Statistical analysis of VEXLPE wear against alumina produced by a new 200-station, multidirectional pin-on-disk device.

IF 1.7 4区 医学 Q3 ENGINEERING, BIOMEDICAL Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine Pub Date : 2024-04-01 Epub Date: 2024-03-05 DOI:10.1177/09544119241235789
Vesa Saikko
{"title":"Statistical analysis of VEXLPE wear against alumina produced by a new 200-station, multidirectional pin-on-disk device.","authors":"Vesa Saikko","doi":"10.1177/09544119241235789","DOIUrl":null,"url":null,"abstract":"<p><p>With low wear rates shown by contemporary bearing materials of total hip prostheses, the standard deviation of wear rate is relatively high. Therefore, large sample sizes are needed for an adequate power of test. Because wear tests take a long time, it is practical to test several samples simultaneously. A new high-capacity, multidirectional wear test device, called the SuperCTPOD-200, was introduced. A 3 million-cycle wear test with an unprecedented sample size of 200 was performed for VEXLPE. The duration of the test was 6 weeks. The wear factor was normally distributed with a mean ± SD of 1.64 × 10<sup>-7</sup> mm<sup>3</sup>/Nm ± 0.22 × 10<sup>-7</sup> mm<sup>3</sup>/Nm (<i>n</i> = 200). The observation that SD was 13.1% of the mean can be useful in power analyses of future tests with other highly cross-linked polyethylenes. Burnishing was the most typical feature on the worn pins, which was in agreement with clinical findings on retrieved acetabular liners. The present study emphasizes statistics that often plays a minor role only in wear studies.</p>","PeriodicalId":20666,"journal":{"name":"Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine","volume":" ","pages":"438-443"},"PeriodicalIF":1.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1177/09544119241235789","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/5 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

With low wear rates shown by contemporary bearing materials of total hip prostheses, the standard deviation of wear rate is relatively high. Therefore, large sample sizes are needed for an adequate power of test. Because wear tests take a long time, it is practical to test several samples simultaneously. A new high-capacity, multidirectional wear test device, called the SuperCTPOD-200, was introduced. A 3 million-cycle wear test with an unprecedented sample size of 200 was performed for VEXLPE. The duration of the test was 6 weeks. The wear factor was normally distributed with a mean ± SD of 1.64 × 10-7 mm3/Nm ± 0.22 × 10-7 mm3/Nm (n = 200). The observation that SD was 13.1% of the mean can be useful in power analyses of future tests with other highly cross-linked polyethylenes. Burnishing was the most typical feature on the worn pins, which was in agreement with clinical findings on retrieved acetabular liners. The present study emphasizes statistics that often plays a minor role only in wear studies.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
对新型 200 工位多向盘上销装置产生的 VEXLPE 对氧化铝的磨损进行统计分析。
由于现代全髋关节假体轴承材料的磨损率较低,磨损率的标准偏差相对较高。因此,需要大量样本才能获得足够的测试能力。由于磨损测试需要较长的时间,因此同时测试多个样本是切实可行的。新型大容量、多方向磨损测试设备 SuperCTPOD-200 问世了。为 VEXLPE 进行了史无前例的 300 万次循环磨损测试,样品数量达到 200 个。测试持续了 6 周。磨损因子呈正态分布,平均值(± SD)为 1.64 × 10-7 mm3/Nm ± 0.22 × 10-7 mm3/Nm (n = 200)。观察发现,SD 为平均值的 13.1%,这对今后使用其他高交联聚乙烯进行测试的功率分析非常有用。烧蚀是磨损销钉上最典型的特征,这与取回的髋臼衬垫上的临床发现一致。本研究强调了在磨损研究中通常只起次要作用的统计数据。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.60
自引率
5.60%
发文量
122
审稿时长
6 months
期刊介绍: The Journal of Engineering in Medicine is an interdisciplinary journal encompassing all aspects of engineering in medicine. The Journal is a vital tool for maintaining an understanding of the newest techniques and research in medical engineering.
期刊最新文献
Improving arterial stiffness prediction with machine learning utilizing hemodynamics and biomechanical features derived from phase contrast magnetic resonance imaging. Synthesis methods of Mg-based scaffolds and their applications in tissue engineering: A review. User experience of lower extremity exoskeletons and its improvement methodologies: A narrative review. A wavelet and local binary pattern-based feature descriptor for the detection of chronic infection through thoracic X-ray images. Optimization and control of robotic vertebral plate grinding: Predictive modeling, parameter optimization, and fuzzy control strategies for minimizing bone damage in laminectomy procedures.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1