Daniel Conrad Kirchhoff, Wei Zhang, Athanasia Chandras, Damodara Rao Mendu
{"title":"Analytical assessment and validation of the ProteinSimple ELLA serum B-cell maturation antigen assay","authors":"Daniel Conrad Kirchhoff, Wei Zhang, Athanasia Chandras, Damodara Rao Mendu","doi":"10.1016/j.plabm.2023.e00354","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><p>Soluble B-Cell Maturation Antigen (sBCMA) is a degradation product of plasma cell-bound BCMA found in serum. Serum sBCMA concentrations correlate with bone marrow plasma cellularity, making it an attractive biomarker for monitoring plasma cell disorders, such as multiple myeloma. Here we evaluated the automated BCMA immunoassay for the ProteinSimple ELLA, for the analysis of sBCMA.</p></div><div><h3>Design & methods</h3><p>Inter and intra-run precision was assessed through replicate sBCMA measurements at 3 different concentration levels. Linearity was determined through serial dilution of a high sBMCA patient sample. Accuracy was assessed through split specimen analysis on two separate lots of reagents. Stability was assessed at 3 temperature levels over 14 days. Cross-reactivity was assessed on BCMA targeting and non-targeting chemotherapeutics. A reference range was established through the analysis of 146 healthy donor samples. The effect of endogenous interferents was assessed through spiking and recovery studies.</p></div><div><h3>Results</h3><p>Inter and intra-run precision studies afforded CVs of <10% at all three concentration levels. Analytical measurement range was confirmed from 0.1 to 7 ng/mL. Accuracy studies afforded a slope of 0.976, intercept of 1.22, R<sup>2</sup> of 0.996. Assayed sBCMA values were unaffected by endogenous interferents and non-BMCA targeting antibodies. BCMA targeting therapeutics negatively affected assayed sBCMA concentrations. The reference range was established at 19–58 ng/mL sBCMA is analytically stable.</p></div><div><h3>Conclusions</h3><p>The ProteinSimple ELLA sBCMA assay shows acceptable performance for the clinical assessment of sBCMA. The assay was highly affected by BCMA targeting therapeutics, thereby patients undergoing this therapy should not have their sBCMA levels assessed by this method.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"38 ","pages":"Article e00354"},"PeriodicalIF":1.7000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551723000483/pdfft?md5=9b9411a7476dae998d012d5ec1e6ae83&pid=1-s2.0-S2352551723000483-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551723000483","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives
Soluble B-Cell Maturation Antigen (sBCMA) is a degradation product of plasma cell-bound BCMA found in serum. Serum sBCMA concentrations correlate with bone marrow plasma cellularity, making it an attractive biomarker for monitoring plasma cell disorders, such as multiple myeloma. Here we evaluated the automated BCMA immunoassay for the ProteinSimple ELLA, for the analysis of sBCMA.
Design & methods
Inter and intra-run precision was assessed through replicate sBCMA measurements at 3 different concentration levels. Linearity was determined through serial dilution of a high sBMCA patient sample. Accuracy was assessed through split specimen analysis on two separate lots of reagents. Stability was assessed at 3 temperature levels over 14 days. Cross-reactivity was assessed on BCMA targeting and non-targeting chemotherapeutics. A reference range was established through the analysis of 146 healthy donor samples. The effect of endogenous interferents was assessed through spiking and recovery studies.
Results
Inter and intra-run precision studies afforded CVs of <10% at all three concentration levels. Analytical measurement range was confirmed from 0.1 to 7 ng/mL. Accuracy studies afforded a slope of 0.976, intercept of 1.22, R2 of 0.996. Assayed sBCMA values were unaffected by endogenous interferents and non-BMCA targeting antibodies. BCMA targeting therapeutics negatively affected assayed sBCMA concentrations. The reference range was established at 19–58 ng/mL sBCMA is analytically stable.
Conclusions
The ProteinSimple ELLA sBCMA assay shows acceptable performance for the clinical assessment of sBCMA. The assay was highly affected by BCMA targeting therapeutics, thereby patients undergoing this therapy should not have their sBCMA levels assessed by this method.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.