Quantification of Artifacts and Image Distortions in 1.5 Tesla Magnetic Resonance Images of a Commercial Multi-Channel Vaginal Cylinder Brachytherapy Applicator Set.
Abolfazl Kanani, Ali Fatemi-Ardakani, Amir M Owrangi, Mehran Yazdi, Hadi Baghbani, Mohammad Amin Mosleh-Shirazi
{"title":"Quantification of Artifacts and Image Distortions in 1.5 Tesla Magnetic Resonance Images of a Commercial Multi-Channel Vaginal Cylinder Brachytherapy Applicator Set.","authors":"Abolfazl Kanani, Ali Fatemi-Ardakani, Amir M Owrangi, Mehran Yazdi, Hadi Baghbani, Mohammad Amin Mosleh-Shirazi","doi":"10.31661/jbpe.v0i0.2309-1665","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The BEBIG Portio multi-channel applicator provides better target dose coverage and sparing organs-at-risk compared to a single-channel cylinder. However, artifacts and distortions of Portio in magnetic resonance images (MRI) have not yet been reported.</p><p><strong>Objective: </strong>We aimed to quantify the artifacts and distortions in its 1.5-Tesla MR images before clinical use.</p><p><strong>Material and methods: </strong>In this experimental study, we employed a gelatin-filled phantom to conduct our measurements. T2-weighted (T2W) images were examined for artifacts and distortions. Computed tomography (CT) images were used as a reference to assess image distortions. Artifact severity was measured by recording the full-width-at-half-maximum (FWHM) image pixel values at various positions along the length of the applicator/channels. CT and MRI-based applicator reconstruction accuracy were then compared, and signal-to-noise ratio (SNR) and contrast were also determined for the applicator images.</p><p><strong>Results: </strong>The applicator distortion level for the Portio applicator was less than the image spatial resolution (0.5±0.5 pixels). The average FWHM for the tandem applicator images was 5.23±0.39 mm, while it was 3.21±0.37 mm for all channels (compared to their actual diameters of 5.0 mm and 3.0 mm, respectively). The average applicator reconstruction difference between CT and MR images was 0.75±0.30 mm overall source dwell positions. The image SNR and contrast were both acceptable.</p><p><strong>Conclusion: </strong>These findings indicate that the Portio applicator has a satisfactory low level of artifacts and image distortions in 1.5-Tesla, T2W images. It may, therefore, be a promising option for MRI-guided multi-channel vaginal brachytherapy.</p>","PeriodicalId":38035,"journal":{"name":"Journal of Biomedical Physics and Engineering","volume":"13 6","pages":"523-534"},"PeriodicalIF":0.0000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10749412/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Biomedical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31661/jbpe.v0i0.2309-1665","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The BEBIG Portio multi-channel applicator provides better target dose coverage and sparing organs-at-risk compared to a single-channel cylinder. However, artifacts and distortions of Portio in magnetic resonance images (MRI) have not yet been reported.
Objective: We aimed to quantify the artifacts and distortions in its 1.5-Tesla MR images before clinical use.
Material and methods: In this experimental study, we employed a gelatin-filled phantom to conduct our measurements. T2-weighted (T2W) images were examined for artifacts and distortions. Computed tomography (CT) images were used as a reference to assess image distortions. Artifact severity was measured by recording the full-width-at-half-maximum (FWHM) image pixel values at various positions along the length of the applicator/channels. CT and MRI-based applicator reconstruction accuracy were then compared, and signal-to-noise ratio (SNR) and contrast were also determined for the applicator images.
Results: The applicator distortion level for the Portio applicator was less than the image spatial resolution (0.5±0.5 pixels). The average FWHM for the tandem applicator images was 5.23±0.39 mm, while it was 3.21±0.37 mm for all channels (compared to their actual diameters of 5.0 mm and 3.0 mm, respectively). The average applicator reconstruction difference between CT and MR images was 0.75±0.30 mm overall source dwell positions. The image SNR and contrast were both acceptable.
Conclusion: These findings indicate that the Portio applicator has a satisfactory low level of artifacts and image distortions in 1.5-Tesla, T2W images. It may, therefore, be a promising option for MRI-guided multi-channel vaginal brachytherapy.