sizing balloon.
期刊介绍:
Case Reports in Cardiology is a peer-reviewed, Open Access journal that publishes case reports and case series related to hypertension, arrhythmia, congestive heart failure, valvular heart disease, vascular disease, congenital heart disease and cardiomyopathy.