Sathiyan Saminathan, Mageshraja Kannan, K M Ganesh, James Mazarello, Rahul Fernandes
{"title":"Characteristics Performance of Newly Developed Parallel Plate Chamber in Electron Beams.","authors":"Sathiyan Saminathan, Mageshraja Kannan, K M Ganesh, James Mazarello, Rahul Fernandes","doi":"10.4103/jmp.jmp_76_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To investigate the dosimetric performance of newly developed parallel plate chamber in electron beams.</p><p><strong>Materials and methods: </strong>Rosalina Instruments India Private Limited (Mumbai, Maharashtra, India) has designed and fabricated PRATT2 parallel plate chamber. The various dosimetric characteristics, including pre- and post-irradiation leakage, stability, polarity effect, chamber response with bias voltage, dose linearity, dose rate effect, and chamber absorbed dose calibration, were performed for the developed chamber. The electron beam energies of 4, 6, 8, and 15MeV were used in this study.</p><p><strong>Results: </strong>The pre- and post-irradiation leakage of the developed chamber was within the acceptable limit. The chamber shows good stability over the electron beams used in this study. The maximum error in polarity effect was 0.7% for the developed chamber. The chamber shows the good linear response with dose, and its response is independent of the dose rate for all electron beams. The beam quality correction factor (k<sub>Q, Q0</sub>) was determined for the all electron beam energies, which was used for determination absorbed dose in electron beams.</p><p><strong>Discussion: </strong>The developed parallel plate chamber (PRATT2) is suitable for dosimetry of electron beams in radiotherapy. The chamber is cost effective and shows precise and reproducible response. The study carried out confirms that the newly designed and fabricated ion chamber can be used in the measurement of absorbed dose for therapeutic electron beams.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"48 4","pages":"333-337"},"PeriodicalIF":0.7000,"publicationDate":"2023-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10783180/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_76_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/12/5 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Background: To investigate the dosimetric performance of newly developed parallel plate chamber in electron beams.
Materials and methods: Rosalina Instruments India Private Limited (Mumbai, Maharashtra, India) has designed and fabricated PRATT2 parallel plate chamber. The various dosimetric characteristics, including pre- and post-irradiation leakage, stability, polarity effect, chamber response with bias voltage, dose linearity, dose rate effect, and chamber absorbed dose calibration, were performed for the developed chamber. The electron beam energies of 4, 6, 8, and 15MeV were used in this study.
Results: The pre- and post-irradiation leakage of the developed chamber was within the acceptable limit. The chamber shows good stability over the electron beams used in this study. The maximum error in polarity effect was 0.7% for the developed chamber. The chamber shows the good linear response with dose, and its response is independent of the dose rate for all electron beams. The beam quality correction factor (kQ, Q0) was determined for the all electron beam energies, which was used for determination absorbed dose in electron beams.
Discussion: The developed parallel plate chamber (PRATT2) is suitable for dosimetry of electron beams in radiotherapy. The chamber is cost effective and shows precise and reproducible response. The study carried out confirms that the newly designed and fabricated ion chamber can be used in the measurement of absorbed dose for therapeutic electron beams.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.