[Rupture of Artificial Blood Vessel by a Sternal Wire:Report of a Case].

Q4 Medicine Kyobu geka. The Japanese journal of thoracic surgery Pub Date : 2024-03-01
Masahiro Toyama, Shotaro Higa, Mizuki Ando, Tatsuya Maeda, Yuya Kise, Hitoshi Inafuku, Moriyasu Nakaema, Takaaki Nagano, Kojiro Furukawa
{"title":"[Rupture of Artificial Blood Vessel by a Sternal Wire:Report of a Case].","authors":"Masahiro Toyama, Shotaro Higa, Mizuki Ando, Tatsuya Maeda, Yuya Kise, Hitoshi Inafuku, Moriyasu Nakaema, Takaaki Nagano, Kojiro Furukawa","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Formation of a pseudoaneurysm due to blood leakage from the anastomotic site of the vascular graft in large-diameter vessels is often seen, but formation of a pseudoaneurysm from the non-anastomotic site is extremely rare. A 68-year-old woman presented with a history of double valve replacement for combined valvular disease at 37 years old and hemiarch replacement for thoracic aortic dilatation at 65 years old. She visited the emergency room with a 2-week history of chest pain. Contrast-enhanced computed tomography (CT) revealed a 5-cm-diameter pseudoaneurysm and extravasation from the ascending aorta, so emergency surgery was performed. Around the ascending aorta area, we confirmed bleeding from a 5-mm dehiscence in the non-anastomotic part of the graft prosthesis, so hemostasis was performed with a cross-stitch mattress suture over a felt strip. Initially, the cause of the pseudoaneurysm was unknown, but re-examination of CT images from after the previous hemiarch replacement confirmed contact between the sternal wire and graft prosthesis. The wire was thus considered to have caused damage and bleeding. The patient was discharged from the hospital with a good postoperative course and is being followed-up in the outpatient department.</p>","PeriodicalId":17841,"journal":{"name":"Kyobu geka. The Japanese journal of thoracic surgery","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Kyobu geka. The Japanese journal of thoracic surgery","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Formation of a pseudoaneurysm due to blood leakage from the anastomotic site of the vascular graft in large-diameter vessels is often seen, but formation of a pseudoaneurysm from the non-anastomotic site is extremely rare. A 68-year-old woman presented with a history of double valve replacement for combined valvular disease at 37 years old and hemiarch replacement for thoracic aortic dilatation at 65 years old. She visited the emergency room with a 2-week history of chest pain. Contrast-enhanced computed tomography (CT) revealed a 5-cm-diameter pseudoaneurysm and extravasation from the ascending aorta, so emergency surgery was performed. Around the ascending aorta area, we confirmed bleeding from a 5-mm dehiscence in the non-anastomotic part of the graft prosthesis, so hemostasis was performed with a cross-stitch mattress suture over a felt strip. Initially, the cause of the pseudoaneurysm was unknown, but re-examination of CT images from after the previous hemiarch replacement confirmed contact between the sternal wire and graft prosthesis. The wire was thus considered to have caused damage and bleeding. The patient was discharged from the hospital with a good postoperative course and is being followed-up in the outpatient department.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[胸骨线导致人造血管破裂:病例报告]。
在大直径血管中,由于血管移植吻合口部位的血液渗漏而形成假性动脉瘤的情况经常出现,但非吻合口部位形成假性动脉瘤的情况则极为罕见。一名 68 岁的女性患者曾在 37 岁时因合并瓣膜病接受过双瓣膜置换术,在 65 岁时因胸主动脉扩张接受过半弓置换术。她因胸痛两周前来急诊就诊。对比增强计算机断层扫描(CT)显示升主动脉有直径 5 厘米的假性动脉瘤和外渗,因此进行了急诊手术。在升主动脉周围,我们确认移植假体的非吻合口部位有一个 5 毫米的裂口出血,因此用交叉缝合褥式缝合毡条进行止血。最初,假性动脉瘤的原因不明,但重新检查上次半弓置换术后的 CT 图像后证实,胸骨钢丝与移植物假体之间存在接触。因此认为是钢丝造成了损伤和出血。患者术后恢复良好出院,目前正在门诊部接受随访。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
213
期刊最新文献
[Concurrent Thymoma and Mature Teratoma in the Mediastinum]. [Coronary Artery Bypass Grafting in a Patient with Hemophilia A Assisted by Thromboelastography: Report of a Case]. [Double-chambered Right Ventricle with Difficult Preoperative Diagnosis of Ventricular Septal Defect in Adulthood]. [Mediastinal Dedifferentiated Liposarcoma Reconstructing Defected Chest Wall Using Prosthetic Materials:Report of a Case]. [Minimally Invasive Aortic Valve Replacement for Aortic Valve Infective Endocarditis Complicated by Septic Arthritis of the Sternoclavicular Joint:Report of a Case].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1