Imprecision of high-sensitivity cardiac troponin assays at the female 99th-percentile

IF 2.5 3区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY Clinical biochemistry Pub Date : 2024-02-13 DOI:10.1016/j.clinbiochem.2024.110731
Peter A. Kavsak , Lorna Clark , Saranya Arnoldo , Amy Lou , Jennifer L. Shea , Shaun Eintracht , Andrew W. Lyon , Vipin Bhayana , Laurel Thorlacius , Joshua E. Raizman , Albert Tsui , Rose Djiana , Michael Chen , Yun Huang , Ali Haider , Ronald A. Booth , Chris McCudden , Paul M. Yip , Daniel Beriault , David Blank , Allan S. Jaffe
{"title":"Imprecision of high-sensitivity cardiac troponin assays at the female 99th-percentile","authors":"Peter A. Kavsak ,&nbsp;Lorna Clark ,&nbsp;Saranya Arnoldo ,&nbsp;Amy Lou ,&nbsp;Jennifer L. Shea ,&nbsp;Shaun Eintracht ,&nbsp;Andrew W. Lyon ,&nbsp;Vipin Bhayana ,&nbsp;Laurel Thorlacius ,&nbsp;Joshua E. Raizman ,&nbsp;Albert Tsui ,&nbsp;Rose Djiana ,&nbsp;Michael Chen ,&nbsp;Yun Huang ,&nbsp;Ali Haider ,&nbsp;Ronald A. Booth ,&nbsp;Chris McCudden ,&nbsp;Paul M. Yip ,&nbsp;Daniel Beriault ,&nbsp;David Blank ,&nbsp;Allan S. Jaffe","doi":"10.1016/j.clinbiochem.2024.110731","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>An analytical benchmark for high-sensitivity cardiac troponin (hs-cTn) assays is to achieve a coefficient of variation (CV) of ≤ 10.0 % at the 99th percentile upper reference limit (URL) used for the diagnosis of myocardial infarction. Few prospective multicenter studies have evaluated assay imprecision and none have determined precision at the female URL which is lower than the male URL for all cardiac troponin assays.</p></div><div><h3>Methods</h3><p>Human serum and plasma matrix samples were constructed to yield hs-cTn concentrations near the female URLs for the Abbott, Beckman, Roche, and Siemens hs-cTn assays. These materials were sent (on dry ice) to 35 Canadian hospital laboratories (n = 64 instruments evaluated) participating in a larger clinical trial, with instructions for storage, handling, and monthly testing over one year. The mean concentration, standard deviation, and CV for each instrument type and an overall pooled CV for each manufacturer were calculated.</p></div><div><h3>Results</h3><p>The CVs for all individual instruments and overall were ≤ 10.0 % for two manufacturers (Abbott CV<sub>pooled</sub> = 6.3 % and Beckman CV<sub>pooled</sub> = 7.0 %). One of four Siemens Atellica instruments yielded a CV &gt; 10.0 % (CV<sub>pooled</sub> = 7.7 %), whereas 15 of 41 Roche instruments yielded CVs &gt; 10.0 % at the female URL of 9 ng/L used worldwide (6 cobas e411, 1 cobas e601, 4 cobas e602, and 4 cobas e801) (CV<sub>pooled</sub> = 11.7 %). Four Roche instruments also yielded CVs &gt; 10.0 % near the female URL of 14 ng/L used in the United States (CV<sub>pooled</sub> = 8.5 %).</p></div><div><h3>Conclusions</h3><p>The number of instruments achieving a CV ≤ 10.0 % at the female 99th-percentile URL varies by manufacturer and by instrument. Monitoring assay precision at the female URL is necessary for some assays to ensure optimal use of this threshold in clinical practice.</p></div>","PeriodicalId":10172,"journal":{"name":"Clinical biochemistry","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-02-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0009912024000250/pdfft?md5=52e249591a3160ac3ecd0fabbb6789b2&pid=1-s2.0-S0009912024000250-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical biochemistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0009912024000250","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background

An analytical benchmark for high-sensitivity cardiac troponin (hs-cTn) assays is to achieve a coefficient of variation (CV) of ≤ 10.0 % at the 99th percentile upper reference limit (URL) used for the diagnosis of myocardial infarction. Few prospective multicenter studies have evaluated assay imprecision and none have determined precision at the female URL which is lower than the male URL for all cardiac troponin assays.

Methods

Human serum and plasma matrix samples were constructed to yield hs-cTn concentrations near the female URLs for the Abbott, Beckman, Roche, and Siemens hs-cTn assays. These materials were sent (on dry ice) to 35 Canadian hospital laboratories (n = 64 instruments evaluated) participating in a larger clinical trial, with instructions for storage, handling, and monthly testing over one year. The mean concentration, standard deviation, and CV for each instrument type and an overall pooled CV for each manufacturer were calculated.

Results

The CVs for all individual instruments and overall were ≤ 10.0 % for two manufacturers (Abbott CVpooled = 6.3 % and Beckman CVpooled = 7.0 %). One of four Siemens Atellica instruments yielded a CV > 10.0 % (CVpooled = 7.7 %), whereas 15 of 41 Roche instruments yielded CVs > 10.0 % at the female URL of 9 ng/L used worldwide (6 cobas e411, 1 cobas e601, 4 cobas e602, and 4 cobas e801) (CVpooled = 11.7 %). Four Roche instruments also yielded CVs > 10.0 % near the female URL of 14 ng/L used in the United States (CVpooled = 8.5 %).

Conclusions

The number of instruments achieving a CV ≤ 10.0 % at the female 99th-percentile URL varies by manufacturer and by instrument. Monitoring assay precision at the female URL is necessary for some assays to ensure optimal use of this threshold in clinical practice.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
女性第 99 百分位数高灵敏度心肌肌钙蛋白测定的不精确性
背景高灵敏度心肌肌钙蛋白(hs-cTn)测定的一个分析基准是在用于诊断心肌梗死的第 99 百分位数参考上限(URL)处变异系数(CV)小于 10.0%。很少有前瞻性多中心研究对测定的不精确性进行评估,也没有研究对所有心肌肌钙蛋白测定法的女性参考值上限(URL)的精确性进行测定,因为女性参考值上限低于男性参考值上限。方法构建人体血清和血浆基质样本,以获得雅培、贝克曼、罗氏和西门子 hs-cTn 测定法女性参考值上限附近的 hs-cTn 浓度。这些材料被(用干冰)送往参与大型临床试验的 35 家加拿大医院实验室(n = 64 台评估仪器),并附有储存、处理和一年内每月检测的说明。结果有两家制造商(雅培 CVpooled = 6.3 %,贝克曼 CVpooled = 7.0 %)的所有单个仪器和总体 CV 均小于 10.0 %。四台西门子 Atellica 仪器中有一台的 CV 值为 10.0%(CVpooled = 7.7%),而 41 台罗氏仪器中有 15 台(6 台 cobas e411、1 台 cobas e601、4 台 cobas e602 和 4 台 cobas e801)在全球使用的女性 URL 为 9 ng/L 时的 CV 值为 10.0%(CVpooled = 11.7%)。结论在女性第 99 百分位数URL处实现 CV≤10.0 %的仪器数量因制造商和仪器而异。对某些检测方法而言,监测女性第99百分位数URL的检测精度是必要的,以确保在临床实践中最佳使用这一阈值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Clinical biochemistry
Clinical biochemistry 医学-医学实验技术
CiteScore
5.10
自引率
0.00%
发文量
151
审稿时长
25 days
期刊介绍: Clinical Biochemistry publishes articles relating to clinical chemistry, molecular biology and genetics, therapeutic drug monitoring and toxicology, laboratory immunology and laboratory medicine in general, with the focus on analytical and clinical investigation of laboratory tests in humans used for diagnosis, prognosis, treatment and therapy, and monitoring of disease.
期刊最新文献
Deteriorated thiol-disulphide and oxidized-reduced glutathione status in blood in Alzheimer’s disease Editorial Board In vivo and in vitro relationship between ionized magnesium and ionized calcium Heart failure biomarkers and prediction of early left ventricle remodeling after acute coronary syndromes Serum glucose mediated association of serum lactate with acute kidney injury among AIS patients
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1