{"title":"Intentional Oversizing of Valve in Transcatheter Aortic Valve Replacement: Is Bigger Better? A Large, Single-Center Experience","authors":"Khawaja Afzal Ammar MD , Alexandria Graeber BS , Abdur Rahman Ahmad MD , Jodi Zilinski MD , Daniel P. O’Hair MD , Renuka Jain MD , Suhail Q. Allaqaband MD , Tanvir Bajwa MD","doi":"10.1016/j.shj.2023.100278","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>The current clinical practice standard is 10% to 20% oversizing of self-expanding valves in transcatheter aortic valve replacement. We aimed to determine whether >20% oversizing of self-expanding valves (Medtronic Evolut) would lead to better valve performance with similar or better outcomes.</p></div><div><h3>Methods</h3><p>From October 2011 to December 2016, we approached all transcatheter aortic valve replacement patients with a conscious attempt at large oversizing (>20%). The most common valve used, excluding those used in valve-in-valve patients, was the 29-mm Evolut R (29%). We used a retrospective chart review to compare moderate oversizing (group 1; 10% to 20%) with large oversizing (group 2; >20%).</p></div><div><h3>Results</h3><p>Of 556 patients, 45% were male; the overall mean Society of Thoracic Surgeons risk score was 5.8 ± 3.8. Eighty-five (15%) patients needed a pacemaker, and 21 (3.8%) developed significant paravalvular leak. Mean oversizing was 20.3% ± 6.0%, with 41.4% of patients included in group 1 and 54.5% in group 2. Incidences of complications in group 2 vs. group 1 were as follows: a) paravalvular leak (2.0 vs. 6.1%; odds ratio = 0.31, <em>p</em> = 0.01), b) pacemaker (15 vs. 14%), c) gastrointestinal bleed (n = 4 vs. 0; 1.3 vs. 0.0%; <em>p</em> = 0.03), d) annular dissection (n = 1 vs. 0; 0.3 vs. 0%; <em>p</em> = 0.29), e) mortality (n = 5 vs. 4; 1.6 vs. 1.7%). Incidence of paravalvular leak was higher in those who died than survivors (13 vs. 1.3%; <em>p</em> ≤ 0.0001).</p></div><div><h3>Conclusions</h3><p>These data suggest that, in current self-expanding valves, >20% oversizing delivers a significantly lower prevalence of paravalvular leak without an increase in other complications. Since paravalvular leak is associated with increased mortality, >20% oversizing may represent a superior prosthesis choice.</p></div>","PeriodicalId":36053,"journal":{"name":"Structural Heart","volume":"8 3","pages":"Article 100278"},"PeriodicalIF":1.4000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2474870623002294/pdfft?md5=618648d083cfd5a7c1c483c5ce1b3ec9&pid=1-s2.0-S2474870623002294-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Structural Heart","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2474870623002294","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
The current clinical practice standard is 10% to 20% oversizing of self-expanding valves in transcatheter aortic valve replacement. We aimed to determine whether >20% oversizing of self-expanding valves (Medtronic Evolut) would lead to better valve performance with similar or better outcomes.
Methods
From October 2011 to December 2016, we approached all transcatheter aortic valve replacement patients with a conscious attempt at large oversizing (>20%). The most common valve used, excluding those used in valve-in-valve patients, was the 29-mm Evolut R (29%). We used a retrospective chart review to compare moderate oversizing (group 1; 10% to 20%) with large oversizing (group 2; >20%).
Results
Of 556 patients, 45% were male; the overall mean Society of Thoracic Surgeons risk score was 5.8 ± 3.8. Eighty-five (15%) patients needed a pacemaker, and 21 (3.8%) developed significant paravalvular leak. Mean oversizing was 20.3% ± 6.0%, with 41.4% of patients included in group 1 and 54.5% in group 2. Incidences of complications in group 2 vs. group 1 were as follows: a) paravalvular leak (2.0 vs. 6.1%; odds ratio = 0.31, p = 0.01), b) pacemaker (15 vs. 14%), c) gastrointestinal bleed (n = 4 vs. 0; 1.3 vs. 0.0%; p = 0.03), d) annular dissection (n = 1 vs. 0; 0.3 vs. 0%; p = 0.29), e) mortality (n = 5 vs. 4; 1.6 vs. 1.7%). Incidence of paravalvular leak was higher in those who died than survivors (13 vs. 1.3%; p ≤ 0.0001).
Conclusions
These data suggest that, in current self-expanding valves, >20% oversizing delivers a significantly lower prevalence of paravalvular leak without an increase in other complications. Since paravalvular leak is associated with increased mortality, >20% oversizing may represent a superior prosthesis choice.