Yi-Fan Wang, Yu Wu, Xiao-Wei Liu, Jian-Guo Li, Yan-Qiong Zhan, Bin Liu, Wen-Ling Fan, Zi-Heng Peng, Jin-Tao Xiao, Bing-Bing Li, Jian He, Jun Yi, Zhao-Xia Lu
{"title":"Effect of a disposable endoscope precleaning kit in the cleaning procedure of gastrointestinal endoscope: A multi-center observational study","authors":"Yi-Fan Wang, Yu Wu, Xiao-Wei Liu, Jian-Guo Li, Yan-Qiong Zhan, Bin Liu, Wen-Ling Fan, Zi-Heng Peng, Jin-Tao Xiao, Bing-Bing Li, Jian He, Jun Yi, Zhao-Xia Lu","doi":"10.4253/wjge.v15.i12.705","DOIUrl":null,"url":null,"abstract":"BACKGROUND\n Precleaning is a key step in endoscopic reprocessing.\n AIM\n To develop an effective and economic endoscope cleaning method by using a disposable endoscope bedside precleaning kit.\n METHODS\n Altogether, 228 used gastrointestinal endoscopes were selected from five high-volume endoscopy units and precleaned by a traditional precleaning bucket (group T) or a disposable endoscope bedside precleaning kit (group D). Each group was further subdivided based on the replacement frequency of the cleaning solution, which was replaced every time in subgroups T1 and D1 and every several times in subgroups Ts and Ds. The adenosine triphosphate (ATP) level and residual proteins were measured three times: Before and after precleaning and after manual cleaning.\n RESULTS\n After precleaning, the precleaning kit significantly reduced the ATP levels (P = 0.034) and has a more stable ATP clearance rate than the traditional precleaning bucket. The precleaning kit also saved a quarter of the cost of enzymatic detergent used during the precleaning process. After manual cleaning, the ATP levels were also significantly lower in the precleaning kit group than in the traditional precleaning bucket group (P < 0.05). Meanwhile, the number of uses of the cleaning solution (up to four times) has no significant impact on the cleaning effect (P > 0.05).\n CONCLUSION\n Considering its economic cost and cleaning effect, the use of a disposable endoscope bedside precleaning kit can be an optimal option in the precleaning stage with the cleaning solution being replaced several times in the manual cleaning stage.","PeriodicalId":23953,"journal":{"name":"World Journal of Gastrointestinal Endoscopy","volume":"35 6","pages":""},"PeriodicalIF":1.4000,"publicationDate":"2023-12-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"World Journal of Gastrointestinal Endoscopy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4253/wjge.v15.i12.705","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"GASTROENTEROLOGY & HEPATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
BACKGROUND
Precleaning is a key step in endoscopic reprocessing.
AIM
To develop an effective and economic endoscope cleaning method by using a disposable endoscope bedside precleaning kit.
METHODS
Altogether, 228 used gastrointestinal endoscopes were selected from five high-volume endoscopy units and precleaned by a traditional precleaning bucket (group T) or a disposable endoscope bedside precleaning kit (group D). Each group was further subdivided based on the replacement frequency of the cleaning solution, which was replaced every time in subgroups T1 and D1 and every several times in subgroups Ts and Ds. The adenosine triphosphate (ATP) level and residual proteins were measured three times: Before and after precleaning and after manual cleaning.
RESULTS
After precleaning, the precleaning kit significantly reduced the ATP levels (P = 0.034) and has a more stable ATP clearance rate than the traditional precleaning bucket. The precleaning kit also saved a quarter of the cost of enzymatic detergent used during the precleaning process. After manual cleaning, the ATP levels were also significantly lower in the precleaning kit group than in the traditional precleaning bucket group (P < 0.05). Meanwhile, the number of uses of the cleaning solution (up to four times) has no significant impact on the cleaning effect (P > 0.05).
CONCLUSION
Considering its economic cost and cleaning effect, the use of a disposable endoscope bedside precleaning kit can be an optimal option in the precleaning stage with the cleaning solution being replaced several times in the manual cleaning stage.