Reference Ranges for Regional Cerebral Oxygen Saturation with Masimo O3 after Birth and Differences with Other Devices.

IF 1.5 4区 医学 Q3 OBSTETRICS & GYNECOLOGY American journal of perinatology Pub Date : 2024-09-01 Epub Date: 2024-01-25 DOI:10.1055/a-2253-8740
Luis Bachiller Carnicero, Sonia Caserío Carbonero
{"title":"Reference Ranges for Regional Cerebral Oxygen Saturation with Masimo O3 after Birth and Differences with Other Devices.","authors":"Luis Bachiller Carnicero, Sonia Caserío Carbonero","doi":"10.1055/a-2253-8740","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong> Cerebral oximetry using near-infrared spectroscopy (NIRS) is a noninvasive optical technology widely used in neonatology. The present study aimed to define reference ranges for cerebral tissue oxygen saturation (crSO<sub>2</sub>) with a new four-wavelength NIRS device, Masimo O3 oximeter, during immediate transition after birth and compare values with those obtained previously with NIRO 200NX<sup>®</sup>.</p><p><strong>Study design: </strong> This was a prospective observational study using Masimo O3 device to measure crSO<sub>2</sub> and regional cerebral fractional tissue oxygen extraction (cFTOE) in healthy term newborns delivered by primary cesarean section, during the 15 minutes after cord clamping. The neonates who required any medical support were excluded. The NIRS sensor was placed on the right forehead. Peripheral oxygen saturation and heart rate were continuously measured by pulse oximetry. Previous studies which established centiles for crSO<sub>2</sub> with NIRO 200NX were used for comparison.</p><p><strong>Results: </strong> A total of 44 newborns were included. The median crSO<sub>2</sub> and cFTOE (interquartile range) at 2, 5, and 7 minutes was 54% (49-54), 71% (64-86), and 79% (73-84) and 0,25 (0,18-0,33), 0,19 (0,15-0,23), and 0,16 (0,12-0,21), respectively, with no further changes afterwards. The crSO<sub>2</sub> measurements were significantly higher with Masimo O3 compared with NIRO-200NX.</p><p><strong>Conclusion: </strong> The present observational study presented reference ranges for crSO<sub>2</sub> and cFTOE measured with Masimo O3 oximeter during the immediate neonatal transition. Values obtained with O3 were higher than those obtained with other oximeters. For this reason, crSO<sub>2</sub> is device-specific so there must be known reference values for each oximeter to define therapeutic interventions based on crSO<sub>2</sub> and assess cerebral oxygenation in clinical studies.</p><p><strong>Key points: </strong>· Masimo O3 uses four wavelengths to measure regional oxygen saturation value.. · O3 values of crSO2 and cFTOE differ with other neonatal oximeters at birth.. · Knowledge of reference range of O3 at birth is essential to guide resuscitation..</p>","PeriodicalId":7584,"journal":{"name":"American journal of perinatology","volume":null,"pages":null},"PeriodicalIF":1.5000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of perinatology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2253-8740","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/25 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"OBSTETRICS & GYNECOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective:  Cerebral oximetry using near-infrared spectroscopy (NIRS) is a noninvasive optical technology widely used in neonatology. The present study aimed to define reference ranges for cerebral tissue oxygen saturation (crSO2) with a new four-wavelength NIRS device, Masimo O3 oximeter, during immediate transition after birth and compare values with those obtained previously with NIRO 200NX®.

Study design:  This was a prospective observational study using Masimo O3 device to measure crSO2 and regional cerebral fractional tissue oxygen extraction (cFTOE) in healthy term newborns delivered by primary cesarean section, during the 15 minutes after cord clamping. The neonates who required any medical support were excluded. The NIRS sensor was placed on the right forehead. Peripheral oxygen saturation and heart rate were continuously measured by pulse oximetry. Previous studies which established centiles for crSO2 with NIRO 200NX were used for comparison.

Results:  A total of 44 newborns were included. The median crSO2 and cFTOE (interquartile range) at 2, 5, and 7 minutes was 54% (49-54), 71% (64-86), and 79% (73-84) and 0,25 (0,18-0,33), 0,19 (0,15-0,23), and 0,16 (0,12-0,21), respectively, with no further changes afterwards. The crSO2 measurements were significantly higher with Masimo O3 compared with NIRO-200NX.

Conclusion:  The present observational study presented reference ranges for crSO2 and cFTOE measured with Masimo O3 oximeter during the immediate neonatal transition. Values obtained with O3 were higher than those obtained with other oximeters. For this reason, crSO2 is device-specific so there must be known reference values for each oximeter to define therapeutic interventions based on crSO2 and assess cerebral oxygenation in clinical studies.

Key points: · Masimo O3 uses four wavelengths to measure regional oxygen saturation value.. · O3 values of crSO2 and cFTOE differ with other neonatal oximeters at birth.. · Knowledge of reference range of O3 at birth is essential to guide resuscitation..

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
出生后使用 Masimo O3® 测量区域脑氧饱和度的参考范围以及与其他设备的差异。
- 目的:使用近红外光谱(NIRS)进行脑氧饱和度测量是一种广泛应用于新生儿科的无创光学技术。本工作室的目的是使用新型 4 波长近红外设备 Masimo O3® 血氧仪确定出生后过渡时期脑组织血氧饱和度(crSO2)的参考范围,并与之前使用 NIRO 200NX® 获得的数值进行比较。-研究设计:这是一项前瞻性观察研究,使用 Masimo O3 设备测量初次剖宫产的健康足月新生儿在脐带钳夹后 15 分钟内的 CrSO2 和区域脑组织分数氧萃取 (cFTOE)。需要任何医疗支持的新生儿被排除在外。近红外传感器置于右前额。通过脉搏血氧仪持续测量外周血氧饱和度和心率。之前的研究使用 NIRO 200 NX 确定了 crSO2 的百分位数,并将其用于比较。-结果共纳入 44 名新生儿。在 2、5 和 7 分钟时,crSO2 和 cFTOE 的中位数(四分位间范围)分别为 54% (49-54)、71% (64-86) 和 79% (73-84),以及 0,25 (0,18-0,33); 0,19 (0,15-0,23) 和 0,16 (0,12-0,21),之后没有进一步变化。与 NIRO-200NX 相比,Masimo O3 的 crSO2 测量值明显更高。-结论:本观察性研究提供了新生儿过渡期使用 Masimo O3 血氧计测量的 crSO2 和 cFTOE 的参考范围。使用 O3 获得的值高于使用其他血氧仪获得的值。因此,crSO2 具有设备特异性,必须知道每种血氧仪的参考值,才能根据 crSO2 确定治疗干预措施,并进行脑氧合临床研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
American journal of perinatology
American journal of perinatology 医学-妇产科学
CiteScore
5.90
自引率
0.00%
发文量
302
审稿时长
4-8 weeks
期刊介绍: The American Journal of Perinatology is an international, peer-reviewed, and indexed journal publishing 14 issues a year dealing with original research and topical reviews. It is the definitive forum for specialists in obstetrics, neonatology, perinatology, and maternal/fetal medicine, with emphasis on bridging the different fields. The focus is primarily on clinical and translational research, clinical and technical advances in diagnosis, monitoring, and treatment as well as evidence-based reviews. Topics of interest include epidemiology, diagnosis, prevention, and management of maternal, fetal, and neonatal diseases. Manuscripts on new technology, NICU set-ups, and nursing topics are published to provide a broad survey of important issues in this field. All articles undergo rigorous peer review, with web-based submission, expedited turn-around, and availability of electronic publication. The American Journal of Perinatology is accompanied by AJP Reports - an Open Access journal for case reports in neonatology and maternal/fetal medicine.
期刊最新文献
A Description of IVIG Use in Term Neonates with ABO Incompatibility. Early-Pregnancy Resilience Characteristics before versus during the COVID-19 Pandemic. Infant Mortality Categorized by Birth Weight Percentiles for Deliveries between 22 and 28 Weeks of Gestation. Are Racial Disparities in Cesarean Due to Differences in Labor Induction Management? Factors Associated with the Uptake of Long-Acting Reversible Contraception and Contraceptive Use in Postpartum People with HIV at a Single Tertiary Care Center.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1