Successful Retrieval of Filter Embolic Protection Device Fragment Trapped by a Carotid Stent: A Case Report.

Journal of neuroendovascular therapy Pub Date : 2024-01-01 Epub Date: 2024-02-01 DOI:10.5797/jnet.cr.2023-0085
Jun Karakama, Mariko Ishikawa, Sakyo Hirai, Satoru Takahashi, Hikaru Wakabayashi, Hirotaka Sagawa, Shoko Fujii, Kyohei Fujita, Nobuyuki Hirotsune, Kazutaka Sumita
{"title":"Successful Retrieval of Filter Embolic Protection Device Fragment Trapped by a Carotid Stent: A Case Report.","authors":"Jun Karakama, Mariko Ishikawa, Sakyo Hirai, Satoru Takahashi, Hikaru Wakabayashi, Hirotaka Sagawa, Shoko Fujii, Kyohei Fujita, Nobuyuki Hirotsune, Kazutaka Sumita","doi":"10.5797/jnet.cr.2023-0085","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Embolic protection devices are useful for preventing distal embolism during carotid artery stenting (CAS); however, complications have been reported. The successful removal of a filter fragment trapped at the distal edge of a carotid stent during the retrieval procedure is described.</p><p><strong>Case presentation: </strong>CAS was performed for internal carotid artery stenosis in a patient in his 70s, and the carotid stent was successfully placed. During the retrieval procedure, the tip of the filter was trapped at the distal edge of the stent and detached from the filter. Using a snare kit, the filter tip was successfully retrieved, and no postoperative neurological symptoms occurred.</p><p><strong>Conclusion: </strong>The edge of a carotid stent can potentially trap devices. When trapping or fragmentation of a device is suspected, it is necessary to evaluate the situation and cause, and the device should be appropriately retrieved without using force.</p>","PeriodicalId":73856,"journal":{"name":"Journal of neuroendovascular therapy","volume":"18 2","pages":"53-57"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10878737/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of neuroendovascular therapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5797/jnet.cr.2023-0085","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/1 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: Embolic protection devices are useful for preventing distal embolism during carotid artery stenting (CAS); however, complications have been reported. The successful removal of a filter fragment trapped at the distal edge of a carotid stent during the retrieval procedure is described.

Case presentation: CAS was performed for internal carotid artery stenosis in a patient in his 70s, and the carotid stent was successfully placed. During the retrieval procedure, the tip of the filter was trapped at the distal edge of the stent and detached from the filter. Using a snare kit, the filter tip was successfully retrieved, and no postoperative neurological symptoms occurred.

Conclusion: The edge of a carotid stent can potentially trap devices. When trapping or fragmentation of a device is suspected, it is necessary to evaluate the situation and cause, and the device should be appropriately retrieved without using force.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
成功取回被颈动脉支架卡住的过滤器栓塞保护装置碎片:病例报告。
目的:栓塞保护装置可在颈动脉支架置入术(CAS)中有效防止远端栓塞,但也有出现并发症的报道。本文描述了在取回支架过程中成功取出卡在颈动脉支架远端边缘的滤器碎片:病例介绍:一名70多岁的患者因颈内动脉狭窄接受了CAS手术,并成功植入了颈动脉支架。在取回过程中,过滤器的尖端被支架的远端边缘卡住,并从过滤器上脱落。使用套管套件成功取回了过滤器尖端,术后未出现神经症状:结论:颈动脉支架边缘有可能卡住装置。结论:颈动脉支架边缘有可能卡住器械,当怀疑器械卡住或碎裂时,有必要对情况和原因进行评估,并在不使用暴力的情况下适当取回器械。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
A Case of Acute Ischemic Stroke due to Tandem Lesion Treated with Endovascular Thrombectomy by Internal Carotid Artery Direct Puncture. A Case of Cavernous Sinus Dural Arteriovenous Fistula with Persistent Left Superior Vena Cava. A Case of Flow Diverter Placement for a Recurrent Large Thrombosed Middle Cerebral Artery Aneurysm. A Single-Center Experience of Mechanical Thrombectomy for Cancer-Associated Ischemic Stroke. Efficacy and Safety of Lumbar Drainage before Endovascular Treatment for Ruptured Intracranial Aneurysms.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1