{"title":"Biomechanical Comparison of All-Suture, All-Inside Meniscus Repair Devices in a Human Cadaveric Meniscus Model.","authors":"Patrick A Massey, Wayne Scalisi, Carver Montgomery, Drayton Daily, James Robinson, Giovanni F Solitro","doi":"10.1177/19476035241234315","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Newer all-suture, all-inside meniscus repair devices utilize soft suture anchors. The purpose of this study was to compare the biomechanical performance of 4 meniscus repair devices in human cadaver menisci: the JuggerStitch (all-suture, all-inside), the FiberStitch (all-suture, all-inside), a polyether ether ketone (PEEK) all-inside, and an inside-out device.</p><p><strong>Design: </strong>Forty human cadaver menisci were tested after creating 20 mm longitudinal tears in the posterior meniscus. Each knee was randomized to 1 of 4 meniscus repair groups: JuggerStitch (all-suture, all-inside), FiberStitch (all-suture, all-inside), FAST-FIX 360 (PEEK-based anchor all-inside), and inside-out (with Broadband<sup>TM</sup> tape meniscus needles). For each meniscus, 2 devices were used to prepare vertical mattress repair construct. The specimens were tested by pre-conditioning 20 cycles between 5 N and 30 N and then the tear diastasis was measured, followed by distraction to failure phase after imposing a displacement at a rate of 0.5 mm/s.</p><p><strong>Results: </strong>Ten menisci were tested in each of the 4 groups. After pre-conditioning, there was no significant difference in the gap formation among groups (<i>P</i> = 0.212). The average failure load for the JuggerStitch, FiberStitch, PEEK all-inside, and inside-out was 384 N, 311 N, 207 N, and 261 N, respectively, with a significant difference between groups (<i>P</i> = 0.034). <i>Post hoc</i> analysis showed the JuggerStitch failure load was higher than the PEEK all-inside and inside-out (<i>P</i> = 0.005, and <i>P</i> = 0.045, respectively). There was no significant difference between the failure load of the JuggerStitch and FiberStitch (<i>P</i> = 0.225).</p><p><strong>Conclusion: </strong>The JuggerStitch all-suture device, FiberStitch all-suture device, PEEK all-inside, and inside-out devices have similar biomechanical properties for gapping and stiffness. The JuggerStitch all-suture, all-inside device has superior failure load compared with the PEEK all-inside and inside-out repair for longitudinal meniscus tear repair.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/19476035241234315","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Newer all-suture, all-inside meniscus repair devices utilize soft suture anchors. The purpose of this study was to compare the biomechanical performance of 4 meniscus repair devices in human cadaver menisci: the JuggerStitch (all-suture, all-inside), the FiberStitch (all-suture, all-inside), a polyether ether ketone (PEEK) all-inside, and an inside-out device.
Design: Forty human cadaver menisci were tested after creating 20 mm longitudinal tears in the posterior meniscus. Each knee was randomized to 1 of 4 meniscus repair groups: JuggerStitch (all-suture, all-inside), FiberStitch (all-suture, all-inside), FAST-FIX 360 (PEEK-based anchor all-inside), and inside-out (with BroadbandTM tape meniscus needles). For each meniscus, 2 devices were used to prepare vertical mattress repair construct. The specimens were tested by pre-conditioning 20 cycles between 5 N and 30 N and then the tear diastasis was measured, followed by distraction to failure phase after imposing a displacement at a rate of 0.5 mm/s.
Results: Ten menisci were tested in each of the 4 groups. After pre-conditioning, there was no significant difference in the gap formation among groups (P = 0.212). The average failure load for the JuggerStitch, FiberStitch, PEEK all-inside, and inside-out was 384 N, 311 N, 207 N, and 261 N, respectively, with a significant difference between groups (P = 0.034). Post hoc analysis showed the JuggerStitch failure load was higher than the PEEK all-inside and inside-out (P = 0.005, and P = 0.045, respectively). There was no significant difference between the failure load of the JuggerStitch and FiberStitch (P = 0.225).
Conclusion: The JuggerStitch all-suture device, FiberStitch all-suture device, PEEK all-inside, and inside-out devices have similar biomechanical properties for gapping and stiffness. The JuggerStitch all-suture, all-inside device has superior failure load compared with the PEEK all-inside and inside-out repair for longitudinal meniscus tear repair.