{"title":"Assessment of patient characteristics influencing the complexity of leadless pacemaker implantation","authors":"Hiroshi Miyama MD, PhD, Yukihiro Himeno MD, Shuhei Yano MD, Shuhei Yamashita MD, Koki Yamaoka MD, Susumu Ibe MD, Otoya Sekine MD, Yoshinori Katsumata MD, PhD, Takahiko Nishiyama MD, PhD, Takehiro Kimura MD, PhD, FHRS, Seiji Takatsuki MD, PhD, FHRS, Masaki Ieda MD, PhD","doi":"10.1016/j.hroo.2023.12.004","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>The complexity of leadless pacemaker (LP) implantation varies widely. However, the predictive factors determining this difficulty are poorly understood.</p></div><div><h3>Objective</h3><p>The purpose of this study was to evaluate the factors influencing LP implantation difficulty, specifically procedural time during right atrial (RA) and right ventricular (RV) manipulation, based on patient background, cardiac function, and anatomic characteristics.</p></div><div><h3>Methods</h3><p>Analysis included LP implantation cases between 2017 and 2023, excluding the initial 3 implants performed by each operator. The relevance of patient background, cardiac function, and anatomic features on procedural and fluoroscopy times was evaluated.</p></div><div><h3>Results</h3><p>Fifty-four patients (mean age 82.2 ± 10.0 years; 57.4% male) were included in the study. Median procedural and fluoroscopy time was 45.8 minutes and 16.0 minutes, respectively, with an average of 2.0 ± 1.4 device deployments. Univariate analysis showed associations between procedural time and older age, RA and RV diameter, and severity of tricuspid regurgitation (TR). After adjustment for physician and potential contributing factors, RV dilation (midventricular diameter ≥35 mm) and severe TR were identified as independent predictors of prolonged procedural time. Medical history exhibited no association with procedural time. Consistent results were observed in analyses using fluoroscopy time as the outcome.</p></div><div><h3>Conclusion</h3><p>RV dilation and severe TR were associated with prolonged procedural time for LP implantation. Anatomic features obtained from preprocedural echocardiography could provide valuable insights into both the safety and efficiency of LP implantation, thereby enhancing tailored treatment strategies for patients undergoing pacemaker implantation.</p></div>","PeriodicalId":29772,"journal":{"name":"Heart Rhythm O2","volume":"5 2","pages":"Pages 97-102"},"PeriodicalIF":2.5000,"publicationDate":"2024-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2666501823003343/pdfft?md5=c6e41691aff36599bac3f355451fa363&pid=1-s2.0-S2666501823003343-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Heart Rhythm O2","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2666501823003343","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
The complexity of leadless pacemaker (LP) implantation varies widely. However, the predictive factors determining this difficulty are poorly understood.
Objective
The purpose of this study was to evaluate the factors influencing LP implantation difficulty, specifically procedural time during right atrial (RA) and right ventricular (RV) manipulation, based on patient background, cardiac function, and anatomic characteristics.
Methods
Analysis included LP implantation cases between 2017 and 2023, excluding the initial 3 implants performed by each operator. The relevance of patient background, cardiac function, and anatomic features on procedural and fluoroscopy times was evaluated.
Results
Fifty-four patients (mean age 82.2 ± 10.0 years; 57.4% male) were included in the study. Median procedural and fluoroscopy time was 45.8 minutes and 16.0 minutes, respectively, with an average of 2.0 ± 1.4 device deployments. Univariate analysis showed associations between procedural time and older age, RA and RV diameter, and severity of tricuspid regurgitation (TR). After adjustment for physician and potential contributing factors, RV dilation (midventricular diameter ≥35 mm) and severe TR were identified as independent predictors of prolonged procedural time. Medical history exhibited no association with procedural time. Consistent results were observed in analyses using fluoroscopy time as the outcome.
Conclusion
RV dilation and severe TR were associated with prolonged procedural time for LP implantation. Anatomic features obtained from preprocedural echocardiography could provide valuable insights into both the safety and efficiency of LP implantation, thereby enhancing tailored treatment strategies for patients undergoing pacemaker implantation.