[Application Research of Vector Flow Technique on Convex Array Ultrasonic Probe of Abdomen].

Penghui Hao, Yigang Du, Shuangshuang Li, Lei Zhu, Xujin He
{"title":"[Application Research of Vector Flow Technique on Convex Array Ultrasonic Probe of Abdomen].","authors":"Penghui Hao, Yigang Du, Shuangshuang Li, Lei Zhu, Xujin He","doi":"10.3969/j.issn.1671-7104.230086","DOIUrl":null,"url":null,"abstract":"<p><p>Vector flow imaging (VFI) is an innovative ultrasound flow measurement technology. Compared with the traditional color Doppler and spectral Doppler, VFI has the advantages of independence of angle correction and direct acquisition of real-time amplitude and direction of flow. Transverse oscillation (TO) method is one of the effective methods for vector flow imaging. However, a complete and detailed algorithm validation process based on commercial ultrasound machines is still lacking due to more complex convex probes. This study starts with introducing the imaging process and principle of transverse oscillation vector flow technique, and calculates the error between the set velocity value and the measured velocity value through the simulation experiment, and verifies the error between the set velocity value and the measured velocity value through the Doppler flow phantom experiment. Among them, the velocity value measured by the TO vector flow technique in the simulation experiment is 0.48 m/s and the preset value is 0.50 m/s, the error between them is -4%. The velocity values are 8.33, 11.14, 14.44 and 16.67 cm/s measured by the Doppler flow phantom experiment, the actual velocity values are 7.97, 10.78, 14.06 and 17.34 cm/s, the errors between them are all within ±5%. Both experiments verify the feasibility of using vector flow technique on abdominal convex probe.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-01-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.3969/j.issn.1671-7104.230086","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Vector flow imaging (VFI) is an innovative ultrasound flow measurement technology. Compared with the traditional color Doppler and spectral Doppler, VFI has the advantages of independence of angle correction and direct acquisition of real-time amplitude and direction of flow. Transverse oscillation (TO) method is one of the effective methods for vector flow imaging. However, a complete and detailed algorithm validation process based on commercial ultrasound machines is still lacking due to more complex convex probes. This study starts with introducing the imaging process and principle of transverse oscillation vector flow technique, and calculates the error between the set velocity value and the measured velocity value through the simulation experiment, and verifies the error between the set velocity value and the measured velocity value through the Doppler flow phantom experiment. Among them, the velocity value measured by the TO vector flow technique in the simulation experiment is 0.48 m/s and the preset value is 0.50 m/s, the error between them is -4%. The velocity values are 8.33, 11.14, 14.44 and 16.67 cm/s measured by the Doppler flow phantom experiment, the actual velocity values are 7.97, 10.78, 14.06 and 17.34 cm/s, the errors between them are all within ±5%. Both experiments verify the feasibility of using vector flow technique on abdominal convex probe.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[腹部凸阵超声探头的矢量流技术应用研究]。
矢量血流成像(VFI)是一种创新的超声血流测量技术。与传统的彩色多普勒和频谱多普勒相比,矢量血流成像具有无需角度校正、可直接获取实时血流振幅和方向等优点。横向振荡(TO)法是矢量血流成像的有效方法之一。然而,由于凸探头较为复杂,目前仍缺乏基于商用超声设备的完整而详细的算法验证过程。本研究从介绍横向振荡矢量流技术的成像过程和原理入手,通过模拟实验计算出设定速度值与测量速度值之间的误差,并通过多普勒流动模型实验验证设定速度值与测量速度值之间的误差。其中,模拟实验中通过 TO 矢量流技术测量的速度值为 0.48 m/s,预设值为 0.50 m/s,两者误差为-4%。多普勒流动模型实验测得的速度值分别为 8.33、11.14、14.44 和 16.67 厘米/秒,实际速度值分别为 7.97、10.78、14.06 和 17.34 厘米/秒,它们之间的误差均在±5%以内。这两项实验都验证了在腹部凸面探头上使用矢量流技术的可行性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍:
期刊最新文献
[12-Lead Holter Integrated with Sleep Monitoring Module]. [Bowel Sounds Detection Method Based on ResNet-BiLSTM and Attention Mechanism]. [Clinical Application of Equivalent Uniform Dose in Intensity-Modulated Rotational Radiotherapy Based on Eclipse TPS]. [Clinical Validation of a Prototype Smart Non-Invasive Pregnancy Glucose Monitor]. [Clinical Validation Study of Deep Learning-Generated Magnetic Resonance Images].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1