Discovery Elbow System arthroplasty polyethylene bearing exchange: outcomes and experience.

IF 1.8 Q2 ORTHOPEDICS Clinics in Shoulder and Elbow Pub Date : 2024-03-01 Epub Date: 2024-01-26 DOI:10.5397/cise.2023.00668
Daniel L J Morris, Katherine Walstow, Lisa Pitt, Marie Morgan, Amol A Tambe, D I Clark, Timothy Cresswell, Marius P Espag
{"title":"Discovery Elbow System arthroplasty polyethylene bearing exchange: outcomes and experience.","authors":"Daniel L J Morris, Katherine Walstow, Lisa Pitt, Marie Morgan, Amol A Tambe, D I Clark, Timothy Cresswell, Marius P Espag","doi":"10.5397/cise.2023.00668","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The Discovery Elbow System (DES) utilizes a polyethylene bearing within the ulnar component. An exchange bearing requires preoperative freezing and implantation within 2 minutes of freezer removal to allow insertion. We report our outcomes and experience using this technique.</p><p><strong>Methods: </strong>This was an analysis of a two-surgeon consecutive series of DES bearing exchange. Inclusion criteria included patients in which exchange was attempted with a minimum 1-year follow-up. Clinical and radiographic review was performed 1, 2, 3, 5, 8 and 10 years postoperative. Outcome measures included range of movement, Oxford Elbow Score (OES), Mayo Elbow Performance Score (MEPS), complications and requirement for revision surgery.</p><p><strong>Results: </strong>Eleven DESs in 10 patients were included. Indications were bearing wear encountered during humeral component revision (n=5); bearing failure (n=4); and infection treated with debridement, antibiotics and implant retention (DAIR; n=2). Bearing exchange was conducted on the first attempt in 10 cases. One case required a second attempt. One patient developed infection postoperatively managed with two-stage revision. Mean follow-up of the bearing exchange DES was 3 years. No further surgery was required, with no infection recurrence in DAIR cases. Mean elbow flexion-extension and pronosupination arcs were 107° (±22°) and 140° (±26°). Mean OES was 36/48 (±12) and MEPS was 83/100 (±19).</p><p><strong>Conclusions: </strong>Our results support the use of DES bearing exchange in cases of bearing wear with well-fixed stems or acute infection. This series provides surgeons managing DES arthroplasty with management principles, successful and reproducible surgical techniques and expected clinical outcomes in performing DES polyethylene bearing exchange. Level of evidence: IV.</p>","PeriodicalId":33981,"journal":{"name":"Clinics in Shoulder and Elbow","volume":" ","pages":"18-25"},"PeriodicalIF":1.8000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10938009/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinics in Shoulder and Elbow","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5397/cise.2023.00668","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/26 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The Discovery Elbow System (DES) utilizes a polyethylene bearing within the ulnar component. An exchange bearing requires preoperative freezing and implantation within 2 minutes of freezer removal to allow insertion. We report our outcomes and experience using this technique.

Methods: This was an analysis of a two-surgeon consecutive series of DES bearing exchange. Inclusion criteria included patients in which exchange was attempted with a minimum 1-year follow-up. Clinical and radiographic review was performed 1, 2, 3, 5, 8 and 10 years postoperative. Outcome measures included range of movement, Oxford Elbow Score (OES), Mayo Elbow Performance Score (MEPS), complications and requirement for revision surgery.

Results: Eleven DESs in 10 patients were included. Indications were bearing wear encountered during humeral component revision (n=5); bearing failure (n=4); and infection treated with debridement, antibiotics and implant retention (DAIR; n=2). Bearing exchange was conducted on the first attempt in 10 cases. One case required a second attempt. One patient developed infection postoperatively managed with two-stage revision. Mean follow-up of the bearing exchange DES was 3 years. No further surgery was required, with no infection recurrence in DAIR cases. Mean elbow flexion-extension and pronosupination arcs were 107° (±22°) and 140° (±26°). Mean OES was 36/48 (±12) and MEPS was 83/100 (±19).

Conclusions: Our results support the use of DES bearing exchange in cases of bearing wear with well-fixed stems or acute infection. This series provides surgeons managing DES arthroplasty with management principles, successful and reproducible surgical techniques and expected clinical outcomes in performing DES polyethylene bearing exchange. Level of evidence: IV.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Discovery 肘关节置换系统聚乙烯轴承交换:结果和经验。
背景:发现者肘关节系统(DES)在尺侧组件内使用聚乙烯轴承。交换支座需要术前冷冻,并在取出冷冻装置后 2 分钟内植入,以便插入。我们报告了使用这种技术的结果和经验:这是一项由两名外科医生连续进行的DES轴承交换系列分析。纳入标准包括至少随访1年并尝试过置换的患者。术后1年、2年、3年、5年、8年和10年进行临床和放射学复查。结果指标包括活动范围、牛津肘评分(OES)、梅奥肘关节表现评分(MEPS)、并发症和翻修手术要求:结果:纳入了10名患者的11个DES。结果:纳入了10名患者的11个DES,适应症包括肱骨组件翻修过程中遇到的轴承磨损(5例);轴承故障(4例);以及通过清创、抗生素和植入物保留(DAIR;2例)治疗的感染。10例患者在第一次尝试时进行了轴承交换。1例需要第二次尝试。一名患者术后发生感染,经两期翻修处理。轴承交换 DES 的平均随访时间为 3 年。无需进一步手术,DAIR病例无感染复发。平均肘关节屈伸弧度为107°(±22°),前屈弧度为140°(±26°)。平均OES为36/48(±12),MEPS为83/100(±19):我们的研究结果支持在轴承磨损、柄固定良好或急性感染的病例中使用DES轴承交换。该系列研究为管理DES关节置换术的外科医生提供了管理原则、成功且可重复的手术技巧以及进行DES聚乙烯轴承交换时的预期临床结果。证据级别:IV级。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
0.30
自引率
0.00%
发文量
55
审稿时长
15 weeks
期刊最新文献
Use of posteroanterior reference guides for bone block or coracoid process transfer in anterior glenohumeral instability: a cadaveric study of the relationship to the suprascapular nerve. Surgical technique for secondary pectoralis major transfer after reverse total shoulder arthroplasty to treat residual internal rotation dysfunction. Epithelioid hemangioma of the acromion causing shoulder pain: a case report. Glenohumeral internal rotation deficit: insights into pathologic, clinical, diagnostic, and therapeutic characteristics. Persistent lateral elbow pain from overlooked posterolateral impingement of the elbow: a literature review and guidance for treatment.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1